An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. UW Engineering DOM Studies Kael Hanson University of Wisconsin – Madison LBNL DOM Workshop 23 July 2003

  2. What Testing was Done? • Guiding principle “Attempt to carry out measurements and see how DOM behaves as a data acquisition platform then follow our noses.” • Initial measurements desired: • Phototube gain • Peak-to-valley • Phototube noise (not yet /w/ DOM) • These measurements implied a priori characterization of DOM analog F.E. • Summary of measurements captured in Engineering DOM Tests document. LBNL DOM Workshop

  3. DOMs Under Test (DUTs) • UW now has 12 DOM-MBs: • 5 in ‘active’ DOMs (XE3P0001-0005) • 6 in DOMs awaiting testing • 1 DOMMB in use by NK for HV base tests • 2 DOMMBs returned to LBNL because of hardware problems: • 1 flash RAM failure • 1 CPLD failure (could not program) LBNL DOM Workshop

  4. DUT Configurations • Currently all 4 above DOMs running in chest freezer at PSL • XE3P0003 is in freezer at Chamberlin undergoing analog tests • Full DOM Work-in-Progress table maintained by Jim Hoffman LBNL DOM Workshop

  5. ATWD Pedestal Study • Purpose • Monitor system electronic noise • ATWD gain calibration • Procedure • Set ambient temperature to -20 ºC, … • Set FE bias to various levels • Acquire CPU-triggered ATWD shots LBNL DOM Workshop

  6. FE Gain – ATWD Channel 0 Clamping amplifier in ATWD ch0 introduces nonlinearity in analog FE! LBNL DOM Workshop

  7. FE Gain – ATWD Channel 1 No clamping amp – linear but ATWD overflows around 1 V – incorrect ATWD tuning parameters here or intentional behavior? LBNL DOM Workshop

  8. FE Pedestal/Gain Calibration Proposal • If you take delta between 5th and 1st curves on previous plot, you would naively expect a flat line. • Resulting line is not flat – features persist – do we need separate gain calibration for each ATWD bin? Is this subscript necessary? I LBNL DOM Workshop

  9. Pedestal Noise – ISEG DOM 718F PMT HV turned off! max(abs(wfd-ped)) LBNL DOM Workshop

  10. ATWD Pedestal Noise (2) • HV base turned off but ISEG base (718F) just disables HV – power not turned off to base. Disconnecting HV ribbon lowers noise somewhat. • Oddly enough, Ch1 shows about same level of noise – would expect 4x less scaling with op-amp gain. LBNL DOM Workshop

  11. Pedestal Noise – ISEG DOM 7DAC (ISEG base /w/ no ground wire) LBNL DOM Workshop

  12. Pedestal Noise – EMCO DOM 71DB LBNL DOM Workshop

  13. Floating Baseline? • Tried to test noise with HV on/off but run into problem of wobbling ATWD baseline and odd undershoot at pulse beginning (RHS). • (See next page) overall stability of pedestal on 2 hr timescale OK /w/ HV off but HV on causes large rms and drift. LBNL DOM Workshop

  14. ATWD Drift vs. Time NO HV HV ON Approx 2 hr. run time LBNL DOM Workshop

  15. Summary on Pedestal Testing • Significant noise in FE seems due to HV base (no noise seen in DOMMB tests?). Of options EMCO/ISEG, latter produces clearly visible hash in FE while former is clean. • Clamping amp nonlinearity significant. Can we tune this away or are we stuck with it? Do we really want to put a clamping amp on any more channels? • ATWD channel 1 overflows at about 1 V input (50 pe) – again, is this a tuning problem? • Is ATWD FE characterization simply a pedestal fingerprint or do we need more calibration knobs such as individual ATWD bin pedestals and gains. LBNL DOM Workshop

  16. Pulser Tests • Purpose • Calibrate op-amp gains • SPE/MPE discriminator studies • Procedure • Disable HV • Set internal pulser amp to 1/3 pe, 1 pe, … • Do discriminator sweep (FSCAN) • Acquire pulser waveforms LBNL DOM Workshop

  17. Pulser Discriminator LBNL DOM Workshop

  18. Pulser Waveforms • Unable to acquire pulser waveforms with synchronous pulser trigger – advertised in FPGA register level docs but either I called incorrectly or problem at FPGA • This type of trigger – synch with FE pulser, LED pulser, LED flasherboard required for some tests. LBNL DOM Workshop

  19. SPE Discriminator Scan – PMT Pulses Input (71DB) LBNL DOM Workshop

  20. SPE Waveforms CH0 CH2 CH1 LBNL DOM Workshop

  21. SPE Waveforms (2) • Presence of HV again causes drift of front-end baseline. • This may be due to ATWD overvoltage • ATWD channel 1 can saturate • ATWD channel 2 seems OK – small drift but probably functional • Why is ATWD channel 0, protected by clamping amplifier, having problems? • Current situation with drifting baselines makes analysis of ch0/ch1 waveforms very tricky – hope we can alter the clamping voltage to workaround – need information on this system. LBNL DOM Workshop

  22. Gain and P/V Measurement – DOM 0001-3C62-71BD LBNL DOM Workshop

  23. Cheap P/V & Gain Analysis • Used ATWD ch2 because of baseline wobble – gain of this channel is 2/3× so had to crank HV up to 1600, 1800, 2000 V. • Clear P/V appears at 2000 V (apx. 2.2:1) • Assuming ATWD gain is 1 mV/count, FE gain is 2/3×, the peak occurs around 40 mV – resulting in gain of about 5 × 107 – this just makes target! LBNL DOM Workshop

  24. Signal Frequency Analysis LBNL DOM Workshop

  25. Conclusions • Noise on ISEG HV base is potential trouble – analog tests indicate that level is less than 1 mV p-p but really need to understand characteristics after FE op-amps. • Several issues with ATWD-captured waveforms: • ‘startup’ bug: 1st sample after several seconds idle is corrupted – this is more annoying than problematic • Crosstalk from clock when mux’ed into channel 3 – OK once AG explained the problem and the fix. However, crosstalk from ch0 to ch1, ch2 potential problem. Do we test? • Clamping amplifier causes log rolloff of channel 0 • Saturated ATWD inputs cause sizable baseline drift • ATWD channel 0 gain seems a bit high – 16x implies approx 100 mV spe, about 100 ATWD counts. Could probably reduce gain by factor of 2, perhaps up to 4. LBNL DOM Workshop