An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Chapter 9 Basic MRI I Mark D. Herbst, MD, PhD

  2. Notice • This lecture contained many drawings on the whiteboard, so get these from one of the other students if you missed the class.

  3. Spatial characteristics of the MR image • The body is 3-dimensional • The images we look at are 2-dimensional • Pixels have 2 dimensions • Voxels have 3 dimensions • The third dimension is slice thickness • Slicing and dicing the body for MR imaging is like going from whole potato to chips to fries to home fries.

  4. K-space • Signals from the patient are stored in the computer in a place called “k-space” and when it is filled, we use Fourier Transform to transform the information from k-space to “image space,” that is, an image.

  5. Gradients • We use gradients in 3 directions to get images • Slice selection gradient • Phase encoding gradient • Frequency encoding gradient • All these are orthogonal (at right angles) to each other

  6. Slice select gradient • Selective excitation • Apply a gradient in the slice selection (CC for axial slices, RL for sagittal slices, AP for coronal slices) • Two determinants of slice thickness • Strength of gradient • Transmit bandwidth • Remember the previous discussion about multislice imaging? We use the gradient to select the slice we want to excite with RF.

  7. SELECTIVE EXCITATION 2D SLICE ACQUISITION

  8. Slice Selection • 3D volume acquisition • Uses two phase encoding gradient pulses, one in the usual phase encoding way, to be discussed below, and the other in the slice select direction. We need to collect a full set of echoes in a separate k-space for each slice, so we have to multiply the time for one slice by the number of slices to get total time. This is why we only use very fast GE methods to get 3D volume acquisitions.

  9. Phase Encoding Gradient • Applied after the excitation pulse (90 degree or alpha pulse) • Changed to different level for each echo collected • If 256 x 256 image, need 256 echoes, therefore 256 different phase encoding pulses are applied, each at a different strength. • Acquisition of each of these echoes takes one TR if conventional spin echo, less if fast spin echo or EPI

  10. Frequency Encoding Gradient • Applied during the echo, can be seen on some pulse diagrams to be applied also during the FID. • Changes the frequency of precession of the net mag vector in each row of voxels in the frequency encoding direction

  11. MULTISLICE IMAGING

  12. 3D VOLUME ACQUIASITION

  13. SUMMARY