Lecture 21

Lecture 21. UV/Vis Pick up Lecture Problem 7. This Week in Lab: Work on 1st Synthetic Next Week in Lab: Ch 9 Final Report Due Synthetic #2 PreLab Due. UV/Vis Spectroscopy. Basic Idea: Expose compound to UV/Vis radiation UV: 200-400 nm Vis: 400-800 nm

naoko
Download Presentation

Lecture 21

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Lecture 21 • UV/Vis • Pick up Lecture Problem 7 • This Week in Lab: • Work on 1st Synthetic • Next Week in Lab: • Ch 9 Final Report Due • Synthetic #2 PreLab Due

  2. UV/Vis Spectroscopy Basic Idea: Expose compound to UV/Vis radiation UV: 200-400 nm Vis: 400-800 nm 2. Absorption of radiation causes electronic excitations at specific wavelengths (lmax) Main Use: Detects the presence of pi systems in a compound. Unlike NMR, IR, and MS, UV/Vis is NOT used to determine pieces (specific bonds) of a molecule.

  3. Electronic Transitions Types of Electronic Transitions: The length of the arrow is roughly proportional to the energy difference between two levels. The longer arrow means a higher Energy transition. Higher energy means lower wavelength of radiation.

  4. UV/Vis Spectroscopy Transitions observed by UV/Vis: p to p* (strong) n to p*(weak, often not observed) n to p* transition is lower energy (higher wavelength) p to p* transition is higher energy (lower wavelength)

  5. UV/Vis of Acetone p --> p* Shorter wavelength, stronger absorption n --> p* Longer wavelength, weaker absorption

  6. The more conjugation, the lower the energy, the higher the Wavelength (lmax value). p to p* transitions

  7. Quantitative Data from UV/Vis Extinction Coefficients or Molar Absorptivity (): A numerical value of the intensity of the absorptions. Calculate  by using the Beer-Lambert Law: A (Absorbance) =  b c Where b = path length of cell (usually 1.0 cm) c = concentration of the solute When reporting data, chemists only report lmax’s and ‘s from UV/Vis data.

  8. Plant Pigments (Chapter 9) Color wheel: complementary colors

  9. Plant Pigments (Chapter 9) • What you’re looking for: • Example: Chlorophyll: • Observe the visible color: green • Absorption of red light (~ 620 to 680 nm); should be low % • transmittance