An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 自動化溫室效應氣體分析系統之建立與應用 Construction of an automated GC system for measuring atmospheric greenhouse gases Speaker:Ming-Xia Li Advisor:Prof. Jia-Lin Wang

  2. Introduction GWP100:1 364 ~ 408 ppm GWP100:298 303 ~ 326 ppb GWP100:25 1758 ~ 2133ppb

  3. Analytical Techniques

  4. Our analytical system Oven:70℃ Sampling system • Sample automatically fills up to 700 torr

  5. Column Configuration CH4 & CO2 N2O

  6. Actions other compounds Injection O2, N2O, other compounds O2, N2O other compounds Sampling Back-Flush O2, N2O O2 N2O 6

  7. Channel switching Methanizer • Composition:30~50%NiO(nickel(II) oxide)and 50~70% SiO2 • Function : reduced CO2into CH4 • Temperature : 375°C CH4 CO2 H2 CO2 + 4H2→ CH4 + 2H2O

  8. Chromatograms CO2 FID CH4 ECD N2O O2

  9. Experimental • Carrier gases • Oxygen cutting or not • Parameter optimization

  10. Different carrier gases – N2 vs. CH4/Ar N2 CH4/Ar N2O • When the carrier gas uses 5% CH4/Ar, the single of N2Owill increase

  11. Cutting oxygen or not Not cut oxygen Cut oxygen N2O O2 • When cut oxygen, the single of N2O just increase slightly, but the analyze time will increase

  12. Optimize the parameters

  13. Results and discussion • Linearity • Limit of Detection • Reproducibility

  14. Connectingwith Vacuum System vacuum system • When system connects with vacuum system, makes sure samples are not be contaminated.

  15. Quantitative – Working Standard Calibration curve - CO2 • Using NOAA standards to make calibration curves

  16. Linearity • Pressurerange:4 ~ 44 psi

  17. Limits of Detection N = 7 • Pressure:4 psi • Based on 3 times standard deviation  Reproducibility N = 7 • Pressure:10 psi

  18. Canister Analysis

  19. Canister sampling in Taipei area in 2012 • Sampling time:2012/02/14 05:00~06:00 am 18:00~19:00 pm • 53 sampling sites • Sampling time less than 1 min • Monitoring compounds:CFCs, CO2, CH4, N2O, NMHCs • Look for emissions hot spots – sources of GHGs. Ref. Chemosphere 1998, Vol.36, No. 10, 2391-2400 Ref. Analytica Chimica Acta 2006, 576, 91-99

  20. CO2 Concentration range: 387.04 ~ 578.12 ppmv Morning (05:00 ~ 06:00) Evening(18:00 ~ 19:00)

  21. CH4 Concentration range: 1837.57 ~ 4496.74 ppbv Morning (05:00 ~ 06:00) Evening(18:00 ~ 19:00)

  22. N2O Concentration range: 392.06 ~ 421.88 ppbv Morning (05:00 ~ 06:00) Evening(18:00 ~ 19:00)

  23. Conclusions • The system measures CH4、CO2、N2O in 10 minutes for each sample. • It has very good reproducibility (RSD < 1%), and linearity (R2= 0.9999). • The limits of detection for CH4is 138.34 ppb (vs. ~1.7 ppm), CO2 is 3.42 ppm (vs. >360 ppm), and N2O is 2.3 ppb (vs. >300 ppb).

  24. Thanks For Your Attention