1 / 48

CT Image Quality & Quality Control

CT. CT Image Quality & Quality Control. Spatial Resolution. Quantifies image blurring “Ability to discriminate objects of varying density a small distance apart against a uniform background” Minimum separation required between two high contrast objects for them to be resolved as two objects.

tbraswell
Download Presentation

CT Image Quality & Quality Control

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. CT CT Image Quality & Quality Control

  2. Spatial Resolution • Quantifies image blurring • “Ability to discriminate objects of varying density a small distance apart against a uniform background” • Minimum separation required between two high contrast objects for them to be resolved as two objects

  3. Spatial Resolution

  4. Geometric Factors affectingSpatial Resolution Finite focalspot size focal spot - detector distance focal spot - isocenter distance

  5. Focal Spot Size & Resolution The larger the focal spot the more it will blur a tiny place on the patient.

  6. Detector Size & Resolution The shadow of both objects falls on the same detector. Detector

  7. Focal Spot Size & Resolution Combination of finite focal spot and object position blurs object across multiple detectors. Detectors

  8. Resolvable Object Size &Limiting Resolution • Smallest resolvable high contrast object • Often expressed as line pairs / cm • “Pair” is one object + one space One Pair

  9. Resolvable Object Size:Limiting Resolution • Smallest resolvable high contrast object is half the reciprocal of spatial frequency • Example: • Limited resolution = 15 line pairs per cm • Pair is 1/15th cm • Object is half of pair • 1/15th / 2 • 1/30th cm • .033 cm • 0.33 mm 1/15th cm 1/30th cm

  10. Geometric Factors affectingSpatial Resolution • In-slice • Focal spot size • detector aperture width • Z-axis • slice thickness or collimation • Less variation likely for thinner slices • attenuation variations within a voxel are averaged • partial volume effect

  11. High Contrast (or Limiting) Resolution • Phantom • rows of equally spaced high contrast objects • holes get smaller from row to row • Bars of decreasing size & separation • Determine smallest row of holes where all holes can be clearly seen • smaller holes = better resolution • Visual scoring

  12. Contrast Resolution • Ability of an imaging system to demonstrate small changes in tissue contrast • The difference in contrast necessary to resolve 2 large areas in image as separate structures

  13. CT Contrast Resolution • Significantly better than radiography • CT can demonstrate very small differences in density and atomic # This’ll be on your test. I guarantee it. Contrast Resolution Radiography10% CT<1%

  14. CT Contrast Resolution Depends Upon • Dose • reconstruction algorithm • low spatial frequency algorithm smooths image • Loss of spatial resolution • Reduces noise • enhances perceptibility of low contrast lesions • image display

  15. CT Contrast Resolution Depends on Noise

  16. CT Contrast Resolution Contrast depends on noise Noise depends on # photons detected # photons detected depends on …

  17. # of Photons Detected Depends Upon • photon flux (x-ray technique) • slice thickness • patient size • Detector efficiency • Note: • Good contrast resolution requires that detector sensitivity be capable of discriminating small differences in intensity

  18. Noise Measurement in CT • Scan water phantom • Select regions of interest (ROI) • Take standard deviation in each region • Standard deviation measures noise in ROI

  19. CT Image Quality inEquation Form s2(m) = kT/(td3R) Where s is variance resulting from noise k is a conversion factor (constant) T is transmissivity (inverse of attenuation) t is slice thickness d is pixel size R is dose

  20. Slice Thickness • Thinner slices mean • less active detector area • less photons detected • More noise • To achieve equivalent noise with thinner slices, dose (technique factors) must be increased

  21. Noise Levels in CT: • Increasing slice width = less noiseBUT • Increasing slice width degrades spatial resolution • less uniformity inside a larger pixel • partial volume effect

  22. CT Artifacts:Partial Volume Effect • CT #’s based on linear attenuation coefficient for tissue voxels • If voxel non-uniform (contains several materials), detection process will average

