An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Programming A-dec 500/300

  2. Learning Objectives • As a result of this training you will be able to: • Identify an A-dec 500 touchpad from an A-dec 300 touchpad • Configure handpieces on an A-dec 500 and A-dec 300 unit

  3. A-dec 500 Touchpads • Standard Touchpad, Assistant’s or Doctor’s • Deluxe Touchpad (2004 – 2006) • Endo Deluxe (May 2006)

  4. Standard Touchpad – Pg 52 • System Status LED • Dental Light Control with 3-Way Switching • Manual Chair Positioning • Program Button • Presets • Cup Fill / Bowl Rinse • Only if a Cuspidor is installed • Auxiliary Buttons • A-dec relay functionality only if a Cuspidor is not installed

  5. Chair Programming – Pg 53 • Position Programming • Position the chair to the desired operating position, using the manual controls (arrows) on the touchpad or foot switch. • Press and release Program button (one beep). • Within four seconds, press a position button (three beeps). • Position 3 (X-ray/Rinse) • Press and hold Program and Position 3at the same time for three seconds. One beep confirms that the button 3 is now fully programmable. Note: Disabling the x-ray/rinse function is a global function for Dr. A and B.

  6. Cuspidor Functions – Pg 54 • Cuspidor Factory Presets • Cup fill – 2.5 seconds • Bowl rinse – 30 seconds

  7. Handpiece Configuration 2 3 • Overview 1 0 4

  8. Deluxe Touchpad – Pg 56 • Allows switching between Standard and Endo use of the EA51LT Electric Motor • Mode change: • Select with “E” button • 4 memory settings • M1=2,000 RPM • M2=10,000 RPM • M3=20,000 RPM • M4=40,000RPM

  9. Deluxe Touchpad – Pg 57 • Dual Operator Support • Auxiliary Buttons • Cup Fill / Bowl Rinse • Only if a Cuspidor is installed • If a Cuspidor is not installed,buttons are inactive • Water Coolant / Air Coolant • Electric Handpiece / Ancillary Controls

  10. Programming – Pg 58-60

  11. Deluxe Touchpad Displays • Turbine selected • More than one of the“Big Three” selected • Scaler • Electric Handpiece • Pneumatic Handpiece • No Holdback Air • The power is on but there isn’t anyair to the unit

  12. Touchpad Programming • Factory Preset Functions • Disable/Enable the Footswitch or Touchpad • Place the spare jumper in the “EN/DIS TP/FS” position of the test points header. • Push the buttons to be disabled/enabled (PRGM, PRGM 0, PRGM 1, PRGM 2, PRGM 3). • One beep means the button is disabled. • Three beeps means the button is enabled. • Place the spare jumper back in the spare position.

  13. A-dec 300 Deluxe Touchpad • Programmable chair positions • X-ray/rinse button • Cupfill/bowl rinse • Auto Light • Electric handpiece operation • Endodontics

  14. Deluxe Touchpad Setup

  15. Thank you! • What questions do you have?