This presentation provides an overview of the 50 SIC crystals received in January 2004, detailing the quality control procedures, irradiation methods, and comparison of measured parameters with specifications. The distribution of measured dimensions, transmission properties, and correlation with SIC data are analyzed. The effects of lateral and front irradiation on the crystals are also discussed.
E N D
Results of the 50 SIC crystals Received in January 2004 E. Auffray, CERN PH-CMA
Introduction What has been done ? • They have all followed the “routine” QC procedure • Identification • Visual Inspection • Accoce2 measurements • 10 have been irradiated in Geneva hospital • lateral irradiation, 350Gy/h, 400Gy • 3 have been irradiated in lab27 • Front irradiation, 0.15Gy/h Presentation • Distribution of all measured parameters (compared to the specifications) • Comparison with SIC data
AF AR BF BR CR CF Lateral Dimension Measured dimensions- Nominal dimensions
Length Measured dimensions- Nominal dimensions
Face1 Face 2 Face5 Face6 Face3 Face4 Planearity
Distribution of (measured angle-90deg) Angle 1-3 Angle 1-6 Angle 2-3 Angle 2-6 Problem with the geometry No flat pack (No right angle)
Longitudinal Transmission 360nm 420nm 620nm
Longitudinal TransmissionComparison with SIC data 360nm 420nm 620nm
Dispersion Gradient of TTO Transversal transmission SIC
LY LY FNUF RNUF
LY & transmission Correlation with T @360nm No correlation with T@420nm
LY Non Uniformity& transmission No correlation with T420nm but correlation with gradTTO
Decay time Ly in 100ns / Ly in 1000ns (%)