Adverse Event Reports for External Defibrillators 2005 – 2010

Adverse Event Reports for External Defibrillators 2005 – 2010. Roberta Sullivan, BSN, MPH Nurse Consultant, MDR Analyst Office of Surveillance and Biometrics Division of Postmarket Surveillance Product Evaluation Branch I. Introduction. Medical Device Report (MDR) Mandatory - Form 3500A

sahkyo
Download Presentation

Adverse Event Reports for External Defibrillators 2005 – 2010

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Adverse Event Reportsfor External Defibrillators 2005 – 2010 Roberta Sullivan, BSN, MPH Nurse Consultant, MDR Analyst Office of Surveillance and Biometrics Division of Postmarket Surveillance Product Evaluation Branch I AED/DPS-12/15/10-RXS

  2. Introduction • MedicalDevice Report (MDR) • Mandatory - Form 3500A • Voluntary MedWatch - Form 3500 • Manufacturer and User Facility Device Experience (MAUDE) • http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm • Or go to www.fda.gov -> navigate to “report a problem” online AED/DPS-12/15/10-RXS

  3. MDRs: Received by FDA from Jan 1, 2005 to Jan 1, 2010 • Original Product Code LDD: dc-defibrillator, low-energy, (including paddles) Class II, 510K cleared * ExternalDefibrillator, rhythm not diagnosed nor shock delivered automatically • Subsequent Product code MKJ: automated or semi-automated external defibrillators Class III, 510K cleared, used by lay public *Automatic functions, AED AED/DPS-12/15/10-RXS

  4. AED/DPS-12/15/10-RXS

  5. AED/DPS-12/15/10-RXS

  6. AED/DPS-12/15/10-RXS

  7. AED/DPS-12/15/10-RXS

  8. Patient problem codes • No patient involvement (59%) • No patient consequences (10%) • Unknown (9%) • Not applicable (5%) • No information (5%) • Death (4%) • No known impact to patient (3%) • Resuscitation (<1%) • Hospitalization required (<1%) • Cardiac Arrest (<1%) AED/DPS-12/15/10-RXS

  9. Patient problem codes • No patient involvement (59%) • No patient consequences (10%) • Unknown (9%) • Not applicable (5%) • No information (5%) • Death (4%) • No known impact to patient (3%) • Resuscitation (<1%) • Hospitalization required (<1%) • Cardiac Arrest (<1%) AED/DPS-12/15/10-RXS

  10. AED/DPS-12/15/10-RXS

  11. Device problem codes • Device displays error message (28%) • Failure to power up (15%) • Failure to charge, discharge or deliver (9%) • Loss of power (4%) • No display or display failure (4%) • Replace (4%) • Failure to run on AC/DC (3%) • Failure to shock or pace properly (3%) • Device issue (2%) • Device inoperable (2%) AED/DPS-12/15/10-RXS

  12. If returned for analysis, the evaluation identified device failure 1 time out of 3 (Component/subassembly failure, electrical/mechanical problem, design, contamination, manufacturing, software) • If returned for analysis, the conclusion was “device failure directly caused event or was related to event” ~40% of time AED/DPS-12/15/10-RXS

  13. What can MDR information be used for? • Provide a qualitative snapshot of adverse events for a specific device or device type • Signal detection • Unexpected events for a specific device type • Change in severity of expected events • User error/human factors issues • Regulatory actions • Premarket considerations • Public health communications AED/DPS-12/15/10-RXS

  14. Example 1 • “Complaint alleged that while attempting to cardiovert a pt (age and gender unknown), the device inappropriately shut down. Complainant indicated that the patient subsequently expired”. AED/DPS-12/15/10-RXS

  15. Example 2 • “Person requiring CPR, which was started by first responders. Nurse team arrived with AED, continued CPR, placed pads and hooked up to AED. Reading said "MEMORY FULL." CPR continued. Paramedics arrived with ambulance, took over CPR, intubated, inserted IV and transported to hospital. Placed on monitor en route, flatline. Pronounced upon arrival” AED/DPS-12/15/10-RXS

  16. AED/DPS-12/15/10-RXS

  17. Postmarket Surveillance • We need your reports for postmarket surveillance! • www.fda.gov for how to report and to access MAUDE MDR database AED/DPS-12/15/10-RXS