Prepare the patient for endoscopy. Inspect the esophagus and stomach endoscopically and then remove the endoscope.

Download Presentation

Orbera365 Weight Loss Balloon System Placement and Inflation

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Orbera365™ Weight Loss Balloon System Placement and Inflation (Step-By-Step)

  2. Prepare the patient for endoscopy. Inspect the esophagus and stomach endoscopically and then remove contraindications, insert the Placement Catheter Assembly containing the balloon gently down the esophagus and into the stomach. the endoscope. If there are no Visit http://orbera365.net/ for more information

  3. 1. Prepare the patient according to hospital protocol for gastroscopy. 2. Perform gastroscopic inspection of esophagus and stomach. 3. Remove gastroscope. 4. Where there are no contraindications: a. Lubricate the ORBERA365™ System placement sheath with surgical lube-gel. b. Move the ORBERA365™ System gently down the esophagus and into the stomach. 5. Reinsert the endoscope while the balloon is in situ to observe filling steps. The balloon must be below the lower esophageal sphincter and well within the stomach cavity.

  4. 6. Remove the guide wire from the fill tube. 7. Attach the 3-way stopcock and 50 cc syringe to the Luer-Lock. Insert filling kit spike into the saline bag fill tube. 8. Slowly fill the balloon with sterile saline, 50 cc at a time. Repeat up to 700 cc (14 strokes). 9Recommended full volume is up to 700 cc. minimum fill volume is 400cc. 9. 1After the last stroke pull back on plunger to create a vacuum in the valve to ensure closure. 10. Gently pull the tubing out and check valve for leakage.