An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Recent advances in Magnetic Resonance Imaging Peter Fransson MR Research Center, Cognitive Neurophysiology Dept. of Clinical Neuroscience, Karolinska Institute

  2. Overview • Brief recap: MRI Physics • Image acquisition speed is of essence… • Functional Magnetic Resonance Imaging • Diffusion tensor MRI, MR tractography • Parallel Magnetic Resonance Imaging • Outlook

  3. Physical principles of NMR (very briefly) Proton spin angular momentum: Magnetic dipole moment: External magnetic field: Energy levels are split (Zeeman effect): E , anti-parallel spin E 0 , parallel spin B0

  4. Physical principles of NMR (very briefly) Motion of spins in an external magnetic field: NMR experiment: static field and radiofrequency (RF) field): In a rotating frame of reference with the angular frequency B0 M0 M y x B1

  5. Spatial localization in MRI • Let the magnetic field vary in x, y and z-space. Gx x

  6. MR IMAGING IN 1973 P.C. Lauterbur, Nature, 242:190-191, 1973

  7. CONVERTING FREQUENCIES INTO SPATIAL LOCATIONS ”k-space” ”reconstructed image” 2D -FFT 0 1 2 2 0 1

  8. TRAVEL IN ”K-SPACE” WITH THE SPIN ECHO SEQUENCE 180 900 3 1 2 3 4 4 1 • The gradients permits sampling of points in k-space. 2 • Each echo gives us one line in k-space. • Scan time: TR x N_phase ECHO

  9. Conventional gradient echo image acquisition RF SIGNAL N excitations / image N times 0.1-0.2 slices / second

  10. Echo Planar Imaging sequence 90o RF TEeff SIGNAL

  11. EPI image acquisition SIGNAL

  12. EPI and T2*-sensitivity 64 echoes or more are acquired per image EPI is strongly sensitive to variabilities in the magnetic field (T2*)

  13. (Gradient) Echo planar imaging T2*-weighted image contrast TR/TE/flip = 3000ms/40ms/90deg 3.4 x 3.4 x 4 mm3, 30 slices

  14. 1.5T GE Twinspeed Excite MR scanner

  15. 1.5T GE Twinspeed Excite MR scanner

  16. 1.5 Tesla Excite Twinspeed GE MR scanner - console

  17. Functional Magnetic Resonance Imaging

  18. Hypothesis on brain function Paradigm design Physiological and metabolic responses Signal changes in the MR image Post processing / statistical analysis Visualisation / Activation maps

  19. HEMOGLOBIN • 4 subunits, each carrying a heme (red) • one iron atom (Fe2+ ) is carried by each heme • to each heme an oxygen molecule can be attached • with oxygen : oxy-hemoglobin • without oxygen : deoxy-hemoglobin

  20. Oxy-hemoglobin Slightly diamagnetic, same as the surrounding tissue Deoxy-hemoglobin Paramagnetic, susceptibility difference: ppm

  21. The BOLD effect - theoretically Magnetic field distortions: a Outside ”vessel”: r Inside ”vessel”:

  22. Oxygen saturation and magnetic susceptibility Bandettini & Wong, Intern. J. of Imag. Syst. And Techn. 6:133, (1995)

  23. Historical background (II): Initial observations • Ogawa (1990): • Gradient echo imaging (T2*-sensitivity) of mouse brain at 7T • Changed inhalation gas from 100% to 20% oxygen (room air) • Observed a signal decrease in the vicinity of vessels (reversable) • No signal change in corresponding spin echo images (T2-sensitivity) Conclusion: Signal decrease is due to increased magnetic field inhomogeneities caused by an increase in the concentration of paramagnetic deoxy-Hb. Cerebral blood oxygenation (CBO) Signal change in T2*-sensitized MR images BOLD - Blood Oxygenation Level Dependent

  24. Hemodynamic response function (hrf)

  25. rCBF and rCMRO2 mismatch CBF BOLD effect CMRO2 Neuronal activity

  26. fMRI - Blocked design OFF: ON: OFF ON OFF ON OFF ON OFF ON OFF t 0s 30s 60s 90s 120s 150s 180s 210s 240s 270s Continuous EPI image acquisition

  27. fMRI – Blocked design T2*-weightedimage Activation map, p<0.001 • 2T, blipped EPI: TR/TE/flip = 400ms / 54ms / 30 degrees • 10s reversing checkerboard / 20s fixation cross, 6 repetitions • Anatomy: RF-spoiled gradient eko (FLASH) ,TR/TE/flip = 70ms / 6ms / 60 deg.

  28. Blocked fMRI signal intensity time course

  29. Cavernoma

  30. Self-paced fingertapping with left hand

  31. 1.5T GE Twinspeed Excite MR scanner – fMRI set up

  32. MR compatible user feed-back ”glove”

  33. 1.5T GE Twinspeed Excite MR scanner – fmri running

  34. fMRI Summary • fMRI does not directly measure neuronal activity - it relies on vascular and metabolic correlates of changes in the neuronal work load. • Results are dependent on the design of the experiment and the MR parameter settings. • Large intersubject variability in the resulting activation maps • Only relative changes in brain activity can be measured with BOLD fMRI.

  35. Diffusion Tensor Magnetic Resonance Imaging

  36. A stationary molecule in the presence of diffusion gradients f ω < ω0 ω > ω0 180

  37. A moving spin in the presence of diffusion gradients f 180

  38. MR signal intensity in spin echo sequences decreases exponentially: D = diffusion coefficient The diffusion coefficient can be determined by measuring the spin-echo amplitude as a function of gradient strength

  39. Introduce diffusion gradients in the imaging sequence G Skiv- sel. d D Freq. Enc. Phase Enc. 180° 90 RF

  40. b vs. signal intensity log (signal) T2-weighting DWI = diffusion weighted image S0 S1 b b0 b1

  41. The ADC image • ADC = Apparent Diffusion Coefficient • ADC = the slope • CSF  2000 m2/s • Brain  700 m2/s

  42. A clinical example of diffusion-weighted MRI: acute stroke Var är infarkten? ADC T2 DWI

  43. Measurement of the Diffusion Tensor DT-MRI Gray matter CSF White matter

  44. Spatial orientation of the diffusion tensor, red=L-R, green=S-I, blue=A-P FA-map (Fractional Anisotrophy)

  45. MR Tractography – Fiber tracking • Following the direction of the eigenvector corresponding to the largest eigenvalue through the imaged brain volume • e.g. to see if/which two brain regions are connected • several fibres in e.g. the brain stem can be identified • Requires high-resolution & high SNR • Scan times minimum ~20 minutes with SS-EPI • Several methods for improve the results based on the still too noisy data • FACT, Spaghetti model, Continuous tensor field Courtesy of Susumu Mori, Johns Hopkins, Baltimore

  46. Parallelimaging in MRI

  47. Acquisition of MR image Sampling of k-space Imaging scan time is determined by the time it takes to sample k-space. Scan time can be reduced by doing tricks in k-space such as • Fractional NEX sampling of k-space (ky range reduced) • Fractional echo sampling of k-space (kx range reduced) But speed in k-space is crucially determined by gradient strength: Only one point in k-space can be sampled at a time!

  48. The measured signal will depend on the distance to the object being imaged. d Receiver coil Object (Biot-Savarts law)

  49. We can receive MR signals from several coils in parallel... An image from each coil can be generated. The signal intensity in each voxel will depend on the spatial distance of that voxel and the coil.