Riham MARIAM, Matias RODRIGUES , Martin LOIDL CEA, LIST Laboratoire National Henri Becquerel

Riham MARIAM, Matias RODRIGUES , Martin LOIDL CEA, LIST Laboratoire National Henri Becquerel. First measurements of detailed absolute emission intensities of L X-ray emitted by actinides using a MMC. 18 th International Workshop on Low Temperature Detectors. matias.rodrigues@cea.fr.

rigg
Download Presentation

Riham MARIAM, Matias RODRIGUES , Martin LOIDL CEA, LIST Laboratoire National Henri Becquerel

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Riham MARIAM, Matias RODRIGUES, Martin LOIDL CEA, LIST Laboratoire National Henri Becquerel First measurements of detailed absolute emission intensities of L X-ray emitted by actinides using a MMC. 18th International Workshop on Low Temperature Detectors matias.rodrigues@cea.fr

  2. Radioactive material analysis by photon spectrometry • X-ray and gamma spectrometry : • Many fields of applications use photon spectrometry in particular of actinides • Nuclear fuel cycle • Non-proliferation of nuclear weapons • Metrology of ionizing radiations • However the knowledge of the photon emission intensity (uncertainties > 1%) is a fundamental limit to the precision of measurements Radioactive material Photon live time [s] : FEP detection efficiency : Emission intensity [photons / disintegration] Spectrometer 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  3. Knowledge of L X-ray • emission intensities of actinides gand X L-rays 10 - 100 keV : ~ 0.01 to 10 photons / 100 disintegrations •  They are only few and old measurements • Calculations and measurements are not always consistent • The L X-ray lines are not detailed 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  4. Example: emission intensities of Pu-238 Recommended values evaluated by the DDEP (Decay Data Evaluation Project) ~ 10 evaluated measurements of L X-rays  recommended value 10.63 (8) % XL Deviation from recommended value (%) 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  5. Difficulties to measure L X-ray intensities • with semiconductor spectrometers live time [s] : FEP detection efficiency : Intensity [photons / decay] Source Photon • Using semiconductor spectrometers: • Insufficient energy resolution to separate the L X-ray • Correction of the detection efficiency by 15% 241Am L X-ray spectrum Spectrometer >15% La Lb Lg Li 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES L X-ray

  6. A dedicated MMC for Photon spectrometry • Of L X-Ray of actinides Au-Ag absorber KIP Heidelberg 1 mm2 Sensor Thermal bath T0 ~ 10 mK SQUID V V Input coil • 4 absorbers of 1 mm2 • 50 mm of Au + 17 mm of Ag thick • Intrinsic efficiency > 99% between 10-25 keV • 10 – 20 s-1 (td ≈ 4 ms) • Energy resolution FWHM of 26 eV MMC with 4 pixels Thermal link AuAg absorber SQUID chip 2 mm 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  7. FULL ENERGY PEAK detection efficiency calibration USING 241Am AND MC SIMULATIONS Source Stainlesssteel Be window Z Z pixel [PhD thesis of Riham MARIAM] Y Y X MMC Pb shield W collimator W collimator 40 mm thick Be window 15 mm • 10 keV < E < 25 keV ~ 0.7% uncertainty • < 10 keV ~ 4% uncertainty at 2 keV • > 25 keV ~ 1.2% to 2.4% uncertainty Radioactive source Stainlesssteel source support 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  8. Measured radionuclides 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  9. Additional corrections for the absolute emission measurements • Correction of self-absorption in the source by MC simulation: • 239Pu source thickness ~4.5 µm • 237Np source thickness ~6.2 µm • Efficiency transfer by MC simulation: • Source diameter of 12 mm instead of 6 mm  -15.6% correction • Shift of the source position •  -1.4% for 238Pu •  -0.8% for 244Cm [PhD thesis of Riham MARIAM] Activity distribution of 238Pu Detection efficiencymap 12 mm • Correction of -3.1% for 238Pu • Correction of +3.1% for 244Cm • Correction of -1.8% for 237Np

