1 / 13

Dave Lowry, Rebecca Fisher, Mathias Lanoisell é, Sri Sriskantharajah and Euan Nisbet

Picarro Cavity Ring Down Spectrometry (CRDS) for High-Precision Continuous Monitoring of Atmospheric CO 2 and CH 4. Dave Lowry, Rebecca Fisher, Mathias Lanoisell é, Sri Sriskantharajah and Euan Nisbet

monifa
Download Presentation

Dave Lowry, Rebecca Fisher, Mathias Lanoisell é, Sri Sriskantharajah and Euan Nisbet

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Picarro Cavity Ring Down Spectrometry (CRDS) for High-Precision Continuous Monitoring of Atmospheric CO2 and CH4 Dave Lowry, Rebecca Fisher, Mathias Lanoisellé, Sri Sriskantharajah and Euan Nisbet Department of Earth Sciences, Royal Holloway, University of London, Egham, Surrey TW20 0EX, United Kingdom Presented at: APRIL Meeting, London, 12/09 Funded by NERC since 2008

  2. CRDS Theory • When the laser is on, the cavity quickly fills with circulating laser light. A fast photo-detector senses the small mount of light leaking through one of the mirrors (not 100% reflective) to produce a signal that is directly proportional to the intensity in the cavity. • When the photodetector signal reaches a threshold level (in a few tens of microseconds), the continuous wave (CW) laser is abruptly turned off. The light already within the cavity continues to bounce between the mirrors (about 100,000 times). • The light intensity inside the cavity steadily leaks out and decays to zero in an exponential fashion. This decay, or "ring down", is measured in real-time by the photodetector. http://www.picarro.com/technology/whatiscrds.php

  3. The Model We Use • Picarro Inc.WS-CRDS Analyzer for CO2/CH4/H2O in air - Model G1301 (Made in California) • Wavelength-specific • Simultaneous, precise measurement of CO2, CH4 and H2O How much does it cost? At current exchange rates about £40,000 once installed Automated inlet parts - £7,000 Inlet control software development - £5,000 Calibration gas suite - £3,000

  4. Tasks • Instrument Tests • Standardization with NOAA calibration gases • Precision • Water vapour correction • Long-term drift • Development of Automated Inlet • Software • Automated valves and flow • Deployment at Remote Sites • South Atlantic islands • RHUL

  5. Picarro G1301

  6. NOAA standardization • Currently 6 NOAA standards in use in lab - 2 ambient air, 4 synthetic • CRDS only measures 12CCO2 and 12CCH4 peaks • Synthetic are depleted in 13CCO2 and enriched in 13CCH4 so extra correction necessary if not using ambient air standards

  7. Precision - International Requirements for Background Stations • Northern Hemisphere - CO2 ± 0.05 ppm, CH4 ± 2.0 ppb • Southern Hemisphere - CO2 ± 0.025 ppm, CH4 ± 1.0 ppb • Residuals for 5 NOAA standards - CO2 ± 0.018 ppm, CH4 ± 0.16 ppb • Achieved on 10 min ave (120 points) - CO2 ± 0.018 ppm, CH4 ± 0.15 ppb

  8. 0.01898 ppm (18.98 ppb) / 1% H2O 5.075 ppm / 1% H2O Water Vapour Correction = raw CH4 (ppb) x (1 + %H2O x constant) Calculated constant = 0.0101109 Nafion reduces correction from ≈27 to 2 ppb • Factory applied for CO2 but…. • Currently no consensus for the CH4 correction • Calculated by having two instruments side by side on the same air line, one of which has a Nafion-dried airstream - reduces water from 1.5% to 0.1%

  9. Long Term Drift • CO2 - no obvious drift - up to 0.1 ppm fluctuation • CH4 drifted by 3-4 ppb in 1 year (less when air is dried) - <0.2%

  10. Automated Inlet Development 6 standard bottles: 3 calibration 1 target 1 zero 1 long term

  11. Automated Inlet Software • Controls cylinder outlet pressure at 0.5 bar (7 psi) • Controls flow at 300-500 cc/min • Gives real-time data display • Creates calibration file and makes drift correction between calibrations • Remote data transfer capabilities

  12. Deployment in the South Atlantic • Falkland Islands and Ascension Island • Close to airports for communications

  13. Deployment at RHUL • Very local source now affecting all air from W-NW during working hours Source ? LiCor P2 P3

More Related