E N D
1. Quality Control in Diagnostic Radiology
2. Factors driving Q.C.Why do we do it? Legal Requirements
Accreditation
JCAHO
ACR
Clinical improvement
equipment performance
image quality
3. Q.C. Goals Minimize dose to
patients
staff
Optimize image quality
Establish baselines
More on this in a moment
4. Why is Q.C. Important?
5. QC can detect Malfunctions
Unpredictability
may be hard to isolate clinically
Inefficient use of Radiation
high fluoroscopic outputs
Radiation not reaching receptor
inadequate filtration
oversized collimation
6. Goals of a Q.C. Program Obtain acceptable image with least possible radiation exposure to
patients
staff
Attempt to identify problems before they appear on patient films
without QC problems only detected on patient films
7. “Acceptable” Image Image containing information required by radiologist for correct interpretation
goal: minimize exposure while maintaining acceptability
high exposure images often have excellent appearance
Low noise
8. Q.C. & Baselines Baselines
quantitative data on equipment obtained during normal operations
Baselines useful for troubleshooting
isolating problem component, for example
generator
processor
Allows efficient use of engineering / repair personnel
9. X-Ray Quality Control Filtration
Focal Spot Size
Collimation
Maximum Fluoroscopic Output
Calibration Verification
Phototimer Performance
10. Why is Filtration Important? Tube emits spectrum of x-ray energies
Filtration preferentially attenuates low energy photons
low energy photons expose patients
do not contribute to image
low penetration
11. Half Value Layer (HVL) We don’t measure filtration
We measure HVL
HVL: amount of absorber that reduces beam intensity by exactly 50%
12. Half Value Layer Depends upon
kVp
waveform (single/three phase)
inherent filtration
Minimum HVL regulated by law
Maximum HVL regulated only in mammography
13. Radiographic HVL Setup
14. Checking HVL Compliance(Radiographic) How much aluminum must be placed in beam to reduce intensity by exactly 50%?
15. Checking HVL Compliance(Radiographic) Is this machine legal?
2.5 mm Al minimum filtration at 90 kVp
16. Fluoroscopic HVL Setup
17. Fluoroscopic HVL Set desired kilovoltage manually
measure exposure rates instead of exposure
Move absorbers into beam as needed
18. Focal Spot Size We measure apparent focal spot
Trade-off
smaller spot reduces geometric unsharpness
larger spot improves heat ratings
19. Focal Spot Size (cont.) Focal spot size changes with technique
Standard technique required
75 kV (typical)
50% maximum mA for focal spot at kV used
direct exposure (no screen)
NEMA Standardsdefines tolerances
20. Focal Spot Measuring Tools Direct MeasurementPin Hole Camera
Slit Camera
Indirect Measurement of Resolving Power
Star Test Pattern
Bar Phantom
21. Direct Focal Spot Measurement Measure focal spot directly in each direction
Use triangulation to correct for distances
formula corrects for finite tool size
two exposures required for slit
22. Star Test Pattern Measures resolving power
infers focal spot size
Dependent on focal spot energy distribution
measure
largest blur diameter (in each direction)
magnification
use equation to calculate focal spot size
23. Bar Phantom Measures resolving power
Find smallest group where you can count three bars in each direction
24. Bar Phantom Setup
25. Radiographic Collimation X-Ray / Light Field Alignment
Beam Central Axis
should be in center of x-ray beam
Collimator field size indicators
PBL (automatic collimation)
field automatically limited to size of receptor
Bucky Alignment
Using longitudinal bucky light & transverse detent, x-ray field should be centered on bucky film
26. X-Ray / Light Field Alignment Mark light field on table top with pennies
27. Radiographic X-Ray / Light Field Alignment
28. Fluoroscopic Collimation image field is scale seen on monitor
expose film on table above scale
compare visual field (monitor) with x-ray field on film
must check all magnification modes
29. Fluoroscopic Collimation
30. Fluoroscopic Collimation
31. Maximum Fluoro Output put chamber in beam on tabletop
block beam with lead above chamber
fools generator into providing maximum output
10 R/min. limit for ABS fluoro
32. Maximum Fluoro Output
33. Calibration Performance Parameters Timer Accuracy
Repeatability
Linearity/Reciprocity
Kilovoltage accuracy
mA
must be measured invasively
34. Calibration mR/mAs should stay constant for all combinations of mA & kVp at any particular kVp
35. Calibration mR/mAs should stay constant for all combinations of mA & time at any particular kVp
36. Phototiming(check with output or film) Reproducibility
Density Controls
Field Placement
Field Balance
39. Phototimer Field Placement / Balance Placement
cover desired field with lead
select field as indicated
Balance
no fields covered
select field as indicated
40. Phototimer Field Placement / Balance
41. Phototimingchecked with Exposure Index kV Response
phototimer pick-up attenuation may vary with kV
phototimer must track kV response of rare-earth film
Rate Response
Check with varying
phantom (lucite) thickness
mA
43. The End