CTA Physics and Dosimetry

CTA Physics and Dosimetry. Nausheen Akhter, MD Division of Cardiology University of Illinois at Chicago December 21, 2007. Contents. Review of X-ray Production CT Components CT Scanner Generations CT Image Formation and Reconstruction Retrospective vs. Prospective Digitalization

nhi
Download Presentation

CTA Physics and Dosimetry

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. CTA Physics and Dosimetry Nausheen Akhter, MD Division of Cardiology University of Illinois at Chicago December 21, 2007

  2. Contents • Review of X-ray Production • CT Components • CT Scanner Generations • CT Image Formation and Reconstruction • Retrospective vs. Prospective • Digitalization • Hounsfield Unit • Radiation Dosimetery

  3. The CT Scan Defined • Computed tomography is an imaging modality that uses x-rays to image cross-sectional slices through the body. • X-rays are directed into a localized section of the patient. • Attenuated x-rays completely penetrate the patient and are detected. • Detected signals are then constructed into an image.

  4. Three paths of an electron (e-): (1) e- may directly collide with nucleus (unlikely) (2) e- may enter vicinity of nucleus, but not collide (3) e- collides with inner shell e- and outer shell e- shifts position X-ray Production

  5. X-ray Production • Paths (1) and (2) cause e- to lose much energy which is converted into x-ray photons. • Bremsstrahlung phenomena (aka. braking radiation): x-ray photons are created by paths (1) and (2) • Characteristic radiation: x-ray photons created by path (3) • 10-12% of x-ray beam emitted from CT x-ray tube

  6. X-ray Production

  7. CT Components • The gantry: • Aperture • Contains a rotating frame assembly (x-ray tube, detector array, data acquisition system) • Contains slip-ring and high voltage generator

  8. The patient table: X: Left to Right (saggital) Y: Anterior to Posterior (coronal) Z: Head to Feet (axial) CT Components

  9. The x-ray tube Generates and focuses the x-ray beam to strike a target material (Tungsten) and produce photons. Two electrodes: rotating anode and cathode High voltage generator CT Components

  10. Tube Current (mA) • When tube voltage (kV) is applied across the cathode and anode at a specific current (mA), the e- beam flows from the cathode and collides with the target material on the anode. • The cathode contains a heating filament wire. The temperature of this filament controls the current of e-. • Temperature of the filament is controlled by the tube current (mA). • The tube current (mA) determines the number of x-ray photons produced. • Higher mA  hotter filament  greater e- escape and create photons  greater x-ray dose to the patient.

  11. Tube Voltage (kV) • Tube voltage (kV) controls the x-ray photon energy level. • X-ray photon energy level determines how easily the x-ray penetrates the patient’s body and image quality. • 120 kV are higher energy and more penetrating than 100 kV.

  12. Collimation Prevents ‘scatter radiation’ and minimizes radiation dose Restrict the path of x-rays and define slice thickness Two types: pre-patient and post-patient CT Components

  13. CT Components • Detectors • In order for an x-ray photon to generate a signal, four steps must occur: (1) X-ray must enter a detector (ie. ‘capture’) (2) X-ray must collide with a detector atom (3) The collision must result in an electromagnetic conversion suitable for measurement (eg. Light) (4) This event must be amplified and conducted to a data acquisition system.

  14. Detectors Two types: gas (xenon) and solid-state Solid-state (aka. scintillation detector) is made of a solid crystalline substance which is more sensitive to x-rays at various angles. CT Components

  15. Linear Attenuation Coefficient • X-ray photons are attenuated via two processes: absorption and scatter • Amount of attenuation is dependent on: • Atomic number of the tissue, density of electrons in the tissue, thickness of the tissue, and energy of the x-ray photons • Lower energy photons are more easily attenuated than higher energy photons.

  16. Linear Attenuation Coefficient • Exponential decay • The number of photons detected is reduced in an exponential fashion as the thickness of the tissue traversed is increased.

  17. CT Components • Array processor • Primary function is the reconstruction of the projected attenuation raw data into a CT image.

  18. First Generation X-ray tube generated a thin ‘pencil beam’. Single detector Tube and detector translated across gantry while patient stationary Rotation angle 1 degree Scan time: 5 minutes CT Scanner Generations

  19. Second-generation X-ray ‘fan beam geometry’ Detector array (30) Rotation angle >5 degrees Scan time: 20 sec CT Scanner Generations

  20. Third generation Fan-beam x-rays Curved arc detector array (improved reproducibility of data, >750 detectors) Tube/detector array rotates around patient with a ‘continuous slip ring’ Scan time: < 1 sec Fourth generation CT Scanner Generations

  21. Scanning methods: The scout film Conventional/Axial/ Sequential CT Helical/Spiral CT CT Image Formation

  22. CT Image Formation • Conventional/Axial/Sequential CT • The x-ray tube rotates around a stationary patient. The patient table must be incremented after each image is acquired. • For a third generation scanners, a ‘projection’ is made up of all the x-rays emanating from the tube in one position. • Data measured at each projection is called the raw data.

