DP-IR Detecto -Pak Infrared. DP-IR Features. Methane sensitivity 1 ppm detection Built in Self-test 0-100% dynamic range auto ranging and fixed Long use battery No hydrogen fuel Simple to use Menu is similar to RMLD Light weight Same walking and mobile survey methods as DP-4 FID.
E N D
DP-IR Features • Methane sensitivity • 1 ppm detection • Built in Self-test • 0-100% dynamic range • auto ranging and fixed • Long use battery • No hydrogen fuel • Simple to use • Menu is similar to RMLD • Light weight • Same walking and mobile survey methods as DP-4 FID
DP-IR Features • Simple user interface menu and operation • High Sensitivity: 1ppm • Fast response: < 1 sec • Tick or DMD mode: Adjustable alarm level: 1 to 9000 • Extended range to 100% VOL • Auto scaling and fixed range • Built in Self-test with internal cal cell • Internal rechargeable battery or external power • >8 hrs @ 32°F battery operating life • CE compliance • Intrinsic Safety approval • RS232 or Bluetooth optional • Weight: ~ 5.6lbs • Internal logging of self-test with time stamp
DP-IR Use • Surface sampling • Bar-hole testing • Indoor meter / piping • Portable application • Mobile application
Using the DP-IR • Replace dust filter by removing the front dust cap and changing out the filter. Replace the filter at minimum on a daily basis. • Turn on the instrument by pressing the On button and holding until starting • Instrument will count down to indicate that it is warming up. Once warmed up, the instrument will go to the run screen. • Perform Self-test • If Self-test is OK, then the instrument is ready for survey work. • Verify Alarm is on and the detection level is the desired amount. • The instrument may continue to stabilize for up to 15 minutes. Re-zero as needed. • Recharge instrument after use.
Survey Detection Modes • The DP-IR has two survey detection modes: DMD and Tick. • The DMD will alarm when the gas concentration exceeds the Alarm Level • The Tick mode will generate a pulse that is relative to actual gas concentration.
Easy to Operate • Menu: Press to select menu options • Up: Press to increase or select • Down: Press to decrease or deselect • Alarm: Press to enable audible buzzer • Zero: Press to zero the instrument • Power: Press and hold to turn on or off
Survey Run Screen • Zeroing • Press the Zero button to zero the instrument • Disable the Alarm • Press the Alarm button to temporally turn off the audible buzzer. It will reactivate in 5 minutes. • Display range • Auto range will switch from ppm to % gas at 1% • Press the Down button to select fixed ranges (ppm, % gas) • Press the Up button to return to auto ranging
Menu Options • Alarm Level • Use Up/Down button to adjust the detection threshold • Use Zero button to toggle between DMD and Tick mode • Press Alarm button to turn off the alarm disable time out • Press Menu to accept changes and return to run screen • Self-Test • Use Up button to initiate the Self-Test • Press Menu to return to run screen • SEL • Use Up button to turn back light on • Use Down button to turn back light off • Press Zero button to toggle between RS and BT com • Press Menu to accept changes and return to run screen
Other • Battery indicator shows the remaining battery life in 12.5% increments. • Peak indicator (0-1000ppm) holds peak for 2 seconds • Flow Error • Alarm will sound when the flow is restricted
Battery Charging • LI-ion battery can be charged without worry about memory build up • Depending on the supply voltage level of the vehicle battery, the DP-IR battery may also recharge while the instrument is powered through external power.
Bar Holing • Short probe comes standard • 28” probe as option • Other optional probes available • Use auto ranging or fixed % gas • Very fast clearing in the % gas range
Mobile Application • FID to DP-IR coupling adaptor available • Need to match DP-IR intake flow rate • External pump required / sample system • Street / off-road vehicle