An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Analytical & Bioanalytical Techniques Joshua Hicks Catalent Pharma Solutions September 2, 2015

  2. NMR quantification vs. other methods NMR • For automated, internal-standard-free, absolute quantization… • NMR is accurate, precise and versatile • Superior to • CLND • ELSD • Weighing CLND ELSD • Analytical Chemistry (2005), 77(14), 4354-4365

  3. Quantificaiton with Internal Standard (IS) • X mg of Compound • X mg of IS • Purity of IS Concentration of Compound? Purity of Compound?

  4. Using PULCON J Am Chem Soc. 2006 Mar 1;128(8):2571-6. Reference Standard for PULCON NMR of Sample Calibrated to a sample of known concentration, known number of protons 10 mM & P1 Calibration NS TE Integrals # Protons P1 Calibration NS TE Integrals # Protons Concentration of Unknown

  5. Accuracy of Analysis? • Considerations for Reducing Error in Quantification • Sample Prep • Parameter Set • Field Strength (less error at higher fields) • Probe • Instrument stability • Method of integration (Region or Linefit) • Room stability • SOPs • Calibrant Region (pure) Linefit (overlap)

  6. A Tale of Two Probes Single sample, same instrument – does configuration matter?

  7. Quantification of Compounds in Crude Extract * * * * 32.8mM

  8. Youden Ruggedness’ Trial • 600 MHz vs. 400 MHz • NMR Tube Quality • Different concentration • Different gravimetric balance • Different pulse sequence • Different days Mesh Size (Also EtOH:Leaf) Chlorogenic Acid BBE Delta ~ 2% 600 400 7

  9. Benefits of High Quality Data: L-Carvone Red = double presat Blue = Wetdc L-Carvone Water DMSO 100 Hz

  10. QC/QA System • Non-NMR user • Simple interface • Lockdown • Automated • Suitability test • Acqu/Proc • Identification • Quantification • Report

  11. Towards Automatic Analysis • Quantification • Proton, fluorine, and phosphorus • Identification of residual solvents • Assist with identification of ‘unknown’ signals • System: • PDF report • Concentration verification (based on spectra and submitted compound) • Verification by either: • Proton only • Or proton+carbon • Or proton+hsqc • Stop light evaluation of atom assignment

  12. discover more. CATALENT PHARMA SOLUTIONS14 SCHOOLHOUSE ROADSOMERSET, NJ 08873 + 1 866 720 3148 www.catalent.com