1 / 37

IEEE NSS – MIC, Rome 21st October 2004

IEEE NSS – MIC, Rome 21st October 2004. High resolution X-ray analysis of a proximal human femur with synchrotron radiation and an innovative linear detector. M.Bettuzzi , R. Brancaccio, F.Casali, S. Cornacchia, N.Lanconelli, A.Miceli, M. P.Morigi, A. Pasini, D. Romani, A.Rossi

paulaz
Download Presentation

IEEE NSS – MIC, Rome 21st October 2004

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. IEEE NSS – MIC, Rome 21st October 2004 High resolution X-ray analysis of a proximal human femur with synchrotron radiation and an innovative linear detector M.Bettuzzi, R. Brancaccio, F.Casali, S. Cornacchia, N.Lanconelli, A.Miceli, M. P.Morigi, A. Pasini, D. Romani, A.Rossi Department of Physics, University of Bologna and INFN, Section of Bologna

  2. Ostheoporosis Project The experiment was carried out in the framework of a project, powered by the University of Bologna, concerning the characterization of the human bone tissue by means of physical techniques

  3. micro-CT slice Good results were already obtained with the 3D micro-CT technique on small bone samples With our micro-CT system we obtained images with a pixel size from 20 to 30 microns (less than 20 microns with high magnification µ–focus source) A good agreement with results achieved by other systems (i.e. Skyscan) was found

  4. However, such a technique allows to investigate only small samples (1-2 cm maximum size), that means you have to physically cut the bone into slices and make out a carrot of one of them The aim of the present experiment was to verify the possibility of performing the same kind of analysis over a complete human bone (i.e. a femur) without altering its structure by any mechanical treatment Looking forward, this should be a first step toward the hard task of building an X-ray CT system for the structural bone analysis “in vivo”

  5. Trabeculae size in a human bone vary from 0.1 to 0.3 mm while spacing from 0.2 to 2 mm We had to build a new detector with the proper spatial resolution and a sufficiently wide field of view The synchrotron radiation source was choose because it provides a high flux of mono-energetic photons Thus it is the best quality source for this experiment, though it’s not the most suitable one for routine analysis

  6. A new linear intensified detector was designed at the University of Bologna( EU patented ) • The detector is based on two main components: • a digital intensified CCD camera (the EBCCD) • a special coherent fiber optics adapter

  7. The Electron Bombarded CCD camera Produced by Geosphaera (Moscow)

  8. Scheme of the EBCCD tube

  9. The fiber-optics guide permits the conversion of the geometry from linear to rectangular format From 1024x512 to 5600x50 effective pixels Realization Concept Produced by the Vavilov Institute (S.Petersburg)

  10. Detector assembly scheme Scintillating screen stripeGd2O2S:Tb Fiber optics light guide129×1.45 mm2 to 18.4×10.6 mm2Multialkali input window 24.5 mm Back-thinned frame transfer CCD1024×512Intensifier tube voltage6 kV Image size (multislice)5600×50pixel

  11. Mounted detector

  12. SPLINE interpolation of missing data into the spacings between neighbouring FO ribbons 129 m 66 m

  13. Scheme of the CT system (microfocus source) Fiber optics fan adapter Microfocus X-ray tube EBCCD Controlled axis

  14. Radiograph of a calibrated bar-pattern direct measurement of the spatial resolution 5 lp/mm (100 µm)

  15. Modulation Transfer Functionmicrofocus X-ray tube measurements (focal spot 5 µm) Spatial resolution of the detector (at 5% measured MTF value) 60 µm

  16. Noise and DQE

  17. Experiments at Elettra SRFin collaboration with Rizzoli Institute, Bologna August 2003 - April 2004

  18. ELETTRA Syrmep Beamline

  19. Beam energy 34 keV max Beam size width:120 mm (our detector is 130 mm long) height:4 mm (our detector is 1.5 mm high)

  20. Experimental set-up at the SYRMEP beamline beam sample proximal femur (pig) detector

  21. position of the CT slices Each CT section is made of about 35 slices, one pixel thick (23 µm) 1 Original projections size 5600 x 35 pixel 2 Exposure Time: 50 ms Number of averaged frames : 16 Total exposure time for 500 projections: 400 s Total CT time: 8000 s (2 hours e 13 minutes) • Proximal femur of a pig • Maximum size 7,41 x 5,42 cm2 • CT parameters • Energy 34 keV • Number of projections 500 • Angle 180° • Voxel side 23 m

  22. 1

  23. 2

  24. The trabeculae structure becomes visible, thought, spatial resolution is not enough at this level to define it properly

  25. The use of a high resolution linear detector with synchrotron light allowed the definition of the trabecular structure in a CT slice of a complete proximal femur bone of a pig The image quality was too low and it was not suitable for a proper segmentation and for the calculation of the isomorphometric parameters of the bone structure

  26. Improvements • better alignment of the detector • collection of a large number of projection • recalibration and accurate reconstruction • human femur analysis

  27. Human femur CT • Parameters of the CT • 1400 projections over 180 degrees • 16 frame averaged for each projection • 25 ms exposure time for each frame • 2 hours 15 minutes total scanning time

  28. Radiography of a human femur

  29. Micro-CT slice

  30. A high resolution computed tomography of a complete human femur was performed The high quality of the images obtained after the CT reconstruction allows the calculation of the histomorphometric parameters of the bone Thus, it is possible to study the trabeculae structure of the complete human femur

  31. Further steps: • Replicate the experiment with a microfocus X-ray tube • Optimize the source-detector system • Optimize the number of projections for a good image quality with a reduced dose

  32. THANK YOU FOR YOUR ATTENTION

More Related