1 / 21

Operational Modes

Operational Modes. Chapter 2 Stewart C. Bushong. Major Early Developments. Major early computed tomography developments were given the misnomer generation, as in genealogy Progress was rapid so that fourth-generation CT imagers appeared in 1978, just 6 years after the first CT imager

sarila
Download Presentation

Operational Modes

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Operational Modes Chapter 2 Stewart C. Bushong

  2. Major Early Developments • Major early computed tomography developments were given the misnomer generation, as in genealogy • Progress was rapid so that fourth-generation CT imagers appeared in 1978, just 6 years after the first CT imager • Unlike Hounsfield’s early experiments, the patient does not move during CT, except for spiral CT, rather, the x-ray source and the image receptor move

  3. First Generation • Finely collimated x-ray beam (pencil beam) was used in first-generation CT imagers • Fan-shaped x-ray beam (fan beam) is used in all current CT imagers • Single radiation detector • Translate-rotate motion

  4. First Generation • 180 translation with 1 degree between translations • Single image projection per translation • Single image projection per translation • Five minute image time • Head imager only, not capable of body imaging

  5. Second Generation • Fan-shaped x-ray beam • Multiple radiation detectors (detector array) • Translate-rotate motion • Usually 18 translations with 10 degree rotation between translations • Multiple image projections per translation

  6. Second Generation • Approximately, 30 s imaging time • Head and body imager

  7. Third Generation • A fan beam x-ray source is used and it views the entire patient during imaging • As many as several hundred radiation detectors are incorporated into the curvilinear detector array • The curvilinear detector array provides constant distance between source and each detector, resulting in good image reconstruction

  8. Third Generation • This development is based on 360 degree rotate-rotate motion. Both the x-ray source and the detector array rotate about the same axis • Hundreds of image projections are acquired during each rotation, resulting in better contrast resolution and spatial resolution

  9. Third Generation • Imaging time is reduced to 1s or less • Various arc scans are possible in order to improve motion blur-half scan, full scan • Ring artifacts are characteristic of third generation imagers

  10. Fourth Generation • Fourth generation was developed principally to suppress ring artifacts • The x-ray source is collimated to a fan beam as in third generation • The detector array can contain several thousand individual detectors

  11. Fourth Generation • The mechanical motion is rotation of the x-ray source around a fixed detector array (rotate-stationary) • There is a modest sacrifice in geometry; however, the un-attenuated leading edge and un attenuated trailing edge of the fan beam allows for individual detector calibration during each scan

  12. Fourth Generation • Patient dose may be somewhat higher with fourth-generation scanners because of interspace between detectors • When there is an interspace between detectors, some x-radiation falls on the interspace, resulting in a wasted dose • As the fan beam passes across each detector, an image projection is acquired

  13. Fourth Generation • Imaging time is 1s or less • Various arc scan are available – half scan, full scan, over scan

  14. Electron Beam CT (EBCT) • This CT imager was developed specifically for fast imaging • Images can be obtained in less than 100ms, about the time of a radiograph • The x-ray source is not an x-ray tube but rather a focused, steered, and microwave accelerated electron beam incident on a tungsten target

  15. EBCT • The target covers one-half of the imaging circle; the detector array covers the other half • The electron beam is steered along the curved tungsten target creating a moving source • There are four targets, or focal tracks, and four detector arrays, resulting in four contiguous images simultaneously

  16. EBCT • Electron beam CT is principally applied to cardiac imaging and frequently advertised as a heart scan • Electron beam CT has no moving parts • Electron beam CT uses a focused electron beam on a tungsten target ring as an x-ray source • Heat dissipation is no problem in EBCT

  17. EBCT • Electron beam CT can produce up to eight slices simultaneously • Electron beam CT scan times as short as 50ms are possible • Principal application for EBCT is cardiac imaging

  18. Spiral CT • Spiral CT was introduced to clinical practice in 1989 and is now the standard CT imager • If a third or fourth generation is CT imager is caused to continually rotate while the patient couch is moved through the imaging plane, spiral CT results

  19. Spiral CT • The development of slip rings was the technology breakthrough that made spiral CT possible • Spiral CT requires slip ring technology for data transfer from the rotating gantry • Spiral CT requires either an on-board high voltage supply so that coiled high voltage cables are unnecessary or slip rings for high voltage transfer

  20. Spiral CT • The principal advantage to spiral CT is the ability to image large volumes of anatomy in less time • Single breath-hold imaging of the entire torso is possible with spiral CT

  21. Comparison

More Related