E N D
1. CovidienPuritan Bennett 840
2. Objectives Overview
840 Disk
Modes
Special Features
3. Overview Manufactured by COVIDIEN
Massachusetts
PB 840
Headquartered in Boulder, Colorado
Patients
Neo
Peds
Adult
4. 840 DISK
5. Modes Assist/Control
SIMV
Spontaneous
BiLevel
PAV+
NIV
6. Modes Assist/Control (A/C)
A/C VC
Machine [time] or Patient [flow/press] triggered
Flow Limited
Volume Cycled
7. Modes Assist/Control (A/C)
A/C VC+
Pressure Regulated Volume Control [PRVC]
Machine or Patient Triggered
Pressure Limited
Volume Targeted
Time Cycled
Delivered Pressure cannot exceed Set upper Pressure limit Minus 5 son
8. Modes Assist/Control (A/C)
A/C PC
Machine or Patient Triggered
Pressure Limited [targeted]
Time Cycled
9. Modes SIMV
Synchronized Intermittent Mandatory Ventilation
Mandatory breaths
PC or VC
Spontaneous breaths
PS & PEEP
10. Modes SPONTaneous
PEEP
Baseline of 1.5son PS if PS setting is zero
PS
VS
Volume targeted Pressure breath
TC (ATC)
PAV
11. Special Feature Active Exhalation Valve
Allows for Patient to receive extra flow when necessary
when pt would normally override the demand flow of the ventilator in VC mode
Allows for venting of excess circuit pressure
Such as would happen during a cough
During Pressure breaths only