270 likes | 412 Views
MX/MF MOISTURE ANALYZERS. MX-50. MF-50. Super Hybrid Sensor Technology RS232 Interface Easy to Read VFD Quick Reference Operation Guide Automatic Self-Checking Function. Clearly Marked Operation Panel Front Spirit Level 400W Halogen Lamp Easy and low cost to replace 5000 hrs. rating
E N D
MX-50 MF-50
Super Hybrid Sensor Technology RS232 Interface Easy to Read VFD Quick Reference Operation Guide Automatic Self-Checking Function Clearly Marked Operation Panel Front Spirit Level 400W Halogen Lamp Easy and low cost to replace 5000 hrs. rating Fast Heating 50-200C in < 2minutes DESIGN FEATURES
SUPER HYBRID SENSOR (SHS) • MFR-High Resolution • Extremely Fast • Less sensitive to temperature than previous MFR cells • Small and compact • Less expensive to produce than older MFR • Durable
Easy Maintenance • Removable Heater Cover
Low Cost Replacement Lamp Halogen Lamp Unit 400W 5000 Hours
STANDARD ACCESSORIES • Reusable Sample Pans (20) • Pan Handles (2) • Tweezers • Spoons • Test Standard (Sodium Tartrate Dihydrate-30 g) • Display Cover • Dust Cover • Data Cable (MX-50 only)
Washable Sample Trays Pan Handles
Easy to Use Operation Instruction Card
Four Measurement Programs Standard Automatic Timer Manual Program Memory MX-50 20 sets MF-50 10 sets Four Measurement Modes Wet Base Dry Base Solid content Ratio Results Memory MX-50 100 results MF-50 50 results GLP/GMP/ISO Conformity Software Communication Software Win-CT Moisture Standard with MX-50 Win-CT Standard with MF-50 (Win-CT Moisture optional) OPERATIONAL FEATURES
OPERATIONSTANDARD MODE • Operator sets accuracy, HI, MID, or LO, (see accuracy parameters on next slide. • Operator sets temperature, and moisture measurement unit. • Analyzer determines sample weight necessary based on accuracy required. • Auto stop at %/minute programmed.
Auto Stop Mode • Operator Sets: • Temperature, • % of Change, • Moisture Type & • Weight Resolution. • Auto Stop Occurs at %/min. Setting.
Setting 0.005% / min. 0.01% / min. 0.02% / min. 0.05% / min. 0.1% / min. 0.2% / min. 0.5% / min. 1.0% / min. 2.0% / min. MX-50MF-50 - - Accuracy Settings for AUTO Mode
TIMER Mode • Operator Sets: • Temperature (between 50 C to 200 C), • Timer (between 1 – 480 Minutes), • Moisture Type & • Weight Resolution. • Test Auto Completed when Timer Expires
MANUAL MODE • Operator Sets: • Temperature • Moisture Type & • Weight Resolution. • Measurement is Stopped Manually at Any Time.
WinCT-Moisture SOFTWARE • Windows software for MX-50 • Optional for MF-50 • Shows Drying Curve on the Display • Determines The Best Temperature for Moisture Measurement
Green 100 C Light Blue 120 C Red 140 C purple 160 C yellow 180 C Blue 200 C
“OTS”Optimal Temperature Setting Allow the Moisture Analyzer to determine the best test temperature and heating time.
Average Std Dev CV Max-Min MX-50 15.72 0.015 0.095 0.04 HB43 15.79 0.058 0.367 0.14 MA30 15.76 0.053 0.336 0.13 Performance Comparison Testing Sodium Tartrate Dihydrate (5g, 160 C) Ideal Value = 15.66%
Test Results…..A&D Outperforms the Higher Priced Competition ! ! !
NEW A&D MX/MF MOISTURE ANALYSER - SUMMARY • Perform better than competitors • Easier to use and set-up than competition • Modern, updated design and styling • Durable, reliable, and cost less to maintain • Priced significantly lower than competition • Which represents… …Clearly a Better Value