Sartorius Susceptometer - for Precise Measurement of: Susceptibility and Magnetization of Weights. Benno Gatzemeier Market Manager Mass Metrology Sartorius AG / Germany June 2007. : Contents. : Introduction – Magnetic Properties of weights : Susceptometer Method
E N D
Sartorius Susceptometer - for Precise Measurement of:Susceptibility and Magnetizationof Weights Benno Gatzemeier Market Manager Mass Metrology Sartorius AG / Germany June 2007
: Contents • : Introduction – Magnetic Properties of weights • : Susceptometer Method • : The Sartorius Susceptometer • : Calibration Procedure and Factory Calibration • : Long term stability of md • : Comparison Measurement
: Introduction • Influence Parameters in Mass Comparison : • Air buoyancy • Contamination • Air draft • Object temperature • Magnetic properties • The golden rule in metrology is: • Factors that influence the measurement • are switched off, kept constant or considered. • Magnetic properties
N S N N S S : Magnetic properties • The OIML R111 recommends to check the magnetic properties. Magnetization Susceptibility Standard Weights with a Susceptibility F Magnetic Forces Magnetic Forces F Standard Weight with Magnetization Magnet
H - Hz : Susceptibility and Magnetization 0M (µT) • Susceptibility Magnetization 0M (µT) H
: Recommended methods regarding the R111 • Permeability Indicator Gauss meter
Susceptibility: =f (Fa...) F1 + F2 F1 - F2 Fa = Fb = 2 2 F2 = - m2 * g F1 = - m1 * g F2 F1 µ0MZ =f (Fb...) Magnetization: : The Susceptometer Principe - Regarding the OIML R111
F : The Susceptometer Principe - Regarding the OIML R111 • The R111 describes methods for the determination of the magnetic properties. • One of them is the Susceptometer principle. • A) Magnet • B) Weighing Pan • C) Bridge • D) Gauge blocks • Test Weight • Pedestal
: Sartorius Suszeptometer • The building guidance was the R111: • A micro mass comparator • Internal magnet • 5 different distances Z0 • Load plate for weights up to 50 kg • Software to compute the formulae • Determination of:- Susceptibility ““- Magnetization “0M” (T)
Magnet with md produces a maximum field H Field H should not exceed initially: H 2000 A/m when testing class E1 H 800 A/m when testing class E2 H 200 A/m for classes F1 and F2. Table 1: Initial values for testing class E1, E2, F1 and F2, magnetic (dipole) momentmd0.1 Am2 : Vertical Distances Z0; Magnet <-> Weight This is important to avoid permanent magnetization. Distance may be reduced only if the Susceptometer signal is too weak.
: Computation factors for Susceptibility and Magnetization: • Weighing Result of the Magnet H • Distance Z0 : Magnet <-> Weight H • Geometry of the test weight S • magnetic (dipole) moment md [Am2] S • gravitational acceleration [m/s2] S • Local magnet field BEZ–48-60 [µT] S • To measure the Magnetization, we have to rotate the magnet! H
: Calculation of the Magnetic properties • Calculation of the susceptibility Calculation of the Magnetization
: The vertical rotation mechanics of the magnet • Changes the orientation of the magnet • Parts: • Magnet • Pedestal • Gear • Knob
: Application Software • Easy operating • Step by step guide through the measurement procedure • Initial distance is proposed • Results via a serial connection • Calculations, report and export • Recalibrating the necessary constants • Default parameters and user defined configurations • Shape description, OIML knob weights predefined • Export and import function for the sharp of the weights
: 6. Tare balance : 7. Place test weight
: 8. Determine measured value m1 for Z4 : 9. Remove test weight
: 11. Tare balance : 12. Load test weight
: 13. Determine measured value m2 for Z5 : 14. Remove test weight
: Technical specifications Sartorius Susceptometer • Base area 338 x 286 mm • Height 249 mm • Maximum load 50 kg • Dipole moment of the magnet m ~ 0.1 Am2 • Geometry ratio of the magnet h/d = 0.87 • Height Z0 adjustable in fixed steps Z1=18 / Z2=20 / Z3=27 / Z4=35 / Z5=43mm • Field strength 2700 / 2000 / 800 / 360 / 200 A/m • Readability of the Mass Comparator 10 µg or 1 µg • Reverse gear for magnet external rotary knob with N-S marking
Calibration, check of the Susceptometer • Calibration of the Mass comparator (10 g) • Using a Susceptibility Referencewith certificate of the susceptibility • Measure the Susceptibility Reference on the Sartorius Susceptometer • Compare the result of the Susceptometer with the PTB-certificate. • The difference has to be less than 10%
Factory calibration • We use a 1 kg stainless steel susceptibility standard (=0.004069) • Additional information is used as check for the factory calibration: • Value of the vertical distance Z0 from the mechanical adjustments in the manufacturing • We use always the same three additional magnets.historical data (md )
F1-2 F2-3 F1-4 F1-3 F2-4 F3-4 PTB Calibration procedure / Adjustment • Calibration of the mass comparatoruses a 10 g weight • Calibration of the dipole moment md,uses 3 additional magnets and measure the forces between each pair of magnets6 equations and 4 unknown dipole moments • Calibration of the distance Z0,uses a susceptibility standard at known
Comparison Measurement • Our references Susceptibilities • 4 x NPL Standards 1 x PTB Standard • Question: Calibration with susceptibility standard : =0.00401 • Application range : 0 < 1
Long term stability of our reference Susceptibility • The change of the Susceptibility is in the range of the uncertainty and less than 2 %
Thank you for your attentionBenno GatzemeierMass MetrologySartorius AG / Germany