Results of IAEA supported TPS audit in Europe Eduard Gershkevitsh North Estonia Medical Center Tallinn, Estonia
Principles of operation • TPS audit uses IAEA TECDOC 1583 methodology • IAEA provides CIRS Thorax phantom on loan for 6 month to the Member State • IAEA together with national audit coordinator (Medical Physics group, RT department nominated by the national authorities) organise one day workshop in the country • National audit coordinator’s centre is audited by independent auditor • National audit coordinator is performing audits through on site visits
IAEA TECDOC 1583 methodology • Based on anthropomorphic phantom • To verify that logistic chain: • CT scanning • Treatment planning • Data transfer • Dose delivery • is operational and leads to desired results with sufficient accuracy • Employs ionisation chamber measurements
IAEA TECDOC 1583 methodology • Eight test cases with 15 measurement points • Single 10x10cm2 field at nominal SSD • Tangential field with wedge • Corner blocks • 4-field “box” • Customised blocking • Oblique incidence with L-shape block • Half fields with wedges • Non-coplanar field arrangement • Agreement criteria 2-5% depending on complexity
National TPS audit coordinators • B. Petrovic– Institute of Oncology Vojvodina, Sremska Kamenica, Serbia • C. Pesznyak– National Institute of Oncology, Budapest, Hungary • K. Chelminski– M. Skłodowska-Curie Memorial Cancer Center and Institute of Oncology, Warsaw, Poland • J. Grezdo– St. Elizabeth Institute of Oncology, Bratislava, Slovakia • M. do Carmo Lopes– Portuguese Institute of Oncology, Coimbra, Portugal • E. Gershkevitsh– North Estonia Medical Centre, Tallinn, Estonia
Participants of the TPS Audit • 8 countries • 61 centres • 195 datasets(combination of algorithms and beam quality)
CT to RED conversion • CT to RED conversion curve required adjustment in 2/3 of centres(criteria used adopted from IAEA TRS 430 – for the same electron density the variation should not exceed ± 20HU for all materials except water (± 5HU)
Dosimetry problems • Discrepancies requiring intervention and not related to algorithm limitations were found in approximately 9% of datasets Reasons for deviations
Calibration • Use of chamber with outdated calibration factor • Use of plastic phantom instead of water for calibration • Incorrect value in TPS
Input beam data & model fitting • Typographic errors • Use of standard data • Quality of measurement data • Sub-optimal beam fitting
Problems with treatment unit • New couch top was installed and unaccounted. This lead to 8% underdose from posterior fields • Loose mechanical wedge
Accuracy achievable • With corrected data and advanced algorithms the majority of the measurements are within agreement criteria
Conclusion • Input beam data and suboptimal beam modelling werethe largest contributors to observed deviations • CT to RED conversion is customised in minority of centres • The majority of observed deviations have been corrected • Contribution to better understanding of TPS performance and its limitations
Acknowledgement • Stanislav Vatnitsky for drafting a proposal for TPS audit • Joanna Izewska for support and implementation of the TPS audit at IAEA • To national audit coordinators • To medical physicist at audited RT departments
Thank you for attention! More details: Gershkevitsh E, Pesznyak C, Petrovic B, Grezdo J, Chelminski K, do Carmo Lopes M, Izewska J, Van Dyk J. Dosimetric inter-institutional comparison in European radiotherapy centres: Results of IAEA supported treatment planning system audit. ActaOncol. 2014 May;53(5):628-36.