Top Ten Tips for Success

Top Ten Tips for Success. Continuous Glucose Monitoring. 1. Sensor Insertion. 60. . Ensure the Sensor is placed snugly into the SenSerter so that the o-rings are NOT visible The angle of Sensor insertion is crucial for appropriate Sensor penetration depth and “wetting”

rufus
Download Presentation

Top Ten Tips for Success

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Top Ten Tips for Success Continuous Glucose Monitoring

  2. 1 Sensor Insertion 60  • Ensure the Sensor is placed snugly into the SenSerter so that the o-rings are NOT visible • The angle of Sensor insertion is crucial for appropriate Sensor penetration depth and “wetting” • Insert the Sensor at a minimum of 60° by gently tilting the SenSerter forward 

  3. 2 Restrictive Clothing • Do not insert the Sensor in any region that will sit beneath restrictive or very tight clothing as this may alter interstitial tissue perfusion • Altered tissue perfusion may inhibit “wetting” of the Sensor which can lead to longer initialization times, lost sensor alarms and data gaps

  4. 3 Sensor Backing • Remove the transparent Sensor tape AFTER the Sensor has been placed in the SenSerter and BEFORE the Sensor is inserted • This will eliminate the need to remove the tape after Sensor insertion, and thus minimize pulling or dislodgement of the inserted Sensor

  5. 4 Taut Skin • Hold the skin taut at the Sensor site. Do not pinch the skinas doing so may prevent optimal Sensor penetration 5 Needle Removal • Hold the base of the Sensor firmly. Pull the needle out smoothly in one motion and at the SAME ANGLE as inserted • Do not ROCK the Sensor when pulling out the needle, as this may jeopardize Sensor penetration and may damage the Sensor electrodes 6 Bleeding • If bleeding occurs after Sensor insertion, apply gentle pressure for 3 minutes

  6. 7 Taping • After the MiniLink or iPro is connected to the Sensor, application of IV 3000 tape is highly recommended to: • Ensure the Sensor remains fully inserted underneath the skin • Maintain a tight Sensor to MiniLink or Sensor to iPro connection Reference electrode should sit beneath the skin inthe Subcutaneous Tissue Epidermis Dermis Subcutaneous Tissue Muscle

  7. 8 One Minute Rule • WAIT 1 MINUTE after disconnecting the MiniLink or iPro from a Charger, Sensor or Tester and before connecting the MiniLink or iPro to a Sensor or Tester • The MiniLink or iPro may be connected to a Charger at any time • Once the MiniLink or iPro is connected to a Charger, the MiniLink or iPro must remain in the Charger for a MINIMUM OF 1 MINUTE

  8. 9 MiniLink or iPro Flash • The MiniLink or iPro should flash green within 20 seconds AFTER it is connected to the Sensor • The flashing green light signals: • The Sensor is wet • There is enough power to last for a 3-day Sensor use Instruct patients to diligently check for the flashing green light

  9. 10 Tester Procedure STEP1 Confirm the Sensor isfullyinserted for at least 5 minutes to ensure adequate “wetting” of the Sensor Make sure the MiniLink or iPro has a tight connection to the Sensor.Disconnect and reconnect the Sensor from the MiniLink if needed(utilizing the One Minute Rule). STEP2 Make sure the MiniLink or iPro is charged: Disconnect the MiniLink or iPro from the Sensor. Connect the MiniLink or iPro to the Charger. Allow to fully charge. Disconnect the MiniLink iPro from the Charger. Wait 1 minute. Reconnect the MiniLink or iPro to the Sensor. If the MiniLink flashes, continue with Sensor set-up. If the MiniLink does not flash go to Step 4. STEP3 Tester Procedure: Will Differentiate between a Sensor or a MiniLink / iPro Issue Disconnect the MiniLink or iPro from the Sensor. Wait 1 minute. Connect the MiniLink or iPro to the Tester. The Tester replicates a functioning Sensor with ISIG signals. Go to MENU > Sensor Status Screen. If the ISIG is between 24 and 29 nA, the MiniLink is functioning properly, thus the Sensor needs to be replaced. If the ISIG is < 24 or > 29 nA, call the 24-Hour Help Line, as the MiniLink or iPro may need to be replaced. STEP4 If the MiniLink or iPro does NOT flash when connected to the Sensor: