Gammex Boxes in a line scanned together These ROIs propagated through the entire reconstructed volume. Mean of ROIs. Box causes shift in air CT numbers – about 2 HU. Observations. CT Numbers of foam vary from Gammex box to Gammex box Range 15 HU for 320 kg/m 3 foam
E N D
Gammex Boxes in a line scanned together These ROIs propagated through the entire reconstructed volume
Observations • CT Numbers of foam vary from Gammex box to Gammex box • Range 15 HU for 320 kg/m3 foam • Use of Gammex boxes as reference standards will require calibration of each box • Box alters air CT number above the box