Irradiation Tests Module 3B. L-wire 64. L. U. 485cm. 1. Position. Irradiate small area: 2x2 cm 2 , with 2 mCu 90 Sr, 2 cm distance (~200 nA). Many tests, see last slide for overview. NB: gas flow from Upper to Lower. Source profile. Gas flow. History Module 3. Construction:
E N D
Irradiation Tests Module 3B L-wire 64 L U 485cm 1 Position Irradiate small area: 2x2 cm2, with 2 mCu 90Sr, 2 cm distance (~200 nA) Many tests, see last slide for overview NB: gas flow from Upper to Lower Source profile Gas flow
History Module 3 • Construction: • 12 May 2004 • Flushed with Ar/CO2, 10 l/hr: • 9 June 2006 – 29 June 2004 • (Humidity of gas on input : Tdew=-76 0C, 17 June 2004) • 18 June 2004: Tdew=-22 0C • 22 June 2004: Tdew=-33 0C • 29 June 2004: Tdew=-35 0C • Total gas flow ~4800 liter • Medium Scale Test: flushed in C-frame, 76 l/hr (7 modules) • 7 April 2005 – End of May • ~7500 liter • Flushed with Ar/CO2, 10 l/hr (2.5 – 20): • 6 Jan 2006 – present • Date Time Tmitchell Tcorrected H20 content(ppm) • 23-1-2006 18h28 -38 -35.5 212.3 • 24-1-2006 9h26 -41 -37.89 163.4 • 24-1-2006 18h25 -41 -37.89 163.4 • 25-1-2006 -43.8 -41.00 115.6 • 26-1-2006 16h26 -48 -45.66 67.1 • ~5000 liter NB: Remarkably, the humidity of the module did not increase between 2004 and 2006 !
VI: Module 3B90Sr scan before and after 1st irradiation Module irradiated at 486 cm from lower end
VI: Module 3B90Sr scan before and after 1st irradiation After irradiation: Before irradiation: Ratio: Irr/Non-irr: Source position: 484cm
VI: Module 3B90Sr scan before and after 2nd irradiation Module irradiated at 462 cm from lower end
VI: Module 3B90Sr scan before and after 2nd irradiation After irradiation: Before irradiation: Ratio: Irr/Non-irr: Source position: 462cm
VI: Module 3B90Sr scan before and after 3rd irradiation L-side Module irradiated at 23 cm from lower end
VI: Module 3B90Sr scan before and after 3rd irradiation: L-side After irradiation: Before irradiation: Ratio: Irr/Non-irr: Source position: 20cm
VI: Module 3B 5th irradiation: upper side, around straw 32 60 hours Source profile (linear): Source profile (logar.):
VI: Module 3B90Sr scan before and after 5th irradiation 60 hours Module irradiated at 430 cm from lower end
VI: Module 3B90Sr scan before and after 5th irradiation: 60 hours After irradiation: Before irradiation: Ratio: Irr/Non-irr: Source position: 430cm
VI: Module 3B90Sr scan before and after 6th irradiation • By accident gas flow: off… • Irradiated • 4 hours with HV • 8 hours without HV • Used 2 sources • Source collimated circular at 442 cm • Source collimated with slit at 475 cm(used for 90Sr scans) Point collimated Slit collimated
VI: Module 3B90Sr scan before and after 6th irradiation: No gas flow! After 6th irradiation: ! Before 6th irradiation: Ratio: Irr/Non-irr: • Point collimated: 442 cm • effect is circular, as expected • Slit collimated: 475 cm • gain increased!? • No gas flow • Only 4 hrs at HV=1600V Source positions: 442cm and 475 cm
VI: Module 3B 7th irradiation: upper side, around straw 32 20 l/hr Source profile (linear): Source profile (logar.):
VI: Module 3B90Sr scan before and after 7th irradiation 20 l/hr Gas flow: 20 l/hr Source at 390 cm Irradiated 15 hours
VI: Module 3B90Sr scan before and after 7th irradiation: 20 l/hr After 7th irradiation: Before 7th irradiation: Ratio: Irr/Non-irr: • Gas flow: 20 l/hr Source positions: 390cm
VI: Module 3B 8th irradiation: upper side, around straw 32 2.5 l/hr Source profile (linear): Source profile (logar.):
VI: Module 3B90Sr scan before and after 8th irradiation 2.5 l/hr Gas flow: 2.5 l/hr Source at 375 cm Irradiated 15 hours
VI: Module 3B90Sr scan before and after 8th irradiation: 2.5 l/hr After 8th irradiation: Before 8th irradiation: Ratio: Irr/Non-irr: • Gas flow: 2.5 l/hr Source positions: 375 cm
VI: Module 3B90Sr scan before and after 9th irradiation 2 sources Gas flow: 10 l/hr 2 sources Does the 1st source eats the pollutions?? Irradiated 15 hours • Used 2 sources • Source collimated circular at 340 cm • Source collimated with slit at 360 cm(used for 90Sr scans)
VI: Module 3B90Sr scan before and after 9th irradiation: 2 sources After 9th irradiation: Before 9th irradiation: Ratio: Irr/Non-irr: • Two sources: Source positions: 340 and 460 cm
VI: Module 3B90Sr scan before and after 10th irradiation dryer?? Gas flow: 20 l/hr Source at 310 cm Irradiated 15 hours Humidity on output: Tdew= -43 oC <200 ppm Wire locator
VI: Module 3B90Sr scan before and after 10th irradiation: dryer?? After 8th irradiation: Before 8th irradiation: Ratio: Irr/Non-irr: • Gas flow: 20 l/hr • Humidity: <200 ppm Source positions: 310 cm
VI: Module 3B11th irradiation: upper side, around straw 32 dry: 20-30 ppm Source profile (linear): Source profile (logar.):
VI: Module 3B90Sr scan before and after 11th irradiation dry: 20-30 ppm Gas flow: 20 l/hr Source at 415 cm Irradiated 15 hours Humidity on output: Tdew= -51 oC 20-30 ppm
VI: Module 3B90Sr scan before and after 11th irradiation: dry: 20-30 ppm After 8th irradiation: Before 8th irradiation: Ratio: Irr/Non-irr: • Gas flow: 20 l/hr • Humidity: 20-30 ppm Source positions: 415 cm
Humidity 29 Jan 23 Jan
Conclusions Gas flow L-wire 64 L U 485cm 1 Position