Sartorius Susceptometer - for Precise Measurement of: Susceptibility and Magnetization of Weights

Sartorius Susceptometer - for Precise Measurement of: Susceptibility and Magnetization of Weights. Benno Gatzemeier Market Manager Mass Metrology Sartorius AG / Germany June 2007. : Contents. : Introduction – Magnetic Properties of weights : Susceptometer Method

mirielle
Download Presentation

Sartorius Susceptometer - for Precise Measurement of: Susceptibility and Magnetization of Weights

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Sartorius Susceptometer - for Precise Measurement of:Susceptibility and Magnetizationof Weights Benno Gatzemeier Market Manager Mass Metrology Sartorius AG / Germany June 2007

  2. : Contents • : Introduction – Magnetic Properties of weights • : Susceptometer Method • : The Sartorius Susceptometer • : Calibration Procedure and Factory Calibration • : Long term stability of md • : Comparison Measurement

  3. : Introduction • Influence Parameters in Mass Comparison : • Air buoyancy • Contamination • Air draft • Object temperature • Magnetic properties • The golden rule in metrology is: • Factors that influence the measurement • are switched off, kept constant or considered. • Magnetic properties

  4. N S N N S S : Magnetic properties • The OIML R111 recommends to check the magnetic properties. Magnetization Susceptibility Standard Weights with a Susceptibility F Magnetic Forces Magnetic Forces F Standard Weight with Magnetization Magnet

  5. : The new OIML R111

  6. H - Hz : Susceptibility  and Magnetization 0M (µT) • Susceptibility  Magnetization 0M (µT) H

  7. : Recommended methods regarding the R111

  8. : Recommended methods regarding the R111 • Permeability Indicator Gauss meter

  9. Susceptibility:  =f (Fa...) F1 + F2 F1 - F2 Fa = Fb = 2 2 F2 = -  m2 * g F1 = - m1 * g F2 F1 µ0MZ =f (Fb...) Magnetization: : The Susceptometer Principe - Regarding the OIML R111

  10. F : The Susceptometer Principe - Regarding the OIML R111 • The R111 describes methods for the determination of the magnetic properties. • One of them is the Susceptometer principle. • A) Magnet • B) Weighing Pan • C) Bridge • D) Gauge blocks • Test Weight • Pedestal

  11. : Sartorius Suszeptometer • The building guidance was the R111: • A micro mass comparator • Internal magnet • 5 different distances Z0 • Load plate for weights up to 50 kg • Software to compute the formulae • Determination of:- Susceptibility ““- Magnetization “0M” (T)

  12. Magnet with md produces a maximum field H Field H should not exceed initially: H 2000 A/m when testing class E1 H 800 A/m when testing class E2 H 200 A/m for classes F1 and F2. Table 1: Initial values for testing class E1, E2, F1 and F2, magnetic (dipole) momentmd0.1 Am2 : Vertical Distances Z0; Magnet <-> Weight This is important to avoid permanent magnetization. Distance may be reduced only if the Susceptometer signal is too weak.

  13. : Computation factors for Susceptibility and Magnetization: • Weighing Result of the Magnet H • Distance Z0 : Magnet <-> Weight H • Geometry of the test weight S • magnetic (dipole) moment md [Am2] S • gravitational acceleration [m/s2] S • Local magnet field BEZ–48-60 [µT] S • To measure the Magnetization, we have to rotate the magnet! H

  14. : Calculation of the Magnetic properties • Calculation of the susceptibility Calculation of the Magnetization

  15. : The vertical rotation mechanics of the magnet • Changes the orientation of the magnet • Parts: • Magnet • Pedestal • Gear • Knob

  16. : Application Software • Easy operating • Step by step guide through the measurement procedure • Initial distance is proposed • Results via a serial connection • Calculations, report and export • Recalibrating the necessary constants • Default parameters and user defined configurations • Shape description, OIML knob weights predefined • Export and import function for the sharp of the weights

  17. : 1. Select weighing geometry

  18. : Own cylinder - Geometry of the test weight

  19. : 2. Input parameter

  20. : 3. Remove test weight

  21. : 4 Adjust vertical position Z2

  22. : 5. Adjust test magnet to position “N”

  23. : 6. Tare balance : 7. Place test weight

  24. : 8. Determine measured value m1 for Z4 : 9. Remove test weight

  25. : 10. Adjust test magnet to position “S”

  26. : 11. Tare balance : 12. Load test weight

  27. : 13. Determine measured value m2 for Z5 : 14. Remove test weight

  28. : Push result button

  29. : Technical specifications Sartorius Susceptometer • Base area 338 x 286 mm • Height 249 mm • Maximum load 50 kg • Dipole moment of the magnet m ~ 0.1 Am2 • Geometry ratio of the magnet h/d = 0.87 • Height Z0 adjustable in fixed steps Z1=18 / Z2=20 / Z3=27 / Z4=35 / Z5=43mm • Field strength 2700 / 2000 / 800 / 360 / 200 A/m • Readability of the Mass Comparator 10 µg or 1 µg • Reverse gear for magnet external rotary knob with N-S marking

  30. Calibration, check of the Susceptometer • Calibration of the Mass comparator (10 g) • Using a Susceptibility Referencewith certificate of the susceptibility • Measure the Susceptibility Reference on the Sartorius Susceptometer • Compare the result of the Susceptometer with the PTB-certificate. • The difference has to be less than 10%

  31. Factory calibration • We use a 1 kg stainless steel susceptibility standard (=0.004069) • Additional information is used as check for the factory calibration: • Value of the vertical distance Z0 from the mechanical adjustments in the manufacturing • We use always the same three additional magnets.historical data (md )

  32. F1-2 F2-3 F1-4 F1-3 F2-4 F3-4 PTB Calibration procedure / Adjustment • Calibration of the mass comparatoruses a 10 g weight • Calibration of the dipole moment md,uses 3 additional magnets and measure the forces between each pair of magnets6 equations and 4 unknown dipole moments • Calibration of the distance Z0,uses a susceptibility standard at known 

  33. Comparison Measurement • Our references Susceptibilities • 4 x NPL Standards 1 x PTB Standard • Question: Calibration with susceptibility standard : =0.00401 • Application range : 0 <   1

  34. Our Susceptibility standards

  35. Comparison Measurement

  36. Long term stability of our reference Susceptibility  • The change of the Susceptibility is in the range of the uncertainty and less than 2 %

  37. Long term stability of our reference magnets md

  38. Thank you for your attentionBenno GatzemeierMass MetrologySartorius AG / Germany