FDR Presentation Personal Health Monitoring Device. Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00. Overview. Sensors ECG Blood Pressure Temperature Signal Conditioning Filters, Amplifiers,Logic PIC Microcontroller
E N D
FDR PresentationPersonal Health Monitoring Device Bassam Noaman Dina El-Eissa Mentor: Prof Zhenyu Guo Instructor: Prof J Silverman 4/12/00
Overview • Sensors • ECG • Blood Pressure • Temperature • Signal Conditioning • Filters, Amplifiers,Logic • PIC Microcontroller • program interface • Palm pilot • PC
Specification • Power Supply: 9.0 VDC Battery • Operating time 2.8 hours(from Standard 9.0 VDC Duracell battery) • Inputs • Body Temperature Sensor • Detect Temperature in degree Celsius • Range between 25 C – 50.5 C • Blood Pressure • Range between 0 – 5 Psi(259mmHg) • Accuracy 0.125 @ 25 C • Manual cuff inflation • Maximum systolic pressure 250 mmHg • ECG • Bandwidth 0.3Hz – 150 Hz • Detect Heart Rate in bpm • One lead ECG
Specification • Monitor Performance Specification • Palm Pilot screen record and plays the patient’s data file on the screen • PC screen displays the patient data file • Recorded body temperature • Systolic and diastolic blood pressure • ECG waveform • Heart rate • Email this data file • Interfaces • Cables to interface the human body to the device • Serial interface from the device to the Palm • Serial interface from the palm to PC
Specification • Indicators • LED power status • LED blood pressure visual detector • Memory Storage • 1 MB of Palm is allocated for health monitoring • 55.6 minutes of display range
Output voltage of the signal conditioner of the body temperature sensor
Blood pressure measurement • Blood pressure: • is defined as the intra-arterial pressure generated by the pulsatic flow of blood through an artery. • The blood pressure is indicated from two numbers. • Systolic:represents the pressure while the heart is beating. maximum value for blood pressure • Diastolic: represents the pressure when the heart is resting between beats.
Demonstration • Measure current temperature and blood pressure and display it on the palm. • Display temperature,blood pressure and ECG wave in Hyper terminal, then take the data and save it in a file. Finally display it on java applet.
Conclusion • The Palm pilot is not only an organizer, but also a device to monitor data(Blood pressure, Temperature)