An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Alignment systems for GlueX Courtesy of Dimos Sampsonidis, Aristotle University of Thessaloniki Presented by Zisis Papandreou, University of Regina GlueX Collaboration Meeting Bloomington, Indiana May 20-22, 2004

  2. NIKHEF Web Page Pattern used to measure x- and y-axis, size for z-axis Range of several cm with accuracy less than 1m

  3. Y X The infrastructure - RASNIK systems The main element used in chamber alignment and QA/QC during chamber construction •Three point relative alignment measurement systems (RASNIK) with accuracy ± 1μm. CCD camera 400000 pixels pixel size 7x7μm2 Mask and light source Lens adapted to layout Original system used a bright dot

  4. RASNIK specs Y X Infrared LED Performance, in terms of Image Position on Sensor: x = 0.05 µm y = 0.05 µm scale = 2 x 10-5 RotZ = 20 µrad Coded Mask Alignment information from image pattern recognition

  5. Online Quality Assurance Temperature variation Effect on RASNIK monitoring for tube layer position (z coordinate) Temperature variation at Stiffback The height of the clean room is limited (2,65 m). The stiffback is heated from the lights while it is suspended. Temperature variation at Rasnik Tower 6m shift of stiffback visible!

  6. What’s required • RasNik components: camera, mask, lens • PC with FrameGrabber: data translation • Rasmux: a custom multiplexer • For use with camera and mask • Suitable cables • Software: free from NIKHEF • AUTH will contact NIKHEF for a system or could participate in system development

  7. BCAM: another option • Boston CCD Angle Monitor • ATLAS forward muon chambers • Camera electronics is radiation resistant • 10 mrad field view • At 16m 3rad accuracy • Minimum effect of turbulence on resolution