Workshop on Uncertainty in Humidity Measurements Uncertainty in calibrations with secondary standard Jovan Bojkovski, Domen Hudoklin Laboratory of Metrology and Quality (LMK) Faculty of Electrical Engineering (FE), University of Ljubljana MIRS/FE-LMK Slovenia Dubrovnik, 21.6.2004
Summary • What is secondary humidity calibration ? • Humidity “generator” • Reference standard • Measurement methods • Sources of uncertainty in secondary calibration • Practical methods to quantify the components
Summary • Uncertainty related to the reference standard • Uncertainty related to the calibration equipment • Validation • Conclusions
Secondary calibration • Calibration is the process of determining the difference in value of a measuring device to a primary standard with establishing all the unceratinty sources • Calibration in the stable humidity environment using a calibrated reference dewpoint hygrometer and/or thermometer for relative humidity (RH) calibration
Secondary calibration • Calibrated reference dew-point hygrometer is calibrated against primary standard (two pressure generator, single pressure single temperature generator, ...) • Thermometer is calibrated either by comparison in secondary thermometry laboratory or at fixed points
Humidity “generator” • Its role is producing stable and homogeneous humidity environment • Actual value of the humidity is measured with a reference dew-point meter • Types of humidity generators (two pressure, single pressure single temperature, mixed flow, climatic chamber, ...)
Reference standard • Reference standard for humidity measurement is usually dew point meter calibrated against primary humidity generator • It can be also RH hygrometer and psychrometer, but the uncertainty is then typically larger
Reference standard • Reference standard for temperature measurement can be different • Thermometer which is part of reference dew point meter • Resistance thermometer ((S)PRT, thermistor) with measuring device (bridge, multimeter) • Other temperature measuring devices
Measurement methods • Depending on the DUC (RH/absolute humidity) • Parallel sampling • Serial sampling • Measurement of air temperature • Potential pressure drops • Leakage • Type of gas
Calibration with secondary standard • Set-up for dew-point calibration(Device under calibration –DUC- is dew-point meter) generator Reference dew-point sensor 3,45°C DUC
Calibration with secondary standard • Set-up for RH calibration(DUC is RH hygrometer) generator Reference dew-point sensor 3,45°C DUC (RH) thermometer
DUC Reference standard t90 Sources of uncertainty Consequently, this would result in the following sources of uncertainty: • due to generation of humidity • due to uncertainty of reference standard • due to DUC • due to calibration method and ambient conditions generator
Uncertainty of humidity generation The reference value (RV) is measured by calibrated dew-point sensor, so only the deviations from the measured RV need to be characterised. The major uncertainty contributions are therefore: • instability of generated humidity • inhomogeneity of generated humidity • pressure drop • uncertainty of measurement device, used for evaluation
Uncertainty of humidity generation INSTABILITY OF GENERATED HUMIDITY • instability of dew-point • instability of temperature (for RH sensors) • typically less significant component
Uncertainty of humidity generation INSTABILITY OF GENERATED HUMIDITY • instability can be measured during the calibration with stable reference humidity and/or temperature standard • standard uncertainty due to instability is taken as a combination of standard deviation of both dew-point and temperature
Reference standard DUC Uncertainty of humidity generation INHOMOGENEITY OF GENERATED HUMIDITY-Dew-point • almost insignificant in dew-point calibration (if the inlet of the sampling tube of standard and DUC are in parallel and close to each other) • problems, if reference standard and DUC as dew-point meter are sampling in serial (pressure drop, contamination,...)
Uncertainty of humidity generation INHOMOGENEITY OF GENERATED HUMIDITY - RH • in RH calibration – large influence of temperature on relative humidity • Therefore temperature inhomogeneity needs to be quantified – can be very significant
Uncertainty of humidity generation INHOMOGENEITY OF GENERATED HUMIDITY - RH RH(T) dependence
Uncertainty of humidity generation INHOMOGENEITY OF GENERATED HUMIDITY - RH
Uncertainty of humidity generation INHOMOGENEITY OF GENERATED HUMIDITY - RH • Typical uncertainties of RH due to temperature inhomogeneity – from 0,1%r.h. to several %r.h. • The inhomogeneity can be reduced if smaller calibration space is used or appointed 34 times smaller or more
Uncertainty of humidity generation INHOMOGENEITY OF GENERATED HUMIDITY - RH
Uncertainty of humidity generation INHOMOGENEITY OF GENERATED HUMIDITY - RH • Temperature gradients can be evaluated previous to calibration – less expensive possibility... • ... or each time during the calibration (the set of thermometers has to be used all the time) • the second one is better, because the influence of DUC can also be determined, however more expensive (thermoemeters, bridge, scanner, PC...)
Uncertainty of humidity generation INHOMOGENEITY OF GENERATED HUMIDITY - RH To evaluate inhomogeneity, thermometers are typically placed in corners of calibration area and one in the geometrical center.
Uncertainty of reference standard The uncertainty of reference standard depends naturally on the type of standard used. In general the following uncertainty contributions are to be determined: • uncertainty, taken from the calibration certificate (dew-point standard & temperature standard) • standard’s drift • resolution (if not included in certificate) • uncertainty of calculation of relative humidity from dewpoint,temperature (water sat.pressure eq.,enh.f.eq.) • uncertainty of electrical instrument (bridge, multimeter,...)
Uncertainty of reference standard In RH calibrations, if the reference is dew-point sensor, then the temperature and dew-point contributions have to be considered separately.
Measuring resistance of • built in thermometer • with resistance bridge: • calibration of bridge (type B) • drift of bridge (type B) • calibration of standard • resistor (type B) • drift of standard • resistor (type B) • Measuring • resistance of built • in thermometer • with multimeter: • calibration of • multimeter (type B) • drift of • multimeter (type B) Uncertainty of resistance measurement Measurement method for resistance measurement of dew-point and temperature
Sources of uncertainty of DUC • instability of DUC’s readings • resolution • uncertainty due to curve fit • hysteresis • drift (to be added by the end user) • uncertainty of electrical devices (bridge, resistor,...) • drift of electrical devices • the influence of DUC to calibration environment (self-heat, humidification – psychrometer)
Sources of uncertainty of DUC Uncertainty due to curve fit and hysteresis ufit,h = 0,25 %r.h.
Other sources of uncertainty (errors) • leakage (stem of the sensor, cables) • contamination (salts, oil, dirt) • different gas type used in calibration (nitrogen, air) • difference between calibration measurement range and actual range (temperature, pressure,...) • reproducibility