An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Basic MRI I Chapter 2

  2. The Magnet • The heart of the system • Induces tissue magnetization • Induces resonance with a specific frequency • Frequency is proportional to field strength

  3. Frequency is proportional to field strength • Freq = gyromagnetic ratio x field strength • Freq = “gamma” x B-naught • Freq = γHB0 • Gyromagnetic ratio = γH = 43.58 MHz/T • Field strength = B0

  4. Magnetic field • Magnetic fields have strength and direction – they are vectors • Strength measured in Gauss or in Tesla • 1 Tesla = 1 T = 10,000 Gauss = 10,000 G • 1000 G = 1 kG • Direction is defined as the “z-direction” in MR, in 3D Cartesian coordinates

  5. Type of Magnets

  6. Gradients • Intro to gradient concept • Linear gradients • Non-linear gradients

  7. Gradients • Orientation • Definition of orthogonal = “ at right angles” • Functions • To get images—all types • To get specialized images (gradient echo) • To reduce artifacts • Strength • Units of mT/m • High value = ~ 20 mT/m

  8. Gradients • Rise time = time for gradient to reach its maximum, like 0.20 msec • Slew rate = rate of gradient change, like 100 mT/m/msec

  9. Eddy currents • Electrical currents induced or generated in conducting materials, in metal in the machine, minimized by gradients

  10. Shimming • To achieve homogeneous field strength

  11. Shielding • Passive—large pieces of ferromagnetic metal like iron • Active—electromagnets to decrease the extent of the field

  12. RF system • Coils • Body coils • Surface coils • Special coils • TMJ • Head • Wrist • Linear, quadrature, CP, phased array

  13. RF shielding • Farraday cage • To decrease stray RF energy from lights and motors outside the magnet from getting into the receive coil (antenna) • Copper screen • Sheet metal

  14. Computer • To store and acquire signals in k-space • To reconstruct images • To perform Fourier Transform • Storage, retrieval, transmission of images • Post Processing, zoom, magnification, contrast/windowing, 3D reconstructions