Microelectrode Recording

Microelectrode Recording. Using the FHC micro-drive. Microelectrode Recording. Calculate the micro-drive starting point Assemble the FHC micro-drive Insert the cannulae Insert the microelectrode Additional tracks: Perform a parallel track Using the offset positioner Place the DBS lead.

neo
Download Presentation

Microelectrode Recording

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Microelectrode Recording Using the FHC micro-drive

  2. Microelectrode Recording • Calculate the micro-drive starting point • Assemble the FHC micro-drive • Insert the cannulae • Insert the microelectrode • Additional tracks: • Perform a parallel track • Using the offset positioner • Place the DBS lead

  3. Depth to target is from alignment adapter Offset calculates for FHC stackup Allowable targeting range is 150-180mm Starting Point Calculation

  4. FHC Calculation

  5. Remove alignment adapter Insert Center positioner Attach drive base Attach micro-drive FHC Assembly Note: be sure drive is set to starting depth

  6. Insert stylet through passing cannula Attach collet to passing cannula Pass cannula/collet combo into drive Remove Stylet Insert microelectrode cannula Cannulae insertion Note: be sure drive is set to starting depth

  7. Retract microelectrode within sheath Pass microelectrode through cannula Advance electrode Connect recording system cables Microelectrode insertion Note: be sure drive is set to starting depth

  8. Microelectrode Recording

  9. Parallel Tracts • Retract drive to starting point • Remove cannula & re-insert through offset

  10. Offset Positioner • Retract drive to starting point • Remove cannula • Remove drive and drive base • Remove center positioner • Choose direction for offset • Insert offset positioner • Re-assemble drive and cannula

  11. Center Positioner at 45° Center Positioner 3mm offset at “B” 3mm offset at “G” Possible Tracks

  12. Lead Placement • Set drive to desired target depth • Remove microelectrode and spacer cannula • Attach lead holder • Insert lead - depth stop at 347mm