Chapter 18 Quality Control. Quality Assurance is a JCAHO required 10 step PEOPLE process Quality Control Is an imaging instrumentation and equipment process. QC – 3 step process. Acceptance testing Routine performance monitoring Maintenance. Radiography QC. Filtration - Annually
E N D
Chapter 18 Quality Control Quality Assurance is a JCAHO required 10 step PEOPLE process Quality Control Is an imaging instrumentation and equipment process
QC – 3 step process • Acceptance testing • Routine performance monitoring • Maintenance
Radiography QC • Filtration - Annually • Collimation – Semiannually (+/- 2% SID) • F.S. size – Annually (+/- 50%) FREQUENTLY MISSED PARAMETER! • kVp calibration – Annually ( +/- 10%) • Exp. Timer accuracy – Annually (based on ms used – more leeway with longer exposures) • Linearity – Annually (+/- 10%) of same mAs with differing mA/times • Exp. Reproducibility – Same contr./Dens., etc when exposures are repeated later on
Focal Spot Size Check Tools • Pin hole camera (somewhat hard to use) • Star pattern (down to .3 mm sizes) • Slit camera – preferred method
Other RAD QC • Filtration • Collimation (PBL) • kVp/Timer • Exposure Linearity (same mAs with varied mA and times) • Intensifying screens • Protective apparel • Film illuminators
Fluoro QC • Exposure rate • Spot-fil Exposures • Automatic Exposure System (ABS – AEC)
Tomo QC • Tube travel/ARC • Fulcrum level • Pt. Exposure
Processor QC • See table 18-8 (p. 313) for QC for RAD Processor items/intervals • Processor Maintenance • Scheduled maintenance • Preventative maintenance • Nonscheduled maintenance
Processor Monitoring • Temperatures • Replenishment • Sensitometry/Densitometry