An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Ex-situ NMR LDRDProgress report 01/27/03 Magnetic and mechanic design Paolo Ferracin

  2. Field calculation for a line current • 2D azimuthal component of the field: • Only odd normal (b3, b5, b7,…) and even skew (a2, a4, a6,…) harmonics. y - I + I a p (r, J) F x Paolo Ferracin

  3. y x d t + I - I Line currents in NMR configuration • Distance from the sweet spot: d. • Distance from the y axis: t. • We move the line currents horizontally. • We compute, for a given d and t, the field harmonics. Paolo Ferracin

  4. x d t - I + I Main field y Paolo Ferracin

  5. x d t - I + I Skew quadrupole y Paolo Ferracin

  6. x d t - I + I Normal sextupole y Paolo Ferracin

  7. Comments • Comparison of the curves with Roxie: • line currents: perfect agreement; • blocks ( 20  20 mm) with d  30 mm: very good agreement; • You can cancel either a2 or b3 with 2 horizontal coils (with the same current): •     • You can cancel both a2 and b3 with 3 horizontal coils (with the same current): •       Paolo Ferracin

  8. 2D analysis: HD and RD HD RD • NbTi cables. • Distance d ~ 40 mm, total width 200 mm. • Main field: 2000  3000 Gauss. • Field homogeneity (without full optim.): R = 5 mm  10-3; R = 1.5 mm  10-4  10-5. Paolo Ferracin

  9. 2D analysis: SM coils, upper part • Nb3Sn cables. • Distance d ~ 40 mm, total width 200 mm. • Main field: ~ 6000 Gauss. • Field homogeneity: R = 5 mm  10-3; R = 1.5 mm  10-4  10-5. Paolo Ferracin

  10. 2D analysis: SM coils • Nb3Sn cables. • Distance d ~ 40 mm, total width 200 mm. • Main field: 2000  3000 Gauss (return current effect). • Field homogeneity: R = 5 mm  10-3; R = 1.5 mm  10-4  10-5. Paolo Ferracin

  11. 3D analysis: SM coils • Main field: 2000  3000 Gauss. • Field quality: • strong 3D effect of the ends. • Cube of 3 mm side: • By: 10-4 homogeneity; • Bx and Bz ~ 0.01 By. • Work in progress: • coil geometry optimization; • effect of the iron (2D performed). Paolo Ferracin

  12. 3D analysis: SM coils 10-4 10-4 Byy Byx Bz Bx Paolo Ferracin

  13. Mechanical design: Lorentz forces • Fx (internal coil) = - 0.3 MN. • Fx (two external coil) = + 0.5 MN. • Fx in the SM = + 0.6 MN. • Assuming: • shell 20 mm thick (upper limit) and 100 mm long; • shell azimuthal stress at cold: ~ 100 MPa (cool-down)  Fx on the coils = ~ 0.4 MN. Paolo Ferracin