Mammography Image Quality Optimization. Germán Tirao Grupo de Física Médica e Imágenes de Rayos X FaMAF – Universidad Nacional de Córdoba & CONICET. 2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010. Mammography Image Quality Optimization.
E N D
Mammography Image Quality Optimization Germán Tirao Grupo de Física Médica e Imágenes de Rayos X FaMAF – Universidad Nacional de Córdoba & CONICET
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization The image quality depends on: • Sample properties (breast tissue, kind of micro, breast size, calcifications, etc…) • Detector resolution • Set-up & Irradiation parameters (beam features and inserted devices) • Set-up & Irradiation parameters (beam features and inserted devices) The goal of mammography is the early detection of breast cancer. A mammography image is an special kind of breast radiography. OPTIMIZATION RX IMAGING TECHNIQUES Hoja 2
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Detector Colimator Object z X-ray source Measured Intensity Attenuation factor Mammography Image Quality Optimization CONVENTIONAL RADIOGRAPHY 2. 1st radiography 1895 CONTRAST SOURCE
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 2 mm Conventional X-ray source (Mo-Ka ≈ 17 keV) Conventional X-ray source(Cu-Ka ≈ 8 keV) Mammography Image Quality Optimization CONVENTIONAL RADIOGRAPHY • CONTRAST SOURCE: • ABSORPTION • X-ray energy • Sample thickness • Sample composition 2.
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization MONTE CARLO SIMULATION • CONTRAST SOURCE: • ABSORPTION • X-ray energy • Sample thickness • Sample composition 2. • Simulated images using Monte Carlo Techniques • Mathematical Processing to details detection
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization MONTE CARLO SIMULATION Sample properties • Micro calcifications (details) • Calcium Carbonate • Calcium Oxalate • Breast size • 30mm • 50mm • 70mm • Breast tissue • 100% Glandular • 50% Glandular and 50 % Adipose • 100% Adipose 2.
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization MONTE CARLO SIMULATION Irradiation parameters • X-Ray spectra (Mo anode) • Inserted devices • Anti-scatter grid 2.
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization MATHEMATICAL PROCESSING Quality evaluation & details detection • Central axis profile C and SNR • Image segmentation • Edge filters • Correlation
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization VIRTUAL IMAGE RESULTS Details Detec. & Quality Eval. • Central axis profiles Glandular 50-50 Adipose 34kV and 30mm * G. Tirao et al, Int. J. Low Radiation 7 (2010)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization VIRTUAL IMAGE RESULTS Details Detec. & Quality Eval. • Central axis profiles Calcium oxalate Calcium carbonate * G. Tirao et al, Int. J. Low Radiation 7 (2010)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization VIRTUAL IMAGE RESULTS Details Detec. & Quality Eval. • Central axis profiles 30mm 50mm 70mm * G. Tirao et al, Int. J. Low Radiation 7 (2010)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization VIRTUAL IMAGE RESULTS Details Detec. & Quality Eval. • Central axis profiles 24kV 34kV 40kV Breast absorbed dose [eV/(g*prim)] 92.17 123.54 136.27 * G. Tirao et al, Int. J. Low Radiation 7 (2010)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization VIRTUAL IMAGE RESULTS Details Detec. & Quality Eval. • Central axis profiles SNR value 9 5 8 6 1 7 3 2 4 * G. Tirao et al, Int. J. Low Radiation 7 (2010)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization VIRTUAL IMAGE RESULTS Details Detection • Image segmentation Tol1 Tol2 Tol3 * G. Tirao et al, Int. J. Low Radiation 7 (2010)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization VIRTUAL IMAGE RESULTS Details Detection • Image segmentation A B C Mammographic parameters: 100% glandular breast tissue with 30 mm thick and Mo 40 kV incident spectrum, calcium oxalate μCa composition. * G. Tirao et al, Int. J. Low Radiation 7 (2010)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization VIRTUAL IMAGE RESULTS Details Detection • Edge filters – Canny algorithm Different T1 and T2 thresholds 1- Smooth the image to eliminate noise. (Gaussian filter). 2- Find the image gradient to highlight regions with high spatial derivatives. 3- Nonmaximum suppression to become thinner the edge. 4- Reduces the gradient array by hysteresis to reduce the possibility of occurrence of false edge. T1 and T2 thresholds. 