Fluxgate Magnetometer (FGM) Instrument Status Rumi Nakamura, Werner Magnes

Fluxgate Magnetometer (FGM) Instrument Status Rumi Nakamura, Werner Magnes Institut fuer Weltraumforschung (IWF) Austrian Academy of Sciences Hans-Ulrich Auster Institut fuer Geophysik und Meteorologie Technical University Braunschweig (TU-BS). FGM Overview. Status F1..F6

tanith
Download Presentation

Fluxgate Magnetometer (FGM) Instrument Status Rumi Nakamura, Werner Magnes

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Fluxgate Magnetometer (FGM) Instrument Status Rumi Nakamura, Werner Magnes Institut fuer Weltraumforschung (IWF) Austrian Academy of Sciences Hans-Ulrich Auster Institut fuer Geophysik und Meteorologie Technical University Braunschweig (TU-BS)

  2. FGM Overview • Status F1..F6 • 3 ETU’s available for tests • F4 & F6 test and calibration completed • F1, F2, F3, F5 manufacturing completed, test and calibration finished in February 2005 • Compliance with MRD requirements • See FGM-3 …FGM-6 • Test schedule • See FGM -7

  3. Dependency on electronics temp. without feedback 0.01nT/°C ! with feedback 0.1nT/°C ? acceptable, but could be improved by replacing AD648 by LT1013 MRD requirements • The absolute stability of the FGM shall be less than 1nT Dependency on sensor temp. <0.01nT/°C ! temperature

  4. MRD requirements • The relative stability of the FGM shall be less than 0.2nT/12hrs Measurement of horizontal Earth field by F4 and F6. Over 15 hours no significant difference visible

  5. MRD requirements • The FGM noise level @ 1Hz shall be less than 0.03nT/Sqrt(Hz) 0.03nT/Sqrt(Hz) 0.01nT/Sqrt(Hz)

  6. MRD requirements • The FGM digital resolution shall be less than 0.1nT • The FGM science range shall exceed 0-1000nT Test shows Earth field variation with a resolution better 0.1nT @ a back ground field of 12000nT 1nT

  7. FGM Test Schedule • December 2004 • exchange of AD648 by LT1013 (all boards) • start F1 & F2 test in Europe • January 2005 • repetition of electronics temperature test F4 & F6 in Berkeley • finish F1 & F2 test in Europe • February 2005 • start F3 & F5 test in Europe • finish F3 & F5 test in Europe