An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Development of an Ultrasound Lab Laura Wade April 4th 2012 3970Z

  2. Introduction • Piezoelectric – an alternating voltage across the crystal causes it to flex and contract, emitting sound. • Piezoelectrics also generates alternating voltage in response to a returning sound wave. • It emits sound waves and receives them.

  3. Speed of sound depends on compressibility of a material • Acoustic Impedance (Z) is a measure of resistance to sound waves. • Large differences in Z create strong refections (signals)

  4. B Mode Imaging • Produces a 2D grayscale Image. • Brightness is proportional to amplitude of the reflected sound waves. • The time at which the signals are received indicates depth. • c= 2D/t

  5. Colour Doppler • Velocity information is represented by colour and is overlaid onto a 2D B-Mode image. • Velocity is determined using the Doppler effect: Δf = 2f0 (v/c) cosα

  6. PulseWave Doppler: velocity is measured at a specific depth, which can be adjusted • Continuous Wave Doppler: measures all velocities along the ultrasound beam. It provides no information about depth of the signal.

  7. Colour Power Doppler: • Displays the amplitude of the frequency shift. • Amplitude is a function of the number of reflectors (RBCs) with that velocity. • Colour is still used to determine direction

  8. Objectives • Develop an experiment using sonography to measure blood flow in the carotid artery. • Develop a complete set of instructions for the operation of the equipment as it applies to this lab. • Determine a way to analyze the data acquired from the sonographs.

  9. Approach • Research the theory behind ultrasound • Master the technical systems to be used in the lab • Research possible parameters and treatments to use in the lab • Develop appropriate protocol

  10. Hypotheses • Sonography can be used to verify continuity of flow in the carotid artery. • Increasing both physical and mental activity will increase blood flow in the carotid arteries.

  11. Methods • Carotid Ultrasound Carotid is located at a depth of 3-4cm beneath the surface of the skin. • Remember to calibrate the system to the angle the transducer is held at.

  12. A 38-element linear array transducer is used • Uses frequency of 5MHz • Sonosite 180

  13. Measurements and Calculations • Flow in the right carotid before and after the carotid bifurcation using PWD. • A1v1 = A2v2

  14. Cardiac Output (CO) • Measure peak systolic velocity and end diastolic velocity. • Calculate Volume Flow Rate (CBF) • Use known relationship to calculate CO

  15. Cerebral Blood Flow (CBF) before and after exercise and/or mental activity • Volume Flow = Area * Velocity

  16. Results • CBF = ~750mL/min at rest • Flow in the carotid before and after the bifurcation is equal. • CO = 5 – 5.5 L/min at rest • Flow in the carotid is increased during both exercise and increased mental activity. • Paired t – test results in significance with p<0.05.

  17. Discussion • Why should we incorporate this lab into the 3970Z curriculum? • Ultrasound is covered in both 3rd and 4th year courses • Noninvasive, inexpensive, and therefore common imaging technique

  18. Sources of Error • Inaccurate measurement of cross sectional area. • Inaccurate angle correction. • Noise

  19. Questions for Discussion: • Would the effectiveness of sonography be different for an obese patient? Why? • Would ultrasound be effective for imaging blood vessels in the torso? • How could an occluded blood vessel be detected? • What would be the effect of not using lubrication between the skin and transducer?

  20. Acknowledgements I would like to thank: • Dr. Ian MacDonald, Supervisor • Michelle Belton, Lab Manager

  21. Thank you for your time. Any Questions?