  23. Partial Volume Effect • Can appear as • incorrect densities • streaks • bands • Minimizing • Use thinner slices

  24. Noise Levels in CT: • When dose increases, noise decreases • dose increases # detected photons • Doubling spatial resolution (2X lp/mm) requires an 8X increase in dose for equivalent noise • Smaller voxels mean less radiation per voxel

  25. CT Image Quality Trade-off s2(m) = kT/(td3R) To hold noise constant • If slice thickness goes down by 2 • Dose must go up by 2

  26. Contrast Response FunctionCTF or CRF • Blurring causes loss of contrast • darks get lighter • lights get darker Lower Frequency Higher Frequency Higher Contrast Lower Contrast

  27. CT Spatial vs. Contrast Resolution • Spatial & contrast resolution interact • High contrast objects are easier to resolve • Improve one at the expense of the other • Can only improve both by increasing dose Increasing object size Increasing contrast

  28. Contrast & Detail • Larger objects easy to see even at low contrast Increasing object size Increasing contrast

  29. Contrast & Detail • Small objects only visible at high contrast Increasing object size Increasing contrast

  30. CT Number • Calculated from reconstructed pixel attenuation coefficient (mt - mW) CT # = 1000 X ------------ mW Where: ut = linear attenuation coefficient for tissue in pixel uW = linear attenuation coefficient for water

  31. CT # of Water • What do we check? • Average (ideally 0 by definition) • Standard Deviation (ideally 0) • Flatness • ROI’s at various locations in phantom • Results depend upon protocol • kVp, mA, scan time • slice width • phantom size • ROI position • Problem causes • Mis-calibration • Noise / drift in detector system

  32. Linearity • Relationship of CT #’s to object attenuation coefficients • Checked with phantom of several known materials • average CT # of each material obtained from ROI analysis • Compare CT #’s with known coefficients -100 77 325 -44 50

  33. Preview!CT Artifacts: Causes • motion • metal & high-contrast sharp edges • beam hardening • partial volume averaging • detectors

  34. Image Artifacts:Ring Artifact in 3rd Generation • Causes • 1 or more bad detectors • small offset or gain difference of 1 detector compared to neighbors • detector calibration required • Reason: rays measured by a given detector are all tangent to same circle

  35. Monitor Performance / Hard Copy Output • Standard gray scale image needed • computer generated (SMPTE pattern) • Check monitor & hard copy • 5% patches • Contrast patches • Resolution bars • Hard copy • Measure optical density of steps with densitometer

  36. Table Backlash • Does table return to same position from both directions? • Test • Apply masking tape to table • Load table with weight • Note numerical position • Mark tape at starting position using laser position • Move table away from starting position & then back to same numerical location • Check laser vs. mark on tape

  37. Laser Accuracy • Align phantom on table matching outer laser to external marks on phantom • “Zero” table • Scan phantom • “0” z-location should show correct phantom position on image • Repeat for internal laser

  38. American College of Radiology (ACR) Accreditation Program for CT • CT Accreditation Program involves acquisition of • clinical images • phantom images • dose measurements • submission of scanning protocols. • Providers billing under part B of Medicare Physician Fee Schedule must be accredited to receive technical component reimbursement from Medicare. $$$ $$$

  39. ACR Phantom • Positioning accuracy • CT # accuracy • Image thickness • Low contrast resolution • High contrast (spatial) resolution • CT number uniformity • Image noise

  40. Phantom Section #1 • Alignment • CT # Linearity • Slice Width

  41. ACR Phantom Section #2 • Low Contrast Resolution

  42. ACR Phantom Section #3 • Uniformity & Noise • Distance Accuracy 100 mm

  43. ACR Phantom Secion #4 • High Contrast Resolution

  44. ACR Phantom Criteria • Phantom imaged with clinical protocols • Measurements • CTDTvol • Low contrast performance • Contrast to Noise Ratio

More Related