  23. Helical/Spiral CT Continuous data collection. Tube rotates around the patient as the patient table moves at a fixed speed through the gantry. Need slip-ring scanner and high heat capacity tube The slice thickness of a helical scan is defined as the table increment/pitch CT Image Formation

  24. Advantages vs. Disadvantages • Advantages of Helical Scanning • More coverage in a breath-hold • No misregistration of slices • Less contrast injection • Gapless slices • Images may be constructed at any arbitrary position • Multiplanar reconstruction and 3D • Disadvantage of Helical Scanning • Continuous radiation

  25. Image Reconstruction • Most CT systems use a ‘filtered back projection algorithm’ for reconstruction. • Back projection is a summation technique of all of the projections. • A correction filter (aka. reconstruction filter or kernel) is applied to accentuate the edges and allow for exact representation of the original object. • ‘Convolution’ is the process of applying filtration.

  26. Filtered Back Projection

  27. Retrospective Reconstruction • Simultaneous recording of ECG with continuous x-ray data acquisition. • The ECG is retrospectively used to assign images to the respective phases of the cardiac cycle. • Retrospective, phase-specific segments of the R-R intervals are combined to reconstruct an axial slice. • Strength: oversampling allows for gap-less and motion-reduced imaging • Drawback: higher radiation

  28. Retrospective Reconstruction

  29. Prospective Reconstruction • “Step and shoot modality” • Images obtained at a pre-determined offset of the ECG-detected R wave • Only 1 image per detector per cardiac cycle • Strength: lower radiation • Drawbacks: inability to perform overlapping images, longer acquisition time

  30. Multi-planar Reconstruction and 3D • Multi-planar reconstruction is a post-processing technique that requires only the image data to reconstruct images in planes other than the plane of the original scan. • 3D surface rendering is another post-processing technique that reconstructs images of the surfaces of anatomical structures.

  31. Contrast Media Injection • Scan delay is determined by the contrast transit time • Bolus tracking technique • The coronary CT imaging sequence is initiated when the contrast enhancement in the ascending aorta reaches 100 Hu. • Timing bolus technique • Images acquired at level of carina every 2s starting 10s after injection of constrast (10-20 mL), followed by saline flush (30-50 mL).

  32. Timing Bolus Technique, Scan delay 20s

  33. Digitalization • The first step in generating a digital image is to divide the object into a grid of small regions. • Each region of the grid is called a ‘pixel’. • A grid of pixels is the image matrix. • Assign shades of gray to the pixels. • Gray-scale values for pixels within the reconstructed tomogram are defined as CT numbers or “Hounsfield units” (HU).

  34. Hounsfield Unit (HU) • CT uses X-ray absorption to create images which differ in brightness depending on their physical density (atomic composition). • HU are defined with a reference to the value for water. • Higher densities will be brighter, lower densities darker.

  35. Hounsfield Unit (HU)

  36. CT Parameters • Spatial Resolution • Measure of the size of the smallest object that can be visualized in an image • Directly related to the slice thickness and reconstruction matrix • Best cardiac CT spatial resolution (x, y, and z-axes) is 0.6 x 0.6 x 3mm (nearly cubic/isentropic voxel) • Temporal Resolution • Precision of a measurement with respect to time • High temporal resolution needed to minimize coronary motion artifact • Determined by pitch, gantry rotation time, and the patient’s heart rate

  37. Radiation Dosimetery • 50% of a person’s lifetime radiation exposure is due to medical testing. • Three primary factors that go into radiation dosimetry: x-ray energy (kV), tube current (mA), exposure time. • ‘Effective dose’ is an estimate of the uniform, whole-body equivalent dose that would produce the same level of risk for adverse effects that results from the non-uniform partial body irradiation. • The unit for the effective dose is the milliSievert (mSv).

  38. Radiation Dosimetery

  39. Radiation Dosimetery • There are various choices with MDCT that can dramatically change the patient radiation dose. • Dose modulation was a large advancement. • Tube current (mA) is reduced during systole by 80%, then the full dose is used in diastole (40-80% R-R interval) • May reduce radiation exposure from 30-50% • Slower heart rates, less systole, reduction is less.

  40. Radiation Dosimetery • Other methods to decrease radiation: • Current (mA) can be increased/decreased, most often based on body habitus. • Increase the pitch (table speed/collimator width) • Lower pitch = more radiation • Slice thickness can be increased (faster scan). • Prospective reconstruction

  41. Articles to Review • Morin RL et al. “Physics and dosimetry in computed tomography” Cardiol Clin 21 (2003), pp. 515-520 • Budoff et al. Cardiac CT Imaging: Diagnosis of Cardiovascular Disease • Chapters 1 and 2 • CT Registry Review Program • Chapters 1, 2 and 3 • Budoff et al. “ACCF/AHA Clinical Competence Statement on Cardiac CT and MR” JACC Vol. 46, No. 2, 2005