Programming A-dec 500/300
Learning Objectives • As a result of this training you will be able to: • Identify an A-dec 500 touchpad from an A-dec 300 touchpad • Configure handpieces on an A-dec 500 and A-dec 300 unit
A-dec 500 Touchpads • Standard Touchpad, Assistant’s or Doctor’s • Deluxe Touchpad (2004 – 2006) • Endo Deluxe (May 2006)
Standard Touchpad – Pg 52 • System Status LED • Dental Light Control with 3-Way Switching • Manual Chair Positioning • Program Button • Presets • Cup Fill / Bowl Rinse • Only if a Cuspidor is installed • Auxiliary Buttons • A-dec relay functionality only if a Cuspidor is not installed
Chair Programming – Pg 53 • Position Programming • Position the chair to the desired operating position, using the manual controls (arrows) on the touchpad or foot switch. • Press and release Program button (one beep). • Within four seconds, press a position button (three beeps). • Position 3 (X-ray/Rinse) • Press and hold Program and Position 3at the same time for three seconds. One beep confirms that the button 3 is now fully programmable. Note: Disabling the x-ray/rinse function is a global function for Dr. A and B.
Cuspidor Functions – Pg 54 • Cuspidor Factory Presets • Cup fill – 2.5 seconds • Bowl rinse – 30 seconds
Handpiece Configuration 2 3 • Overview 1 0 4
Deluxe Touchpad – Pg 56 • Allows switching between Standard and Endo use of the EA51LT Electric Motor • Mode change: • Select with “E” button • 4 memory settings • M1=2,000 RPM • M2=10,000 RPM • M3=20,000 RPM • M4=40,000RPM
Deluxe Touchpad – Pg 57 • Dual Operator Support • Auxiliary Buttons • Cup Fill / Bowl Rinse • Only if a Cuspidor is installed • If a Cuspidor is not installed,buttons are inactive • Water Coolant / Air Coolant • Electric Handpiece / Ancillary Controls
Deluxe Touchpad Displays • Turbine selected • More than one of the“Big Three” selected • Scaler • Electric Handpiece • Pneumatic Handpiece • No Holdback Air • The power is on but there isn’t anyair to the unit
Touchpad Programming • Factory Preset Functions • Disable/Enable the Footswitch or Touchpad • Place the spare jumper in the “EN/DIS TP/FS” position of the test points header. • Push the buttons to be disabled/enabled (PRGM, PRGM 0, PRGM 1, PRGM 2, PRGM 3). • One beep means the button is disabled. • Three beeps means the button is enabled. • Place the spare jumper back in the spare position.
A-dec 300 Deluxe Touchpad • Programmable chair positions • X-ray/rinse button • Cupfill/bowl rinse • Auto Light • Electric handpiece operation • Endodontics
Thank you! • What questions do you have?