  10. Comparison of total L X-ray emission intensities 238Pu 10.76 (10) / 100 decays Preliminary 244Cm 237Np 9.12 (12) / 100 decays 63.0 (9) / 100 decays Deviationwithrecommended value (%) 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  11. 10000 1000 Counts per channel 100 10 1 11000 12000 13000 14000 15000 16000 17000 18000 19000 20000 21000 22000 Energy (eV) • L X-ray spectra with a high level of details 238Pu FWHM energy resolution of between 25 and 45 eV  measurements of about 30 L X-ray emission intensities 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  12. Individual absolute L X-ray emission intensities Preliminary 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  13. 10000 50000 10000 1000 1000 Counts per channel 100 100 10 10 1 18600 19600 20600 21600 14600 15600 16600 17600 Energy (eV) Energy (eV) XLi GROUP EMISSION INTENSITIES XL3 XLgof 238Pu XLbof 238Pu XL2 XL1 The Li-X lines separated individually  Determination of the and the XLi groups • Both 238Pu and 239Pu decay to U • But the relative XLi intensities depend on the isotopes Ni :primary vacancies created by IC 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  14. 5000 10000 1000 1000 Counts per channel 100 Counts per channel 100 10 10 12600 13100 13600 14100 12800 13300 13800 14300 Energy (eV) Energy (eV) • L X-Ray Satellite intensities 238Pu 237Np , 233Pa U La2 Solid line : diagram Dashed line : satellites U La1(L3-M5) Pa La1 U La1 (L3-M5) U La2 Pa La2 233Pa: La1(Sat)/La1(Diag) = 18.7% 238Pu: La1(Sat)/La1(Diag) = 2.2% M5 M1 L3 L2 L1 X-ray satellite E’(L3M4-M4M5) > E(L3-M5) Shift up to 100 eV CosterKronig transition L3-L1M4 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  15. Conclusions • Calibration of the FEP detection efficiency with 0.7% uncertainty between 10 and 25 keV • Measurement of L X-ray spectra from 8 radionuclideswith FWHM energy resolution of 25 – 45 eV up to 60 keV • Total L X-ray emission intensities in agreement with recommended values • About 30 absolute individual emission intensities for 4 actinides • Uncertainties dominated by sources positioning and activity inhomogeneity can be improved • New X-ray MMC detector for photon emission intensities < 10 keV  MetroMMC project, Poster [122] RANITZSCH,Phillipp Chung-On Perspectives 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  16. Thank you for your attention

  17. LIVE time determination using extendable dead time Pile-up on the pulse decay Raw stream Filtered stream Pulse start Pulse stop • Dead time window proportional with the pulse duration • Dead time suppress pile-up on the baseline • Pile-up on the decay suppressed using Khi2 with template pulse • Correct the live time with thePoisson law : probability to have 0 pulse during the pulse decay time • Live time and signal processingdeterminationvalidatedwithsimulated data streams Pulse height (a.u.) tstart Time (s) Pulse height (a.u.) Khi2

  18. INTENSE SATELLITE STRUCTURES : EXAMPLE FOR 210Pb L3-M5 (La1) Total fit Diagram Satellite Shake-off Fluorescence Background L3-M4 (La2) 10000 L3-M1 (Ll) Pb Zn W L2-M1 (Lh) Po W L3-M2 1000 Au W L3-M3 W Counts per channel 100 10 9200 10000 10800 11600 M5 M1 L3 L2 L1 M5 M1 L3 L2 L1 Energy (eV) X-ray transition line (diagram line) E(L3-M5) = 10.839 eV X-ray satellite E’(L3M4-M4M5) > E(L3-M5) Shift up to 100 eV CosterKronig transition L3-L1M4

  19. Relative emission intensities of L X-ray from 241Am 241Am 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES

  20. Difficulties to calculate L X-ray emission intensities from radionuclide decays The calculations require nuclear and atomic fundamental parameters (FPs): • g transition probabilities and partial ICCs  Primary vacancy distribution • Coster-Kronig transition probabilities fij Reorganized vacancy distribution • Fluorescence yields wLi Probability of radiative transitions for each subshell Li • Radiative transition rates  If the intensity of each L X-ray transitions is needed Large uncertainties or systematic errors because: • Atomic FPs fijand wLi have large uncertainties (5-50% for fij,5-20% for wLi) [J.L. Campbell, At. Data. 5 (2003) 291–315] • For anomalous g transitions ICCs cannot be calculated (i.e. 241Am, 240Pu) [V.M. Gorozhankina and M.-M. Be, Appl. Radiat. Isot 66 (2008) 722–728] • Atomic FPs calculations simplified by some assumptions (i.e. single vacancy atomic configuration) 18th Workshop on LowTemperature Detectors| Milano | 22-26/07/2019 | Matias RODRIGUES