A Portable Device for Real-time Sample Flow Measurement Jerry Dulude, Glass Expansion, Inc. www.geicp.com
Why Monitor Uptake • Clogged nebulizer • Worn pump tubing • Kinked sample capillary • Worn pump rollers • Incorrect pressure of pump tubing clamp • Faulty peri pump
Heater Temp. Sensor A How it worksThermal Flow Measurement sample Temp. Sensor B Flow Rate is related to (TempA – TempB)
Features of TruFlo • Adjustable damping • Settable alarm limits • Digital display • Recordable graph • Range: 0.05 to 5ml/min
Pump rate vs flow rate & noise 70 rpm 60 rpm 50 rpm 40 rpm 30 rpm
Effect of Clamp Position ¼ turn tighter at each arrow
Applications of TruFlo • ICP-OES • ICP-MS • HPLC-MS Potential Applications of TruFlo • HPLC • UV-Vis • FTIR • Any technique where sample flow rate is important