1 / 56

Quality Management -Standards

Quality Management -Standards. Joan Kelley 28th Nov 2002. Background . Employed UK MoD Assessor in QC system Transferred to UKAS (then NATLAS) on inception Testing lab UKAS (now ISO 17025) Accredited since 1984. Accreditation Bodies. UKAS-UK Accreditation Service

ollie
Download Presentation

Quality Management -Standards

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Quality Management -Standards Joan Kelley 28th Nov 2002

  2. Background • Employed UK MoD Assessor in QC system • Transferred to UKAS (then NATLAS) on inception • Testing lab UKAS (now ISO 17025) Accredited since 1984

  3. Accreditation Bodies • UKAS-UK Accreditation Service • NAB-National Accreditation Board,Ireland • BELCERT and BELTEST, Belgium • Comite Francais d’Accreditation • Deutscher Akkreditierungsrat • Raad voor Accreditatie

  4. Cooperative Bodies • European co-operation for Accreditation (EA) • EA 4/10 Accreditation for Microbiological Laboratories, July 2002 www.european- accreditation.org • International Laboratory Accreditation Cooperation (ILAC) www.ilac.org

  5. Context • Regulations = EN45001, European Standard (others GLP, ISO 9000 etc) • But since early 2000…………. • ISO/IEC 17025 came into force • General Requirements for the Competence of Testing and Calibration Laboratories

  6. Current Standards • ISO 17025- Testing and ‘methods’ • ISO 9000 series-production and management systems • National and international GLP standards-Medical research

  7. Change in Philosophy • Process driven approach • -does the system work, rather than -does it contain specific features • More flexibility for the lab.

  8. Format • Most technical standards have two main sections ……………... • Management requirements • Technical requirements

  9. Management • If part of larger company, key personnel for lab must be defined • Lab must have tech management but can be a team-individual not necessary

  10. Documents • Quality Manual • Methods (Technical Operating Procedures) • Standard Operating Procedures

  11. Quality System • Management review • Audit programme • Document control • Control of complaints and anomolies • Clear job descriptions • Training records • etc etc etc

  12. Contract Review • Lab must have procedures which ensure that…… • methods to be used are defined,documented and understood • lab assured it has capability and resources to carry out work • appropriate methods are used to meet clients needs

  13. Preventive Actions • Pro-active look for opportunities to improve not just corrective actions • Can come from…….. • Internal audits,review of AQC data,staff suggestions,management review

  14. Technical requirements ISO 17025 • Increased emphasis on uncertainty of measurement, always a minefield for microbiol. • In house calibrations given emphasis • Future training to be addressed • Opinions and interpretations can be accredited

  15. Application of Standards to Microbiology labs Some Guidelines

  16. Quality Control • Internal AQC to monitor individual performances,cross checks between microbiologists,duplicate preps. of single samples by different people. • QC of batches of media and diluents ( more later)

  17. Inter-laboratory comparisons • Lab should participate in external AQC schemes when available and appropriate-Collections to establish system?? • eg PHLS Water and Food QC, FAPAS,NEQAS, SMART etc etc • Must be seen to act on results-don’t just file and forget

  18. Staff • Senior member of staff must have appropriate qualifications • Alternatives may be acceptable with extensive experience

  19. Equipment • Records for all essential equipment • Cleaning programmes for appropriate equipment together with documented records • eg water baths, incubators, fridges, freezers

  20. Reagents • Generally ANALAR • Water -deionised or distilled • Regents labelled, identity, opening and expiry dates,conc, storage requirements • Includes all media and diluents

  21. Calibration • Must have a co-ordinated, recorded programme encompassing all lab activities • Measurements of significance to methods must be traceable to national/international standards

  22. Temperature Measuring Devices • Fridges, incubators,water baths,drying ovens, autoclaves • Where temp needs to be better than +/- 1C calibrated thermometer in appropriate range essential • Otherwise BS 593 may be adequate or calibrated ref used to check working thermometers

  23. Cont……. • Ref thermometers, calibrated every 5 years • Annual ice point check • Thermocouples-traceable to national standards • Working thermocouples can be calibrated in house against ref standards

  24. Weights and Balances • Must be calibrated at least annually depending on use and environment • In house or by Accredited service • Daily drift checks

