11mm diameter. 33mm diameter. Oven with generator on the left middle (vessels in the center). 11mm diameter. 33mm diameter. E Field Distribution (Input Signal: 2.45GHz, 1 watt). 11mm. 33mm. Variation of E Field. 33mm diameter 11mm diameter. Configuration of MARS.
E N D
11mm diameter 33mm diameter Oven with generator on the left middle (vessels in the center)
11mm diameter 33mm diameter E Field Distribution(Input Signal: 2.45GHz, 1 watt)
11mm 33mm Variation of E Field
33mm diameter 11mm diameter Configuration of MARS
Relationship of Permittivity of DI-Water withTemperature
Measurements of the Precursor Solution at various Temperature (e’)
Measurements of the Precursor Solution at various Temperature (e’’)
e’ e’’ Extrapolation of permittivity of Precursor Solution at 175degC
Discussion Given other same conditions • E field at middle plane of the reactor is higher than the bottom • E field for precursor is lower than water (due to higher loss tangent of precursor) • For water, temperature goes up -> loss tangent goes down -> E field goes up • For precursor, temperature goes up -> loss tangent goes up -> E field goes down • For MARS-5 oven, E field maxima for 33mm reactor is lower than 11mm reactor
Maxima E field in MARSfor precursor solutionNormalized with sqrt(power)