Dissolution. Mrs. Witinee Kongsuk Bureau of Drug and Narcotic July 9, 2015. Dissolution. Apparatus 1 Basket. Vessel - Glass or other inert transparent material - Cylindrical with hemispherical bottom - 1L, 2L or 4L nominal capacity Stirring shaft and basket Basket
E N D
Dissolution Mrs. Witinee Kongsuk Bureau of Drug and Narcotic July 9, 2015 Bureau of Drug and Narcotic
Dissolution Bureau of Drug and Narcotic
Apparatus 1 Basket • Vessel • - Glass or other inert transparent material • - Cylindrical with hemispherical bottom • - 1L, 2L or 4L nominal capacity • Stirring shaft and basket • Basket • - 40 mesh • - Gold coating of 2.5 um is allowed. • Stirring • - rotating stirrer • - typical speeds: 50-100 rpm • Dosage form placed in the basket Bureau of Drug and Narcotic
Apparatus 2 Paddle • Vessel • - same as for Apparatus 1 • Stirring blade and shaft • - metallic or suitably inert and rigid • - may be coated • Agitation • - rotating stirrer • - typical speeds: 50-75 rpm • Dosage form should remain • at the bottom center of the vessel Bureau of Drug and Narcotic
Dissolution Assembly Water bath Bureau of Drug and Narcotic
Visual Inspection Water bath • Check for leaks: tubing, joints • Circulation pump: check for leaks, trapped air, noise • Good practice to regularly change water • Use clean and clear water • Keep bath clean • Ensure adequate water level Bureau of Drug and Narcotic
Dissolution Assembly Vessel Bureau of Drug and Narcotic
Visual Inspection Vessel • Thorough cleaning is very important. • Sticky residues can build up on glass surface. • Always check the internal surface • Irregular curvature can affect hydrodynamics. • Check periodically during use looking for cracks, • scratches and build up of deposit or residue. Bureau of Drug and Narcotic
Dissolution Assembly Bureau of Drug and Narcotic
Dissolution Assembly Apparatus 1 Basket Bureau of Drug and Narcotic
Dissolution Assembly Apparatus 2 Paddle Bureau of Drug and Narcotic
Visual Inspection Baskets and Paddles • It is very important to carefully observe • the physical condition. • Shaft for both apparatus must be straight. • Loose basket clips will cause wobble. • Basket should be right cylinders (not bent). • Baskets that are old, distorted or frayed can give • non-reproducible results. • Baskets should be properly cleaned and stored. • Ensure vent hole in basket head is clear. • Baskets should be attached carefully to the shaft. • Paddles: smooth coating, if coated Bureau of Drug and Narcotic
Dissolution Assembly Canular TheUSPdefines the sampling position as “A zone midway between the surface of the dissolution medium and the top of the rotating basket or blade, not less than 1 cm from the vessel wall" Bureau of Drug and Narcotic
Other related tools • Calibrated thermometer • Calibrated stopwatch Bureau of Drug and Narcotic
Calibration Bureau of Drug and Narcotic
Mechanical Calibration • Conformance to USP specification • Position designation in the assembly • - Stirring elements • * Shafts • * Blade or baskets • - Vessels • Recommendation: • Maintain the same position for vessels / stirring • elements within the assembly throughout a PVT cycle Bureau of Drug and Narcotic
Mechanical Calibration Shaft verticality Vessel verticality Bureau of Drug and Narcotic
Mechanical Calibration Centering Wobble Bureau of Drug and Narcotic