An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. KEELE UNIVERSITY WORKING WITH X-RAYS Dr D E Dugdale October 2012

  2. ROTATING ANODE GENERATOR

  3. X-RAY SPECTRUM Emax = eV

  4. DOSE RATES • Can be very high • For anode of atomic number Z, operating voltage V, current I •  Unfiltered dose rate = D  ZVI/d2 • at a distance d • Typical figures for V = 50 KeV, I = 10 mA • @ d = 1 m D - 1 Sv per minute • @ d = 10 cm D - 100 Sv per minute

  5. Radiation burns ( 3 Sv ) in a time of • @ 1m ~ 3 mins. • @ 10 cm ~ 2 secs. • Using Ni filter or monochromator D  D/100 • To have radiation level below maximum allowed ( 7.5 micro Sv/hr ) requires shielding of > 108

  6. X-RAY SHIELDING I = Io e- x =linear attenuation coefficient H1/2 = 0.69/

  7.        For 50 keV X-rays H1/2 - 10 m in air H1/2 - 3 cm in body tissue H1/2 - 0.1 mm in Pb         For Pb @ 50 keV X-ray energy I/Io = 1/10 for x = tenth value thickness = 0.23 mm For x = 2.3 mm therefore I/Io = 10-10

  8. X-RAY LEAKAGE - Typical Arrangement Of X-ray Generator And Camera Leakage at joints must be less than 7.5 mSv/hr - but ALARA principle demands much less since extra shielding is reasonably achievable for X-rays

  9. Perspex Shielded Enclosures Leaded Perspex outer shield

  10. MINI MONITORS • Soft X-rays ( < 30 keV ) do not easily penetrate Geiger -Muller ( GM )tube walls. • Ideally tubes with Be end windows should be used - but even cheap general purpose GM tubes have some response down to 9 keV ( Cu K lines used in crystallography) • For GM tube MX-168 @ 10 keV • 1 c/s ~ 1 micro Sv/hr ( background ~ 0.5 c/s ) • 1 micro Sv/hr is only ~ 2 mSv/yr but nevertheless more shielding should be used • if any count rate above background can be detected

  11. PERSONAL MONITORS • Film badges and thermoluminesent detectors can detect doses > 0.05 mSv (~ 1/40 annual background) • Film badges are changed every month ( every 3 months for TLD’s) • Not likely to detect accidental exposure to collimated beams but should detect any diffuse scattered radiation • Wear at chest height with serial number facing outward

  12. WORKING PROCEDURES • Register each year as a radiation worker • (so that dose and training records can be kept ) • Follow the general rules of the University and the local rules of the Department for specific equipment and specific procedures • Local rules and local training are given by research supervisors and RPS • Never overide interlocks • Never operate if warning lights fail or if equipment is malfunctioning • Report all faults to the responsible technician immediately