1 / 8

UHV Cleanliness Requirements: Cleaning/Baking and Contamination Control

UHV Cleanliness Requirements: Cleaning/Baking and Contamination Control. Dennis Coyne HAM-SAS Bidder’s Conference @ LIGO Laboratory, Caltech 25 October 2005. Livingston, LA. Hanford, WA. Basic UHV Compatibility Requirements. Cleanliness Requirements:

stian
Download Presentation

UHV Cleanliness Requirements: Cleaning/Baking and Contamination Control

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. UHV Cleanliness Requirements:Cleaning/Baking and Contamination Control Dennis Coyne HAM-SAS Bidder’s Conference @ LIGO Laboratory, Caltech 25 October 2005 Livingston, LA Hanford, WA LIGO R&D

  2. Basic UHV Compatibility Requirements • Cleanliness Requirements: • Achieve a Non-Volatile Residue (NVR) level of A/50 or better (per MIL-STD-1246C or IEST-STD-CC1246D) on piece parts before clean assembly • Achieve a Particulate Cleanliness Level of 50 or better (per MIL-STD -1246C or IEST-STD-CC1246D) on piece parts before clean assembly • Perform clean assembly and all steps through UHV Packaging in a Class 100 Clean Room (per FED-STD-209) • Comply with Contamination Control Protocols defined in E048228 (M990034 is also a good guide) • Package the Clean Parts per E048227 • No Virtual Leaks • Full Penetration Welds • No Cold Welding • Stainless fasteners into Aluminum • Silver coated Stainless Fasteners into Stainless • Oversized pilot and tap for tapped holes LIGO R&D

  3. Ultra-High Vacuum (UHV) Cleanliness Requirements Related to Manufacturing Processes Notes: The process sequence depicted here is suggestive, and not meant to be prescriptive. Loaded GAS Filter, Spring Blades to be Creep Relaxed in Air Bake LIGO R&D

  4. Practical UHV Requirements • Must pre-clean parts before welding and winding • Must clean parts multiple times during fabrication • Must thoroughly clean parts! • “gross” clean with solvents and acids • Mechanical clean (scrub) parts especially tapped holes – multiple times, with powered rotary tool – during “gross” and “precision” clean steps • “precision” clean with hot, water soluble detergents (steam or hot water spray) – multiple times • Never let parts dry with detergent • Multiple rinses with Deionized water • Must bake parts (air or vacuum) after “precision” cleaning • Must maintain parts in a class 100 clean room • or keep wrapped in “UHV Foil” and keep in at least a class 10,000 clean area) • limit exposure, keep wrapped /covered when possible • Must Control Contamination with Good Clean Room Practice • All tools must be cleaned, air baked, wrapped & kept in Class 100 areas • All personnel must wear full clean room garb, including hair net, face mask and gloves LIGO R&D

  5. UHV Requirement References • Particulate Level 50 and NVR Level A/50 per MIL-STD-1246C or IEST-STD-CC1246D • Cleaning for UHV Service: • Contractor is to certify the cleaning and baking process with representative parts to demonstrate 1246 Level 50-A/50 • NVR Level to be confirmed via FTIR (Fourier Transform Infrared analysis) • Particulate level to be confirmed by particle sizing and counting or equivalent particle area coverage • LIGO Lab will independently assess the contractor’s cleaning/baking process with representative parts by RGA (mass spectrometer outgassing analysis) per E960022 • Guidance: E048227, Process Specification for Ad-LIGO Seismic Isolation System In-Vacuum Mechanical Elements: Fabrication (includes cleaning) • Contamination Control: • Per E048228, Process Specification for Ad-LIGO Seismic Isolation System In-Vacuum Mechanical Elements: Contamination Control • Additional Guidance: • M990034, LIGO Contamination Control Plan • FED-STD-209, Airborne Particulate Cleanliness Classes in Cleanrooms and Clean Zones • ISO-14644, Cleanrooms and controlled environments[ISO-14644 superceeds FED-STD-209] • Packaging: • Per E048227, Process Specification for Ad-LIGO Seismic Isolation System In-Vacuum Mechanical Elements: UHV Packaging LIGO R&D

  6. Class 100 Clean Room Class 10,000 Room HEPA Filter Gowning Room “Sticky” Mat Clean Room & Gowning Room LIGO R&D

  7. HAM Chamber Clean Rooms “Full” Clean Room Dress Concrete Floor – Cleaned Routinely, not ideal, but serviceable LIGO R&D

  8. UHV Foil Annealed (baked) UHV Aluminum Foil for Wrapping Parts LIGO R&D

More Related