Activity monitor accuracy in persons using canes

Activity monitor accuracy in persons using canes. Deborah Michael Wendland, PT, DPT, CPed; Stephen H. Sprigle, PhD, PT. Aim Report accuracy of leg-mounted and cane-mounted StepWatch devices on people ambulating over different surfaces while using a cane. Relevance

tana
Download Presentation

Activity monitor accuracy in persons using canes

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Activity monitor accuracy in persons using canes Deborah Michael Wendland, PT, DPT, CPed; Stephen H. Sprigle, PhD, PT

  2. Aim • Report accuracy of leg-mounted and cane-mounted StepWatch devices on people ambulating over different surfaces while using a cane. • Relevance • Physical activity is important for health and associated with decreased all-cause mortality. • Assistive device use has increased to facilitate increased independence with activities of daily living.

  3. Method • 16 subjects (67 to 85 yr old) who regularly use a cane for ambulation participated. • StepWatch calibration was performed by adjusting sensitivity and cadence. • Following calibration optimization, accuracy was tested on: • Leg-mounted vs cane-mounted devices. • Different surfaces, including linoleum, sidewalk, grass, ramp, and stairs.

  4. Results • Accuracy: • Leg-mounted:93.4% all surfaces. • Cane-mounted:84.7% all surfaces. • Significantly less accurate on stairs. Gray bars = leg-mounted. Black bars = cane-mounted.

  5. Conclusion • When monitoring community mobility, placement of StepWatch on a person and his or her ambulation aid can accurately document both activity and device use.