A point of care diagnostic device for monitoring CD4 levels in HIV patients in a resource poor setting. Group 14: Lina Aboulmouna Peter DelNero Parker Gould Rosie Korman Chris Madison Stephen Schumacher Advisor: Dr. Kevin Seale, BME Vanderbilt University VIIBRE/ SyBBURE Nashville, TN.
E N D
A point of care diagnostic device for monitoring CD4 levels in HIV patients in a resource poor setting Group 14: LinaAboulmouna Peter DelNero Parker Gould Rosie Korman Chris Madison Stephen Schumacher Advisor: Dr. Kevin Seale, BME Vanderbilt University VIIBRE/SyBBURE Nashville, TN
Problem Statement • The Gates Foundation has identified low-cost HIV testing as a primary global health goal (“Grand Challenges in Global Health”). • Current HIV testing methods are slow and expensive • Flow cytometers are not suitable for the point-of-care needs for developing countries • ~$30,000-$150,000 per flow cytometer. • Limitations include energy scarcity, untrained technicians, high capital-cost equipment, patient proximity, low throughput and long feedback period
What is a CD4+ Lymphocyte? CD4 Antigen CD4 White blood cell CD4 Other surface markers Note: Image not to scale
CD4+ counts and Stage of HIV Infection • Patients with HIV who have CD4+ counts above 500cells/uL are in stage 1, CD4+ counts between 500cells/uL and 200cells/uL are in stage 2, and CD4+ counts of 200cells/uL and below are in stage 3 and are classified as having AIDS.
Performance Criteria • Prototype accurately determines CD4 count • Meets $2/test Gates Foundation challenge • Small sample volume (single finger-stick) • Generate results in minutes • Disposable and portable • Minimal energy requirements • Low technical expertise
Primary objective • Create a working prototype that accomplishes the specified goals and meets the performance criteria
Diagnostic Device Design Input PDMS Outputs Filter Device Glass Pump Antibodies Buffer Blood Mixer
Antibody conjugation protocol • Suspend antibodies and CMEUs (carboxylate-modified Europium nanoparticles) at 30 ugIgG/mg CMEU in the coating buffer: 10mM NaPO4 pH 8.0 • Allow the antibodies to coat the CMEUs for 1-2 hours, with gentle shaking. • After the coating, spin down the CMEUs, remove supernatant, and resuspend in blocking buffer: either 10 mg/ml BSA in buffer, or 5% PEG in buffer • Wash 2 or 3 times: Spin down the CMEUs at 10,000-12000g, remove supernatant, resuspend in blocking buffer • Spin down CMEUs, remove supernatant, resuspend in Conjugate Dilution Buffer
CD4 detection Eu NP ɑ-CD4 Antibody conjugated to Eu nanoparticle CD4 White blood cell CD4 Other surface markers Well Note: Image not to scale
Peristaltic Pump (hand-crankable) 1. Stepper Motor Alignment Screws Set Screw Shaft Coupler Fluid Tubing PDMS Washer Thrust Bearing PDMS Device Polycarbonate Base 2. 3. 4. 5. 6. 7. 8. 9.
Peristaltic Pump (hand-crankable) 1. 4. 2. 3. 5. 6. 7. 8. 9. Stepper Motor Alignment Screws Set Screw • Shaft Coupler • Fluid Tubing • PDMS Washer Thrust Bearing PDMS Device Polycarbonate Base
FilterArray White blood cell Red blood cell Silicon Pore Tunable pore size to selectively trap CD4+ cells
FilterArray White blood cell Red blood cell Silicon Pore Tunable pore size to selectively trap CD4+ cells
FilterArray White blood cell Red blood cell Silicon Pore Tunable pore size to selectively trap CD4+ cells
FilterArray White blood cell Red blood cell Silicon Pore Tunable pore size to selectively trap CD4+ cells
FilterArray White blood cell Red blood cell Silicon Pore Tunable pore size to selectively trap CD4+ cells
FilterArray White blood cell Red blood cell Silicon Pore Tunable pore size to selectively trap CD4+ cells
FilterArray White blood cell Red blood cell Silicon Pore Tunable pore size to selectively trap CD4+ cells
Input PDMS Filter Outputs Glass Filter with PDMS/Glass coverings
TRF (Time-Resolved Fluorescence) Imaging In a nutshell: • Precisely timing a xenon excitation flash and image capture with long lifetime fluorophores(europium nanoparticles). • Reduces the impact of background fluorescence.
TRF imaging with a cell trap device EuNP aggregate and water droplet Possibly CD4+ cells, possibly EuNP aggregates
Goals • Convert DC pump to hand-cranked mechanical power generator mechanism • Integrate microfluidic platform with camera and pumps • Separation of white and red cells in whole blood sample • Fluorescent conjugation, labeling, and excitation of anti-CD4 antibodies • Digital image acquisition • Digital image analysis • Accomplish the primary objective