  25. Volume • Pipettes etc must be calibrated and traceable to national standards • Variable volume automatic pipettes discouraged • Disposable pipettes should be batch checked even if supplied from ISO manufacturer

  26. pH • Checked daily or before use • Buffers from quality supplier • Buffers prepared to standard and stored appropriately with an expiry date

  27. Humidity • Must be calibrated and traceable • Easiest to send to experts!

  28. Autoclaves • On purchase-multi point validation for all typical loads and put results to use! • Traceable to national standards • Thereafter calibration of timer and temperature measuring equipment annually • Again act on the results

  29. Cont • Records of all loads and runs must be taken and kept • In addition load monitors must be used with each load

  30. Air velocity • If used in eg safety cabinets (ie not done under contract) must be checked for accuracy at regular intervals.

  31. Method Validation • Methods developed in house, modifications and applications of standard methods must be validated • But also need to prove you can get the correct results even in standard tests

  32. Essential Features • Limits of detection • Selectivity • Precision • Reproducibility • Procedures depend on requirements/qualitative or quantitative/new or existing method? • Reference materials

  33. Qualitative Results (Presence or Absence) • Limit of detection • Selectivity determined to ensure interference not occurring-use a range of micro-organisms

  34. Quantitative methods • Sensitivity • Selectivity • Precision • Expression of uncertainty

  35. Environment • Restricted access • Lab coats • ‘Potential for contamination minimised’ ie non-absorbent surfaces,easy clean, blinds,no plants,no personal possessions,minimal paperwork • Documented cleaning programme

  36. Cont • Microbiological monitoring • Records kept • Action and alert limits • Documented procedures

  37. Sample/Deposit Handling and Identification • System can be simple as notebook or complex LIMS system • Adequate facilities for storage and segregation essential (fridges,freezers) • Samples must be stored and examined under conditions which minimise changes in the numbers or viability of micro-organisms

  38. Disposal of Contaminated Waste • All waste should be autoclaved before leaving the laboratory • If disposal of waste is subcontracted control measures must be in place • However if pathogens involved also follow ACDP guidelines and WHO Manuals

  39. Organisms in Quality Systems Use, preservation and traceability

  40. Organisms in Quality Systems • Challenge testing • Method validation • Media QC • Analytical Quality Control (AQC) • Proficiency Testing

  41. Guidance for ISO 17025 • States two fundamental requirements:- ‘…shall maintain records of checks on batch to batch variation of media and other reagents by using working standards of known micro-organisms’ ‘...shall hold reference cultures obtained from a recognised national collection in order to demonstrate traceability’

  42. Common Confusions • Definitions of reference/stock/working cultures • Can you subculture and if so how many times • Are the newer delivery systems acceptable

  43. Reference Organisms-1 • Reference cultures (i.e. from a collection) may be subcultured once to provide reference stocks • Reference stocks shall be maintained ‘...by techniques which maintain the desired characteristics of the strains’ • Examples-freeze drying, LN,deep freeze methods

  44. Reference Organisms-2 • Reference stocks are used to prepare working stocks for routine use • Working stocks can be subcultured if:- • required by a standard method • documentary evidence exists to show changes do not occur over a defined number of subcultures, and this has been validated

  45. Reference Organisms-3 • Once thawed or otherwise reconstituted reference stocks must not be refrozen or re-used • Working stocks must not be sub-cultured to replace reference stocks

  46. Things often forgotten • Document traceability of strains including date purchased,dates sub-cultured dates beads removed etc.. • Continuous monitoring of freezer temp. or at least max/min. • Place shelf life on all stocks, reference and working

  47. ‘New’ Delivery Systems • E.g Culti loops, Quanticult etc • All previous requirements still apply • Proof of quality, traceability etc falls to the laboratory • Treat like any other goods and service supply but with extra ‘calibration’ requirements

  48. Questions to Ask • ISO 9000.. Company? • But what for -manufacture,supply? Who actually makes the products? • Traceability to national collection? • How many passages? • What preservation techniques are used?

  49. Documents Required • Validated methods and QC data • Certificate of traceability to National Collection number • Statement of uncertainty i.e. if quantitative, number of cfu +/- but also details of how this is calculated, cumulative errors etc.

  50. Finally • Whatever route chosen, the onus falls on the laboratory to demonstrate quality is being maintained.

More Related