7526A Precision Temperature/Pressure Process Calibrator

7526A Precision Temperature/Pressure Process Calibrator. Introducing the 7526A Precision Process Calibrator. Value proposition Typical applications Key features & benefits Product overview Product features Summary specs Options & accessories.

thad
Download Presentation

7526A Precision Temperature/Pressure Process Calibrator

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. 7526A Precision Temperature/Pressure Process Calibrator

  2. Introducing the7526A Precision Process Calibrator • Value proposition • Typical applications • Key features & benefits • Product overview • Product features • Summary specs • Options & accessories A “best fit” for process instrumentation calibration

  3. 7526A Value Proposition • Balances precision and versatility with cost • The 7526A is less expensive than high-end multi-product calibrators, yet more precise and versatile than handheld field calibrators • Calibrates wide ranging workload 7526A Versatility, precision & value The 7526A offers an isolated measurement channel allowing it to precisely, and simultaneously, source and measure temperature & electrical parameters. Pressure measurement can be viewed in both displays in different units of measure.

  4. Typical Applications

  5. 7526A Key Features & Benefits

  6. 7526A Product Overview • Precision Temperature/Pressure • Process Calibrator • Sources and measures: • Thermocouples (13 types) • RTDs (9 types) • Thermistors (YSI-400) • Voltage (0 to 100 Vdc) • Resistance (0 Ω to 4 kΩ) • Current (100 mA output / 50 mA input) • Measure only: • Pressure (0 to 700 bar / 10,000 psi)

  7. 7526A Product Overview (Left Half) DC voltage output terminals 0 Vdc to 100 Vdc Accuracy: 30 ppm (+3 μV) DC current output terminals 0 mA to 100 mA Accuracy: 50 ppm RTD/Ω output terminals (two-wire) 5 Ω to 4 kΩ, Accuracy: ± 0.05 ºC, –200 to 630 ºC Four-wire RTD/Ω input terminals Accuracy: ±0.02 ºC Pt 385, 100 Ω, –80 to 100 ºC Isolated pressure module input Accepts Fluke 700 & 525A series pressure modules Thermocouple input/output terminal Accuracy: ± 0.1 ºC –100 to 800 ºC (type K)

  8. 7526A Product Overview (Right Half) Isolated input controls dc voltage/current Switch test mode 24 Vdc power supply 250 Ω HART resistor Pressure input mode • Numeric and secondary function keys Cursor controls • Input/Output Function Keys • dc voltage/current • TC/RTD • Pressure mode • Unit type • Stby/Operate • Isolated input terminals • Allows simultaneous source and measure • dc voltage (0 to 100 V) • dc current measurement (0 to 50 mA) • Switch-test input • 24 V dc loop power supply

  9. 7526A Product Overview (Rear Panel) Power line voltage selector and fuse compartment Power switch AC power inlet 120/240 AC Service port Chassis ground GPIB IEEE-488 RS-232 Port

  10. 7526A Product Features • STBY/OPR • In Standby mode, changes to the output value are not applied until you select the Operate mode • In Operate mode, each change to the output value is applied immediately • Voltages more than 30 V are not applied. The mode automatically reverts to Standby for safety • In Thermocouple mode, move through the thermocouple types (including millivolts) • In RTD/Ohms mode, move through the RTD types (including ohms) • In Pressure mode, move through the pressure units Select a secondary function from the numeric keypad

  11. 7526A Product Features • HART • Enables a 250 Ω internal resistor • Without the loop resistance in place, the DC power supply would “short out” the communicator’s AC voltage signal • The presence of a loop resistor in the circuit prevents the DC power supply from “loading” the AC voltage signal by the communicator. • Some minimum loop impedance is needed so that the HART signal can be ‘seen’ or ‘read’ by a HART master, like a HART communicator or HART modem. HART transmitter Power supply ↓ ∿↓ 250 < R < 1100 Computer HART communicator ∿ • LOOP PWR • Turns on an internal 24 Vdc power supply in series with the current measurement circuit • Used to power 4 – 20 mA transmitters

  12. 7526A Product Features • RTD and Ohms Measure/Source • Measures & Sources all common RTD types • Stores CVD coefficients for up to five RTDs • Stores coefficients for one SPRT (ITS-90) • Resistance from 0 Ω to 4 kΩ. • RTD types: • Pt 385 100 Ω, 200 Ω, 500 Ω, 1000 Ω • Pt 3926 100 Ω • Pt 3916 (JIS) 100 Ω • Ni120 120 Ω • Cu 427 (Minco) 10 Ω • YSI 400 thermistor

  13. 7526A Product Features Thermocouple (TC) Measure/Source Measure & sources all common thermocouple types and basic millivolts from -10.0 to 75.0 mV Thermocouple types: B, C, E, J, K, L, N, R, S, T, U, XK, BP • Pressure Measure • The isolated pressure display uses the same physical pressure connector as the primary display • Both displays can show pressure at the same time • The same source can be shown in different pressure units

  14. 7526A Pressure Modules

  15. 7526A Product Features • Preset setpoints • Nine preset output setpoints can be stored and recalled for each of these output modes: • Voltage • Current • Each thermocouple type (including mV) • Each RTD type (includes each of the five custom curves) • Switch Test Mode • When in Switch Test mode the Calibrator can sense a contact transition (closed or opened) • At the time of the transition, the Calibrator records the value and units shown on the primary display

  16. 7526A Product Features • Remote Interface • RS-232 or GPIB (IEEE-488) can control 7526A from a PC • A serial to USB adapter cable is included as standard with the 7526A • Individual commands can be typed into terminal emulator software • MET/CAL compatible – write automated procedures and generate calibration reports

  17. 7526A Workload

  18. 7526A Summary Specs

  19. 7526A Summary Specs (cont.)

  20. Convert Spec in PPM to % Reading Convert a spec in ppm to % error of reading

  21. 7526A Standard & Optional Equipment • Included in the box: • 7526A Precision Process Calibrator • Users Manual CD-ROM • Getting Started • AC Mains Cord • Thermocouple Shorting Jumper • NIST Traceable Calibration Report • USB to Serial Adapter Cable • Options and accessories: • 5520A-525A Leads kit • Y7526A Rack Mount kit • Fluke 700 and 525A-P series pressure modules • MET/CAL with 7526A Function Select Code (FSC) • MET/CAL 7526A calibration procedure • Optional accredited calibration

  22. Questions ?

  23. Appendix – P/I Transmitter

  24. Appendix – I/P Transmitter

  25. Appendix – V/I Transmitter

  26. Appendix – RTD Transmitter

  27. Appendix – TC Transmitter

  28. Appendix – RTD Indicator

  29. Appendix – Temperature Switch Test

  30. Appendix – Pressure Switch Test