5- Close open boundary. (Deriche and Cocquerez algorithm). Mammographic parameters: 100% glandular breast tissue with 30 mm thick and Mo 40 kV incident spectrum, calcium oxalate μCa composition. * G. Tirao et al, Int. J. Low Radiation 7 (2010)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization VIRTUAL IMAGE RESULTS Details Detec. & Quality Eval. • Template Matching - Correlation Output Image Input Image Template Image Correlation Quality The matching process moves the template image to all possible positions in a larger source image and computes a numerical index that indicates how well the template matches the image in that position. * G. Tirao et al, Int. J. Low Radiation 7 (2010)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization SUMMARY • Sample properties • Breast size • Breast tissue • Calcification chemical composition • Irradiation parameters • Spectral spectrum composition • Harder spectrum generate better image quality • The dose grows meanwhile the quality increases • The application of a filter in the spectrum produce a lees dose • Mathematical Processing • Quality evaluation: Central axis profiles, Correlation • Details detection: Image segmentation, edge filters, Correlation • Strong dependence • Weak dependence • Low dependence • Strong dependence * G. Tirao et al, Int. J. Low Radiation 7 (2010)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization OPTIMIZATION RX IMAGING TECHNIQUES • Set-up & Irradiation parameters: X ray beam features Phase difference d Image acquisition techniques Attenuation CONTRAST SOURCE CONVENTIONAL UNCONVENTIONAL REFRACTION PHASE SCATTERING ABSORPTION object ABSORPTION
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 • PROPAGATION WAVEFRONT CHANGES TRANSVERSE COHERENCE reflectividad • INTERFEROMETRY LONGITUDINAL & TRANSVERSE COHERENCE OPTICAL PATH DIFFERENCE monocromator detector • DIFFRACTOMETRY R2 detector monocromator X-Rays X-Rays sample sample X-Rays analyzer Mammography Image Quality Optimization UNCONVENTIONAL X RAY IMAGING TECHNIQUES LONGITUDINAL COHERENCE REFRACTION
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 qe qR reflectividad qo qD qi detector • DIFFRACTOMETRY monocromator X-Rays sample analyzer Mammography Image Quality Optimization Imágenes de rayos X no convencionales para diagnóstico médico UNCONVENTIONAL X RAY IMAGING TECHNIQUES Ie IR Analyzer ID Io LONGITUDINAL COHERENCE Ii REFRACTION REFRACTION qR ≈ mrad * D. Chapman et al, Phys. Med. Biol. 42 (1997) DqD
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization Imágenes de rayos X no convencionales para diagnóstico médico UNCONVENTIONAL X RAY IMAGING TECHNIQUES Nylon immersed in methanol in a) maximum b) low-side reflectivity curve of the analyzer crystal. Eo = 10 keV. LNLS
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Dose: 0.8 mGy Dose: 0.5 mGy Conventional Mammography Unconventional Mammography DEI techniques Eo=17 keV Mammography Image Quality Optimization Imágenes de rayos X no convencionales para diagnóstico médico UNCONVENTIONAL X RAY IMAGING TECHNIQUES Perimeter: 259.2 m Energy: 2 / 2.4 GeV SYRMEP (Synchrotron Radiation for MEDicine Physics)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization Imágenes de rayos X no convencionales para diagnóstico médico UNCONVENTIONAL X RAY IMAGING TECHNIQUES • CONTRAST-ENHANCED DIGITAL MAMMOGRAPHY Dual Energy Subtraction Low Energy (24 - 32 keV) I K-edge (33.2 keV) High Energy (45 – 50 keV) 30 s Total X-ray dose delivered 20–50% higher * C. Dromain et al, Eur. J. Radiology 69 (2009)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization Imágenes de rayos X no convencionales para diagnóstico médico UNCONVENTIONAL X RAY IMAGING TECHNIQUES • CONTRAST-ENHANCED DIGITAL MAMMOGRAPHY Temporal subtraction technique High Energy (45 – 50 keV) - Light breast compression - Breast motion + Kinetic of tumor – Clasif. + Low radiation exposure * C. Dromain et al, Eur. J. Radiology 69 (2009)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization Imágenes de rayos X no convencionales para diagnóstico médico UNCONVENTIONAL X RAY IMAGING TECHNIQUES • CONTRAST-ENHANCED DIGITAL MAMMOGRAPHY Dual Energy Subtraction * C. Dromain et al, Eur. J. Radiology 69 (2009)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization Imágenes de rayos X no convencionales para diagnóstico médico UNCONVENTIONAL X RAY IMAGING TECHNIQUES • CONTRAST-ENHANCED DIGITAL MAMMOGRAPHY Temporal subtraction technique Ductal Carcinoma In Situ Invasive Ductal Carcinoma Kinetic curve of enhancement of breast lesion * C. Dromain et al, Eur. J. Radiology 69 (2009)
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 Mammography Image Quality Optimization Imágenes de rayos X no convencionales para diagnóstico médico CONCLUSIONS OPTIMIZATION RX IMAGING TECHNIQUES • Set-up & Irradiation parameters: X ray beam features Image acquisition Simulated images by Monte Carlo Techniques Relevant parameters Mathematical Processing Contrast enhancement, quality, details det.
Germán Tirao Grupo de Física Médica e Imágenes de Rayos X FaMAF – Universidad Nacional de Córdoba & CONICET Mammography Image Quality Optimization Thanks for your attention
2º Jornadas Física Médica en la Frontera. Temuco, Chile, Diciembre de 2010 • PROPAGATION WAVEFRONT CHANGES TRANSVERSE COHERENCE Aeration of the lungs of a newborn rabbit. (From left, fetus, after 12 breaths, then 1 hour). SPring 8, Japan (2005) R2 X-Rays sample Mammography Image Quality Optimization Imágenes de rayos X no convencionales para diagnóstico médico UNCONVENTIONAL X RAY IMAGING TECHNIQUES * S. W. Wilkins et al, Nature 384 (1996)