An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Operation ofPersonal Radiation DetectorsPolimaster PM1703MO1

  2. PM1703M Personal Radiation Detector Polimaster PM1703M Alarm light Battery case Infra red connection to PC Display Clip for belt Menu button mode Menu button light

  3. PM1703M (back cover) • GM tube (in latest model only) located here • Location of the centre of the CsI crystal • Battery Compartment • 1 AA battery • Correct direction for battery insertion shown

  4. PM1703M Display • Counts Per Second (cps) display “ s-1” or Dose Rate display – switchable with mode button • “ s-1” mode: number per sec + Uncertainty in % • Dose Rate mode: unit microSv/h + Uncertainty (%)

  5. Working with the Polimaster PRD • Switching On: press and hold mode button until instrument shows reaction: • test of alarm indicators (LED, sound, vibration) • Display check, firm ware version • Battery check (“batt”) • Device self-test (“test”) • background measurement (“cal”) is taken for 30 seconds • afterwards pager is ready for use! • Working with the radiation pager: • Monitoring mode: wear the pager on the belt during your normal activities • In case of an alarm by the pager, it automatically goes into Search Mode • Switching Off: • Press and hold light button until pager is switching off

  6. The PM menu and IAEA default setting • Search alarm based on 5,3 Sigma, safety alarm at 100microSv/h (set only by using computer) • Limited Menu Selection on PRD: • Recalibration of the background • Alarm indicator vibration ON – OFF • Alarm indicator sound ON – OFF • Changing Modes: • Press mode button until Calibration display is visible • Press mode button again -> Audio: ON-OFF selection by light button • Press mode button again -> Vibration: ON-OFF selection by light button

  7. Search with the Polimaster PRD • No manual activation of the search mode! • In case of an alarm, system automatically goes to search mode • Frequency of acoustic and light pulses is proportional to radiation strength • Move detector slowly, and as close as possible to surface • Keep 5-10 cm from surface to avoid possible contamination! • Do not shield the sensitive point (top side of pager) with your fingers • Watch display (amount of cps) and/or frequency of light and/or frequency of sound

  8. Pager configuration per PC • Infrared connection to PC available • Detector history (alarm logbook) • Device settings (password protected)