Reliability of point-of-care testing for glucose measurement in critically ill adults

Reliability of point-of-care testing for glucose measurement in critically ill adults. Abstract. Objective: To determine the accuracy and clinical impact of three common methods of bedside point-of-care testing for glucose measurements in critically ill patients receiving insulin infusions.

Download Presentation

Reliability of point-of-care testing for glucose measurement in critically ill adults

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Reliability of point-of-care testing for glucose measurement in critically ill adults

  2. Abstract • Objective: To determine the accuracy and clinical impact of three common methods of bedside point-of-care testing for glucose measurements in critically ill patients receiving insulin infusions. • Design: Prospective observational study. • Setting: A 21-bed mixed medical/surgical intensive care unit of a tertiary care teaching hospital. • Patients: Thirty consecutive critically ill patients who were vasopressor-dependent (n = 10), had significant peripheral edema (n = 10), or were admitted following major surgery (n = 10).

  3. Measurements: Findings from three different methods of glucose measurement were compared with central laboratory measurements: • (1) glucose meter analysis of capillary blood (fingerstick); • (2) glucose meter analysis of arterial blood • (3) blood gas/chemistry analysis of arterial blood.

  4. Results: Clinical agreement with the central laboratory was significantly better with arterialblood analysis (69.9% and 76.5% for glucose meter and blood gas/chemistry analysis, respectively) than with capillary blood analysis (56.8%; p = .039 and .001, respectively). • During hypoglycemia, clinical agreement was only 26.3% with capillary blood analysis and 55.6% and 64.9% for glucose meter and blood gas/chemistry analysis of arterial blood (p = .010 and <.001, respectively). • Glucose meter analysis of both arterial and capillary blood tended to provide higher glucose values, whereas blood gas/chemistry analysis of arterial blood tended to yield lower glucose values.

  5. Agreement vs Disagreement • Agreement: the blood glucose measurements under study generated an observed measurement that fell into the same category (i.e., resulting in the same intervention) as that of the hospital’s central laboratory. (fingerstick:4.1~ lab:4.3) • Disagreement: If the two results fell into different categories and would have resulted in different clinical interventions (fingerstick:4.1mmol/L~ lab:5.0mmol/L)

  6. No statistically significant differences between measurement methods for vasopressor-dependent and edematous patients were observed, although there was a trend toward arterial blood providing more accurate results. • Agreement was significantly better for arterial blood measurements by blood gas/chemistry analysis (84.2%) than for capillary blood (fingersticks) (53.8%; p = .0045) among postsurgical patients.

  7. When the three patient groups were combined ----- • hypoglycemic readings: arterial samples (55.6% agreement for glucose meter analysis; 64.9% agreement for blood gas/chemistry analysis) appeared significantly better than agreement for capillary samples (26.3%) • nonhypoglycemic readings:no significant differences between tests were observed • hypoglycemic and nonhypoglycemic readings were combined: arterial sample agreement (69.9% for glucose meter analysis; 76.5% for blood gas/chemistry analysis) was superior to capillary sample agreement (56.8%) for all patients studied.

  8. For all patients, when compared with the reference standard : • capillary sample measurements were 8.8 ± 17.8% higher • glucose meter analysis measurements of arterial blood were 3.6 ± 15.1% higher • blood gas/chemistry analysis measurements were 2.5 ± 6.7% lower

  9. Modified error-grid analysis • display the combined results of individual patient groups for each specimen studied. • blood gas/chemistry analysis of arterial blood: Ninety-nine percent of values fell within the target range • glucose meter analysis of arterial and capillary: 88% and 73% of values fell within the target range. • During hypoglycemia: • glucose meter analysis of arterial and capillary blood overestimated glucose values 5% and 9% • blood gas/chemistry analysis of arterial blood did not overestimate.

  10. Compared with the suggested standard provided by the NCCLS • (NCCLS) National Committee for Clinical Laboratory • correlation coefficients measuring the associations between the reference method and each of the three test methods were calculated • The NCCLS suggests that a correlation above 0.9751 is indicative of a method equivalent to the laboratory standard. • blood gas/chemistry analysis met this standards overall (r = .9902), whereas glucose meter analysis of both capillary blood (r = .9516) and arterial blood (r = .9531) fell short of this guideline.

  11. Compared with the suggested standard provided by the NCCLS • limited to hypoglycemic readings: none of the three methods reached this degree of association (blood gas/chemistry: r = .9251; capillary blood: r = .4942; arterial blood: r = .6153), • normoglycemic ranges: blood gas/chemistry: r = .9506; capillary blood: r = .7837; arterial blood: r = .8576)  none • hyperglycemic range: blood gas/chemistry: r = .9669; capillary blood: r = .8652; arterial blood: r = .8251) none

  12. DISCUSSION • Overall, the performance was poor in all critically ill patients, as clinical agreement with the central laboratory was less than 80% for all tested methods of glucose measurement. • When blood sugars were in the hypoglycemic range, clinical agreement was <70% for all tested methods and as low as 26.3% with glucose meter analysis of capillary blood.

  13. DISCUSSION • Blood gas/chemistry analysis appears to be the most accurate and reliable method. • The blood gas/chemistry analysis may also be the safer approach, since hypoglycemia appears to be overestimated, • whereas glucose meter analysis of capillary or arterial blood would underestimate hypoglycemia.

  14. Thanks for your attention !