Detection of Ziram by LC/APCI/MS Hiroshi MORIWAKI Osaka City Institute of Public Health &

Detection of Ziram by LC/APCI/MS Hiroshi MORIWAKI Osaka City Institute of Public Health & Environmental Sciences. Ziram. use. pesticide antifouling agent. toxicity. endocrine disruptor mutagen carcinogen. ((CH 3 ) 2 NCS.S) 2 Zn M.W. 305.8 m.p. 240 Log Pow 1.09. Analysis of Ziram.

rafiki
Download Presentation

Detection of Ziram by LC/APCI/MS Hiroshi MORIWAKI Osaka City Institute of Public Health &

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Detection of Ziram by LC/APCI/MS Hiroshi MORIWAKI Osaka City Institute of Public Health & Environmental Sciences

  2. Ziram use pesticide antifouling agent toxicity endocrine disruptor mutagen carcinogen ((CH3)2NCS.S)2Zn M.W. 305.8 m.p. 240 Log Pow 1.09

  3. Analysis of Ziram Direct detection by GC/MS → × Conventional method Decomposition of the ligand to CS2 × Chemical speciation of dithiocarbamate pesticides LC/MS ( Direct detection of ziram )

  4. Ion mode Ionization Positive ion mode Negative ion mode ESI/MS × × APCI/MS × Corona current<0.6mA × ○ Corona current>0.6mA

  5. Mass spectrum B) A)

  6. Corona current vs. peak intensity

  7. Cone voltage vs. peak intensity

  8. Conditions of APCI/MS

  9. Column separation

  10. Mass chromatogram

  11. Calibration curve

  12. Summary ・ Direct detection of Ziram by LC/APCI/MS ・ Corona current > 0.6mA ・ LC separation: GPC column ・ Instrumental detection limit: 2mg/L Next step ・ Development of the pretreatment ・ Use of LC/MS/MS ref. H. Moriwaki, Y. Yamaguchi and M. Fukushima: Rapid Commun. Mass Spectro., in press (2001).