An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Measurement of the Oxygen Isotope Anomaly (∆17O) of Tropospheric Ozone Using a Nitrite-Coated Filter William C. Vicars, S.K. Bhattacharya, Joseph Erbland, and Joël Savarino Laboratoire de Glaciologie et Géophysique de l’Environnement Université Joseph Fourier / CNRS, Grenoble, France

  2. The Oxygen Isotope “Anomaly” (Δ17O) of Ozone O3 tropo. δ17O = 0.52*δ18O Δ17O = δ17O - 0.52*δ18O Thiemens, 2006, Annu. Rev. Earth Planet Sci.

  3. Isotope Transfer During NOX Oxidation heterogeneous Higher Δ17O Transfer XO RH Nighttime N2O5 NO2 NO3 X O3 O3 XONO2 XO NO2 HNO3 NO Lower Δ17O Transfer Daytime RO2 OH time scale: minutes hours to days

  4. Δ17O(O3) in the Troposphere • Very few published observations of Δ17O for tropospheric ozone • Low [O3] (ppb level) • High [O2]/[O3] ratio • Initial studies depend on complicated experimental designs • Cryogenic trapping of O3 • Large amounts of liquid helium • Not applicable for field studies • Data may suffer from systematic errors • Low O3 collection efficiency • Interferences from atmospheric xenon Morin et al., 2007, Atmos. Chem. Phys. Krankowsky et al., 1995, Geophys. Res. Lett.

  5. Collection of O3 with Nitrite (NO2-) • Has been applied in many previous studies to determine [O3] using both passive and active sampler designs • Rapid oxidation, quickly goes to completion • Specific collection and measurement of O3 • Negligible interference from other atmospheric oxidants • Yields [O3] measurements in agreement with traditional measurement techniques (UV absorbance, etc.) NO2- + O3 —› NO3- + O2 Michalski and Bhattacharya, 2009, PNAS • Terminal oxygen atom transfer • Δ17O(O3)terminal = 3/2*Δ17O(O3)bulk • Δ17O(NO3-) = 1/3*Δ17O(O3)terminal • Δ17O(NO3-) = 1/2*Δ17O(O3)bulk Koutrakis et al., 1993, Anal. Chem. Pehnec et al., 2003, B. Environ. Contam. Tox..

  6. Collection of Ozone with a Nitrite-Coated Filter Filters (Whatman GF/A, 47mm) are coated with 1 mL NO2- solution and allowed to dry Prefilter for particle removal 15-20% loss of O3 to sampling apparatus • Overall collection efficiency (NO3- on filter/total throughput of O3) is 65-70% • Somewhat higher efficiencies for field collections (~ 75%) • Unexposed coated filters contribute approximately < 0.01 μmol (< 5%) 80-90% reaction yield for NO2- + O3 oxidation Teflon filter holder Flow regulated at ~ 3 L min-1

  7. Isotopic Analysis Bacterial Denitrifier Method He N2 100 nmol NO3- Extracted with MQ Sulfamic Acid Treatment for NO2- Removal Neutralized with NaOH (pH = 7) Incubation (12 hours) N2O NO3- NO3-N2O NO3- Concentration Analysis (CFA) Pseudomonas Aureofaciens N2O Calibration (USGS 34 & 35) O2 N2 O2 + N2 GC MS Au, 900°C

  8. Isotope Transfer Experiments Vacuum gauge ∆17O(O3)bulk N2 Pumps O2 He ∆17O(NO3-) Tesla coil Styrofoam balcony

  9. Isotope Transfer during Nitrite/Ozone Oxidation Δ17O(NO3-)= 1/2*Δ17O(O3)bulk

  10. Δ17O of Ambient Tropospheric Ozone Δ17O(O3)bulk = 18 - 29 ‰

  11. Conclusions Advantages • The nitrite-coated filter method offers several advantages over previous techniques: • Simple technologies, can be implemented nearly anywhere (e.g., alpine and polar environments, remote ocean, stratosphere) • Not susceptible to a systematically low collection efficiency or interferences from other compounds • Provides a direct isotopic measurement of the O-atom of O3 that is transferred during oxidation reactions, rather than a bulk isotopic measurement • Due to a relatively high flow rate and low sample size requirement, samples can be collected at a high frequency allowing for an increased time resolution

  12. Conclusions Limitations • Method may not be appropriate for O3 concentrations less than ~ 30 ppb at the studied flow rate • Filter blanks higher than 5-10% of NO3- formed via O3/NO2- reaction on filter • Increased sampling duration  lower time resolution • Increased flow rate  reduced ozone collection efficiency? • Large “matrix” blanks result in greater uncertainty compared to samples prepared in ultra-pure water • Increased corrections for delta values • Depends on blank/sample ratio

  13. Thank You!