An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Development of X-ray Spectrometry for Nuclear Decay Data Applications at IFIN-HH Aurelian Luca Horia Hulubei National Institute of Physics and Nuclear Engineering (IFIN-HH), Magurele, Romania

  2. Work performed in the frame of the joint research project IFA Romania – CEA France no. C2-05/2012 (collaboration between IFIN-HH and CEA, DRT/LIST/LNHB, 2012-2015): “Creation of national standards for some emerging pharmaceutical radionuclides to ensure the radioprotection of patients and medical staffs”

  3. A new high-resolution spectrometry system is necessary at IFIN-HH for measurements of low energy γ- and X-rays emitted by the radionuclides to be studied for the project (177Lu, 186Re) and other applications. • The energy range (3…50) keV is of particular interest, as it cannot be covered by the existing high-resolution HPGe gamma-ray detector used in the Radionuclide Metrology Laboratory (RML) since 2007.

  4. The following components (Canberra) were purchased, assembled and preliminary tested: -Si(Li) detector (preamplifier included); -Digital spectrometer; -Notebook computer with GENIE2000 S511 and OriginPro v.9.0 (2012) software.

  5. Si(Li) detectorSL30170, with charge sensitive preamplifier model 2008B and cryostat model 7500: active diameter 6.2 mm, active area 30 mm2, 3 mm thick, 5 mm distance from the Be window (0.025 mm). High voltage: -500 V dc. Manufacturer specifications for a time constant 12 ms: FWHM=166 eV and FWTM=306 eV, for 5.9 keV (55Fe) • Digital spectrometer DSA 1000: DSP, MCA- 16384 channels, PHA/MCS, pole zero adjustment, pile-up rejector and dead time correction, pulse shaping, USB/RS-232, support for GENIE 2000 v.2.0A

  6. Experimental Results • Radioactive standard sources available: 55Fe [5.89 keV (25.02(30)%) XKaand(6.49…6.54) keV (3.40(7)%)XKb], T1/2=2.747(8) years 57Co [6.4 keV (50%), 7.1 keV (7%); g: 14.4 keV (9%)] 241Am [XL 12-22 keV (38%), g: 26.3 keV (2.4%) and 59.5 keV (36%)]. 55Fe point standard source: produced by RML; code SEG 7-872, activity: (3793±152) Bq, k=2 (27/02/2013); 7200 s measurement time; 50 mm distance above the detector

  7. Experimental Results -A spectral anomaly caused by a wrong electronic setup was corrected by the specialists from Canberra Romania (2013). After the correction: • Integral background counting rate (1022 channels): 0.234 counts/s • Experimental FWHM: 0.23 keV at 5.7 keV. • Detection efficiency for the mean energy (5.96 keV): 0.87·10-2s-1 Bq-1

  8. X-ray Spectra: 55Fe, 241Am 5.9 keV 6.5 keV

  9. Future Improvements • Resolving (deconvolution) complex X-ray and low energy g-rays spectral regions (e.g. 241Am) • Testing different types of sources to check the influence of self-absorption • Training for the use of GENIE 2000 software

  10. THANK YOU !