An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Analog Devices ICs for In Vitro Diagnostics Equipment

  2. Agenda • In Vitro Diagnostics Overview • Example Application: Flow Cytometry • General block diagram • Detectors & Sensors • Amplifying signals from photodiodes • Digitizing the analog signals • Digital processing • Steering cells into collection chambers • Additional Information & Resources

  3. In Vitro Diagnostics Overview • Testing tissue or fluid samples • Used in diagnosis and monitoring applications in humans and animals • Example applications: • Cellular analysis – hematology, flow cytometry • Chemistry – glucose, cholesterol, electrolytes, proteins, enzymes • Immunoassay – determines concentration of a substance by measuring reaction of an antibody • Electrophoresis – protein, DNA, RNA analysis

  4. Flow Cytometry • Flow Cytometry is a process through which cells are differentiated according to cell characteristics • Size, Granularity, Structures present, Reaction to a compound • Application • Separate specific cell types from heterogeneous mixture • Determine cell reaction to compounds • Development and testing of new drugs • Detect presence of cells in small quantities • Benefits of Flow Cytometry • Detection of biological agents in minutes • Capable of detecting agents at low concentrations

  5. CytometryProcessor ADC Photo-diode LightSource AMP Pulsed voltage CellCharger DAC - + DeflectionPlates How does Flow Cytometry work? Sorting cells, one by one • Stained cells pass in front of light source • Deflection plates can steer charged cells into different containers Amp Requirements • Wide BW for fast signals • Low bias current • Low input capacitance • Fast Settling Time DAC Requirements • Applies charge to each cell • Multiplies pulses – must have wide BW to accommodate different cell velocities (500kHz – 2MHz) • High Speed Interface

  6. Flow Cytometry Detectors/Sensors • Photomultiplier tubes (PMTs) & Avalanche photodiodes (APDs) • Fluorescence detection in 400-1000 nm wavelength range • high sensitivity • high internal gain (106) • fast response (10−7 to 10−9 s) • yield a large S/N due to their internal gain • Next generation: Solid-state lasers and PIN-based photodetectors • a few nA of current in the PIN photodiodes, a lock-in amplification technique is applied to increase the signal to noise ratio • lock-in circuit parameters such as the time constant and sensitivity are selected to maximize signal to noise ratio at the given modulation frequency

  7. C2 f = SIGNAL BW 2 R2 f = OP AMP UNITY _ u GAIN BW PRODUCT I C1 f 1 u Total Input f = p 1 2 R2 C1 Capacitance + –V 1 B f = p 2 2 R2 C2 OPEN LOOP GAIN f = f f • GAIN 2 1 u UNCOMPENSATED C1 NOISE C2 = p 2 R2 f GAIN u COMPENSATED FOR 45° PHASE MARGIN f f 2 u 1 f f = p 2 2 R2 C1 f f 1 u Amplifying Signals from High Speed Photodiodes

  8. Photodiode Preamp Example using AD8067 C2 = 0.35pF CD = 4pF, CIN = 4pF C1 = CD + CIN 33.2k 33.2k 33.2k +12V D1 – AD8067 C1 = 8pF –10V + fu = 350MHz D2 D1,D2: IF-D91 33.2k –12V C3 = 0.35pF Bandwidth: 7MHz Output Noise: 725µV RMS, 21MHz 550µV RMS, 10MHz 33.2k 33.2k

  9. Comparison of Op Amps for Photodiode Preamps Unity GBW fu, MHz 25 65 40 350 24 Input Capacitance CIN, pF 23 6.6 4 4 2.5 fu/CIN MHz/pF 1.1 9.85 10 87 9.6 Ib pA 2 2 1.5 2 <1 VN@10kHz nV/Hz 6 7 11 7 7 *AD8610/20 AD8065/66 AD8033/34 AD8067 G > 9 Stable AD8615/6/8 * Ideal low frequency precision preamps for large area photodiodes operated in photovoltaic mode (zero volt bias)

  10. ADC Requirements: SNR & Sample Rate • Bench-top Equipment • Good SNR (typically 12+ ENOBs ≈ 72+ dB of SNR) • Low power to minimize heat • Larger Equipment • Excellent SNR (16+ ENOBs ≈ 96+ dB of SNR) • Frequently use over-sampling to increase SNR in digital domain • Can average many samples to reduce noise. • New SNR = SNRDATASHEET + 10*Log(FSAMPLE/2*BW) • Example:AD9446 SNR = 82dB at 80MspsCustomer BW is 2 MHzEffective SNR of AD9446 will be:82dB + 10*Log(80MHz/4MHz) = 82dB + 13dB = 95dB • The right ADC choice depends on: • Power • SNR • Sample Rate

  11. Some Recommended ADCs • Benchtop Equipment (low power, multi-channel) • AD9252 & AD9259: 14 bits, 8 & 4 channel ADCs with serial LVDS output • AD7621: 16 bits, 1 channel ADC with very low power (86mW) • Large Equipment (best performance, highest SNR) • AD9446 – 82dB SNR at 80Msps • AD7625 – 90dB SNR at 5Msps • Higher Resolution & Lower Sampling Rate ADCs

  12. Digital Processing in Flow Cytometry • Each cell produces a Gaussian distribution of a/d conversions as it passes the sensors • FPGA Tasks • Take each sample above a threshold. 50Mhz converter. • Do a numerical integration (rectangle addition), this is the Area • Determine Base Width based on threshold • Pass to the DSP • DSP Tasks • Receive FPGA dispersion • Convert to floating point • Compensate for filter overlap • Compare to desired cell type • Steer cell (desired or non desired) via high voltage

  13. Steering the Cells into ContainersGenerating pulsed voltage to charge cell DAC Requirements • Very fast throughput • >10MHz Bandwidth • Fast digital interface • Support Low voltage designs • ± 10V Reference with 2.5V to 5.5V supply • Current output – convert to voltage with op amp • Extended temperature range Pulsed voltage CellCharger - + DeflectionPlates

  14. Additional Information & Resources • Motor Control Solutions • Current Sense Amplifiers • Gate Drivers • Resolver-to-Digital Converters • Digital Potentiometers • Temperature Sensing and Thermal Management • Analog Temperature Sensors • Thermocouple Conditioners • Digital Temperature Sensors • Fan Control • Other Medical Solutions from ADI